Maxim Terekhov1, David Lohr1, Theresa Reiter2, Ibrahim A. Elabyad1, Michael Hock1, and Laura M. Schreiber1
1Chair of Molecular and Cellular Imaging, University Hospital Würzburg, Comprehensive Heart Failure Center, Wuerzburg, Germany, 2Department of Internal Medicine I, Cardiology, University Hospital Würzburg, Wuerzburg, Germany
Synopsis
7T cardiac MRI is a novel field with the
potential to increase the physical sensitivity and diagnostic value of the
clinical cardiac images. Multiple elements transmit/receive phased array coils
in combination with parallel transmit RF-technology allow for an improvement of
the image quality at ultra-high-field (B0≥7T) field strength. In this work, we
present initial experience with a new commercial 8Tx/16Rx thorax array coil
operating in a “pTX-Compatibility Mode” of the scanner for cardiac MRI at
7T.
Introduction
7T cardiac MRI (CMR) is a novel research field
with the potential to increase the physical sensitivity and diagnostic value of
clinical cardiac images when compared to clinical field strengths. Multiple
elements transmit/receive phased array coils in combination with parallel
transmit (pTx) RF-technology promises for an essential improvement of the image
quality in the ultra-high-field (B0≥7T) MRI. Adjustment of the
magnitudes and phases of the driving voltages for the individual elements
allows for the suppression of destructive interferences of the electromagnetic
field and promises better homogeneity of image contrast within the heart.
However, the routine application of the 7T cardiac arrays in full-pTX mode is
currently not feasible due to the complexity of both the B1-shimming
process and, especially, on-line analysis of the SAR-safety for each specific
subject. We present initial results on a
new commercial 8Tx/16Rx cardiac array designed to operate in a virtual
single-channel (“pTX-compatibility”) mode. Materials and Methods
Measurements were performed with the
approval of the local ethics committee using a Magnetom™ “Terra” 7T MR-scanner
(Siemens Healthineers, Erlangen) and a new commercial 8Tx/16Rx cardiac array (RAPID
Biomedical, Rimpar, Germany) with central symmetry of the coil elements (Figure
1a) which has been tested earlier as a 1Tx-channel-prototype [1]. The “pTX-compatibility
mode” mode is a concept allowing usage of pTX-arrays in virtual single TX mode
with predefined B1 phases of the voltages driving the RF-power
amplifier. This specific “phase vector” is fixed in the coil configuration but can
be changed by qualified researchers. This allows for, on one hand, a SAR-safe
“push-button” application of the array in clinical routine and, on the other
hand, keeping flexibility for the targeted B1 adjustment for a
specific subject or patients cohort. As an initial test, a default phase vector
set by the vendor was used. The voltage of an 1800 pulse with 1 ms
duration was determined by an integrated scanner flip angle calibration
routine. CMR was performed using vendor-provided GRE-sequence [1]. Cardiac triggering
was performed using the scanner’s integrated ECG-monitoring system. Imaging
parameters were TR/TE=59/3.6ms, image matrix=288x228 with FOV=340x320mm, and
GRAPPA acceleration factor R=3. The flip-angle (FA) was varied to adjust
optimal blood-tissue contrast such that minimal blood-flow related image artifacts
were observed. The study protocol included localization scans, four-chamber, and
short-axis-view CINE stacks, respectively, with 30 cardiac phases (retrospective ECG
triggering). Technical coil characterization was done using the vendor’s “coil
utility” protocol which comprises of acquisition and reconstruction of SNR and
g-factor maps. For comparison, the same protocol was repeated using a
commercial 1Tx/16Rx cardiac array (MRI.TOOLS, Berlin, Germany)
To explore the potential of the new array concerning
full pTX-based B1-shimming a customer-defined complex B1-vector was computed using the electromagnetic model of the array.
The EM-simulations of individual Tx-channels were done using CST MIcrowave
Studio (Dassault Systeme) in two human models “Duke” and “Ella” (IT’IS Foundation,
Zürich). The dedicated cost-function (Figure 1b) for the B1-vector
optimization was designed taking into account both array geometry and B1-profiles
of individual Tx-channels. It comprises (i) maximization of the mean value of
combined B1-field (ii) minimization of the weighted combination of
the mean gradient of B1 in x,y,z directions, and (iii) maximization
of the transmit efficiency. Results
Figure 2 shows the comparison of the four-chamber
and short-axis views, respectively, for both coils at flip angles optimized for
each coil. Figure 3 shows a comparison of selected slices from a
short-axis-view stack from basal, mid-ventricular, and apical positions at
systole and diastole. Figure 4 shows the comparison of g-factor maps of both
arrays for short-axis and long-axis views. Both long and short-axis views
demonstrate an increase of flow-induced artifacts with increasing flip-angle
for both arrays. The overall amount of such artifacts was found to be lower for
the new array probably due to a lower gradient of B1-field
in the anterior-posterior direction. The maximal FA allowed by the SAR-monitor
for the same volunteer (84kg) at the same protocol was 420 for the new
array using the default B1-vector and 400 for the 1Tx. The
g-factor maps of the tested 8Tx/16Rx array demonstrate lower average values at
acceleration factors R=3 and 4 compared to the 1Tx array. Figure 5 demonstrates
the results of the optimization of the B1-field of the new array in
the heart region of the “Duke” human-body model. Discussion
Overall we observed high coil efficiency and
more homogenous image quality which probably is a result of a more homogeneous B1-profile.
The observed improvement of the g-factor most probably is a result of the symmetrical
arrangement of the coil elements which allows for a more efficient acceleration
at oblique slice orientation. We expect that the usage of a dedicated customer-calculated
B1-vector in the near future will provide further improvement of the
homogeneity of the B1-profile and, thus, ensure even more homogeneous image contrast, particularly
in the posterior and lateral heart wall. Conclusion
The initial experience of measurements and
preliminary analysis is promising and suggests a high potential of the new
commercial 8Tx cardiac array using the virtual single-channel Tx mode with
user-defined B1-phase vector. Further optimization for clinical
usage can be done using dedicated customer B1-vectors computed for
specific patient cohorts. Acknowledgements
Financial
support: German Ministry of Education and Research (BMBF, grants: 01EO1004,
01E1O1504).
Stimulating discussions with Robin Heidemann and Ralph Kimmlingen (Siemens Healthineers), Titus Lanz, and Carsten Kögler (Rapid Biomedical) are greatly
acknowledged.
References
[1]
Terekhov, M, Elabyad, I, Kögler, C, et al. „Customized B1+-Shaping using
multi-Channel transceiver array prototype for 7 T cardiac MRI with central
elements symmetry“. Proc 28th Intl. Soc. Mag. Reson. Med. (Virtual
meeting, 2020)