Congyu Liao1, Xiaozhi Cao1, Ting Gong2, Zhe Wu3, Zihan Zhou2, Hongjian He2, Jianhui Zhong2,4, and Kawin Setsompop1
1Radiological Sciences Laboratory, Stanford University, Stanford, CA, United States, 2Center for Brain Imaging Science and Technology, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China, 3Techna Institute, University Health Network, Toronto, ON, Canada, 4Department of Imaging Sciences, University of Rochester, Rochester, NY, United States
Synopsis
In this work, we developed ViSTa-MRF, which combined Visualization
of Short Transverse relaxation time component (ViSTa)
technique with MR Fingerprinting (MRF), to achieve whole-brain myelin-water fraction
(MWF) and T1/T2/PD mapping at 1mm isotropic resolution in 10 minutes on a clinical 3T scanner. To achieve this fast acquisition,
the ViSTa-MRF sequence also leverages an efficient
3D-spiral-projection acquisition along with spatiotemporal subspace
reconstruction. With the proposed ViSTa-MRF method,
direct MWF mapping was achieved without a need for multicompartment
fitting. In comparison to conventional myelin-water
imaging, the ViSTa-MRF method can provide improved-SNR and faster acquisition with
high image-quality.
Introduction
Myelin-Water Fraction (MWF) mapping has shown
great potential in characterizing brain’s myelination and demyelination. Conventional
MWF-mapping utilizes a multi-echo spin-echo(1) or gradient-echo sequence(2) and multi-compartment fitting of the exponential
decay to extract the shorter relaxation time of myelin-water. However, such fitting is ill-conditioned and
prone to noise. To improve MWF-mapping, the ViSTa technique(3,4) was proposed for direct visualization of
myelin-water. Here, a specifically configured double inversion-recovery sequence
is employed, where the long T1-component is suppressed and those
signals from short T1-components of myelin-water are preserved. The
key challenges of ViSTa are the decreased SNR from the signal suppression and the
long acquisition time. Recently, MWF-mapping has also been performed using a
modified MR Fingerprinting sequence(5,6), specifically designed to provide good signal
separability between myelin-water and other tissue types. Nonetheless,
multicompartment fitting is still required to extract MWF, which can be
challenging, necessitating the use of predefined compartments with fixed T1
and T2 values(7).
In this work, we incorporate ViSTa into MR
fingerprinting (ViSTa-MRF), to improve the speed and accuracy of MWF-mapping.
With the ViSTa inversion-preparation, the first time-point of ViSTa-MRF
will contain direct myelin-water signal. By utilizing spatiotemporal subspace
reconstruction(8,9), the spatio-temporal encoding information
across the whole ViSTa-MRF time-series is employed to reconstruct this direct
myelin-water data at high acceleration, providing MWF-mapping without need for
multicompartment modeling. Using ViSTa-MRF along with an efficient
3D-spiral-projection acquisition (10), we demonstrate high-quality whole-brain 1mm3
MWF and T1/T2/PD maps in ~10 minutes.Methods
Pulse sequence: Figure1(A) shows the diagram of the proposed
ViSTa-MRF sequence, where each acquisition group consists of eight ViSTa blocks
and one MRF block. In each ViSTa block, a double inversion-recovery is
performed, with the first subsequent signal time-point labeled the “ViSTa
signal”. Through extended-phase-graph (EPG) simulation(11), Fig.1(B) shows that the myelin-water
signal with short-T1 is preserved in the ViSTa signal, while the
white-matter (WM), gray-matter (GM) and CSF are suppressed, which enables
direct myelin-water imaging. Figure1(C) shows the ViSTa-MRF signal curves with good signal separability
between the different tissue types. To increase the emphasis on the
encoding of the short-T1 signal, the ViSTa-MRF sequence acquires forty
FISP-readouts with 30° excitations after each ViSTa
block and repeat the block eight times with different spatial-encodings, as
reflected on the signal curves in Fig1.(C). After the ViSTa blocks, 500-time-point FISP-MRF
data are acquired. Between the acquisition blocks, a 90° saturation pulse with a
waiting time of 380ms is used to saturate flow effects and achieve steady-state
longitudinal magnetization. A total of 820 time points is acquired in each
acquisition group across 19.4s.
Acquisitions: 3D-spiral-projection trajectory was used for
ViSTa-MRF acquisition at 1-mm whole-brain on a 3T Siemens Prisma scanner with a
64 channel head-neck coil: FOV:220×220×220mm3, TR/TE=12/2.5ms with a
7.0ms spiral readout. 32 acquisition groups were acquired, where the in-plane
16x-undersampled spiral interleaves in each group were designed to rotate
around kx, ky, and kz axes by tiny-golden-angle to maximize the incoherence
between each time point (Fig.2(A)). This resulted in a scan time of 19.4s×32 = 10.3 minutes,
where the acceleration factor of each time point is 16×(220/2)×π/32=172.8.
