Kathleen Ropella-Panagis1, Jesse Hamilton1, and Nicole Seiberlich1
1Department of Radiology, University of Michigan, Ann Arbor, MI, United States
Synopsis
Increasing
the spatial resolution of MR Fingerprinting can result in longer acquisition
times due to the need to sample a larger extent of k-space. In this work, a
novel data sampling scheme consisting of interleaved high- and low-resolution
spiral trajectories is introduced to achieve high-resolution maps during a
shortened acquisition time, which could have advantages for sequences that
require a breath-hold.
Introduction
MR Fingerprinting1 (MRF) is an imaging technique
that yields quantitative maps of multiple tissue properties from a single rapid
acquisition. Typically, high-resolution property maps require acquisition
patterns that collect high spatial resolution data, or a larger extent of
k-space. These acquisition patterns require more data sampling time, thereby
extending the scan time. However, the individual images used for MRF contain only
a fraction of the total k-space data and, therefore, high spatial resolution
data may not need to be acquired at every time point to enable higher
resolution mapping. In this work we explore the combination of low- and
high-resolution data to generate high resolution maps with a smaller increase
in scan time as compared to the collection of all high-resolution data. Methods
This study employed a 2D FISP MRF sequence2 with
1000 TRs and an initial adiabatic inversion. Two 48-arm spirals with FOV=300mm2
were designed: (1) a lower-resolution variable density spiral with an
undersampling factor of 24 in the central 25% of k-space and 48 at the periphery
of k-space, matrix size 192x192, resolution 1.6x1.6mm2, readout
duration 2.9ms, and minimum TR 5.4ms; and (2) a higher-resolution uniform
density spiral with an undersampling factor of 48, matrix size 300x300,
resolution 1.0x1.0mm2, readout duration 5.6ms, and minimum TR 8.1ms.
During the MRF acquisition, the low- and high-resolution spirals are
interleaved, and the number of higher resolution spirals varied to assess the effect
on image resolution. Figure 1(b) shows an example where 33% of the MRF images
(i.e., every third TR) are collected using the high-resolution spiral. The
non-uniform fast Fourier transform (NUFFT) was used to grid the spiral k-space
data to yield a time series of undersampled images. The low-resolution spiral
data were zero-padded in k-space to yield a matrix size of 300x300 to match that
of the high-resolution data. Figure 1(a) shows the reconstruction pipeline. For
each pulse sequence, a dictionary with 4,966 atoms was calculated for T1 values
[10:10:90,100:20:1000,1040:40:2000,2050:50:3000] and T2 values [2:2:8,10:5:100,110:10:300,350:50:1100].
Maps were generated using a low-rank reconstruction3 followed by
inner product matching with the dictionary.
The proposed method was evaluated in two simulation
experiments. First, a resolution phantom was generated as shown in Figure 1(c)
with a matrix size of 300x300. It contained six clusters of four equal-sized
squares with widths ranging from one to six pixels, each with different T1 and
T2 values; the surrounding pixels were set to T1=250ms, T2=68ms. Six different
acquisition schemes were simulated, where a different fraction of MRF time
points were sampled using the high-resolution spiral (100, 50, 33, 25, 20, and
0%). All acquisitions employed 1000 TRs; however, the total scan times varied
due to the differences in spiral readout duration. Maps were assessed by
measuring the mean T1 and T2 values from each square in the 2x2 square cluster
and the 10x10 pixel surrounding background. For the 33% mixed-resolution
sequence, the T1 and T2 values and full-width at half-maximum (FWHM) were
measured for one square of each size cluster to quantify spatial blurring. Second,
to assess the sensitivity to off-resonance blurring, FWHM of the point spread
function (PSF) for all six acquisition schemes was calculated from the sum of
the undersampled MRF images over time using a point object phantom (assuming
infinite relaxation times) for off-resonance values of 0 Hz, 50 Hz, and 150 Hz.Results
The simulation results (Figures 2-4) show that
high-resolution 1.0x1.0mm3 MRF maps can be collected in 22% less
time (6.3 seconds) using the proposed mixed-resolution spiral sampling compared
to a conventional high-resolution spiral trajectory (8.1 seconds). Acquiring at
least 33% of the data at high-resolution yields accurate T1 and T2 measurements
and minimal blurring in small objects, as shown in Figures 2, 3, and 4(b). Acquiring
less than 33% high-resolution data results in decreased T1 values, as shown in
Figure 2(a).
High-resolution spirals with longer durations may cause
image blurring in the presence of off-resonance. Figure 5 shows the PSF for
low-, high-, and mixed-resolution acquisitions at various off-resonance
frequencies. For off-resonance frequencies close to 0 Hz, the 33% mixed-resolution
acquisition provides improved spatial resolution (FWHM=1.57 pixels) compared to
the low-resolution acquisition (FWHM=2.27 pixels). At larger off-resonance
frequencies near 150 Hz, the off-resonance blurring dominates over the gain in nominal
spatial resolution; the FWHMs of all scans using the high-resolution
acquisition are larger than that of the low-resolution acquisition. Discussion
Using a combination of low-resolution and high-resolution
data for MRF can yield high-resolution maps with a shorter acquisition time
than when collecting all high-resolution data. Acquiring at least 33% of the
data with a high-resolution spiral yields maps with accurate property values
and minimal resolution loss. In contrast, simply zero-padding low-resolution
data to increase the matrix size results in substantial blurring and inaccurate
maps, most likely due to partial volume artifacts. While mixed-resolution
spirals may provide higher-resolution maps in the absence of off-resonance, a
deblurring algorithm may be necessary to mitigate spiral blurring to achieve
higher resolutions in the presence of off-resonance. Conclusion
A data sampling scheme for MRF is introduced consisting of
interleaved high- and low-resolution spiral trajectories, which can decrease
the acquisition time needed to collect high-resolution tissue property maps. Acknowledgements
Research support from Siemens Healthineers and NIH grant R01 HL 094557-10.References
[1] D. Ma et al., “Magnetic
resonance fingerprinting,” Nature, vol. 495, no. 7440, pp. 187–192, 2013.
[2] Y. Jiang, D. Ma, N.
Seiberlich, V. Gulani, and M. A. Griswold, “MR fingerprinting using fast
imaging with steady state precession (FISP) with spiral readout,” Magn. Reson.
Med., vol. 74, pp. 1621–1631, 2015.
[3] J. I. Hamilton et al.,
“Simultaneous multislice cardiac magnetic resonance fingerprinting using low
rank reconstruction,” NMR Biomed., vol. 32, no. 2, Feb. 2019.