With 8 ViSTa blocks, the acceleration factor of the ViSTa signal is 172.8/8=21.6.
Post-processing: Subspace
reconstruction was performed, where ViSTa-MRF dictionary
was generated and the first eight principal components were selected as the
temporal basis Φ shown in Fig2(A). The
desired ViSTa-MRF time-series x is expressed as x=Φc,
where c are the temporal coefficient maps. Figure2(B)
shows the flowchart of the subspace reconstruction, which could be described
as:
$$ min_{c}\bf || MFSΦc|| +\lambda_{1}||c||_{2}+\lambda_{1}||c||_{*} , $$
where S are coil sensitivities,
F are the non-uniform FFT operators, M are the undersampling
patterns, λ1 and λ2 are the
regularization parameters. As shown in Fig.2(C), the reconstructed c are used to generate the ViSTa-MRF time-series. Template matching is then applied
to extract T1/T2/PD maps, while the quantitative MWF map is derived using the
first time-point ViSTa image I(ViSTa) and the PD map I(PD):
$$ MWF =\frac{I_{(ViSTa)}}{I_{(PD)}\cdot S_{(myelin-water)}} , $$
where S(myelin-water) is the EPG-simulated
signal intensity in ViSTa image using nominal T1(120ms) and T2(20ms)
values of myelin-water. All post-processings were performed
using MATLAB and BART(12).Results
Figure3 shows the ViSTa images reconstructed by parallel imaging and
subspace reconstruction at R=21.6. Compared to a direct SENSE reconstruction of
the ViSTa data, the joint subspace reconstruction of ViSTa and MRF time-series
significantly reduces aliasing artifacts on the ViSTa image, demonstrating the benefit
in leveraging the joint spatiotemporal encoding information.
Figure4 shows whole-brain 1mm3 T1/T2/PD/ViSTa
and MWF maps in three orthogonal views, where high-quality results are
observed from this fast acquisition.
Figure5 shows the MWF maps from two representative slices, where the
MWF values from ViSTa-MRF across four WM regions are consistent with the literature
results(3). Low-frequency signal bias can be observed in our MWF results, likely from
B1+ inhomogeneity and the large FAs used in the MRF-ViSTa
sequence (ranged from 10°-82°), which we aim to correct
in future work as per(13).Discussion and conclusion
In this work, we developed a 3D
ViSTA-MRF sequence together with subspace reconstruction to achieve rapid
high-isotropic-resolution MWF and T1/T2/PD mapping in a single scan. In comparison to conventional myelin-water
imaging, the proposed method can provide improved-SNR, faster, and high-quality parametric mapping.Acknowledgements
This work was supported in part by NIH research grants:
R01-EB020613,
R01-EB019437, R01-MH116173, P41EB030006, and
U01-EB025162.References
1. Mackay A, Whittall K, Adler J,
Li D, Paty D, Graeb D. In vivo visualization of myelin water in brain by
magnetic resonance. Magn. Reson. Med. 1994;31:673–677
2. Du YP, Chu R, Hwang D, et al. Fast multislice mapping of
the myelin water fraction using multicompartment analysis of T2* decay at 3T: A
preliminary postmortem study. Magn. Reson. Med. 2007;58:865–870.
3. Oh SH, Bilello M, Schindler M,
Markowitz CE, Detre JA, Lee J. Direct visualization of short transverse
relaxation time component (ViSTa). Neuroimage 2013;83:485–492.
4. Wu Z, Bilgic B, He H, et al.
Wave-CAIPI ViSTa: highly accelerated whole-brain direct myelin water imaging
with zero-padding reconstruction. Magn. Reson. Med. 2018;80:1061–1073.
5. Chen Y, Chen MH, Baluyot KR,
Potts TM, Jimenez J, Lin W. MR fingerprinting enables quantitative measures of
brain tissue relaxation times and myelin water fraction in the first five years
of life. Neuroimage 2019;186:782–793.
6. Cao P, Cui D, Chan Q, Hui ES. A
Multi-Inversion-Recovery (mIR) Myelin Water Mapping (MWF)-MRF Sequence. In:
ISMRM. ; 2020. p. 0048.
7. Deshmane A, McGivney DF, Ma D,
et al. Partial volume mapping using magnetic resonance fingerprinting. NMR
Biomed. 2019;32:e4082.
8. Zhao B, Setsompop K,
Adalsteinsson E, et al. Improved magnetic resonance fingerprinting
reconstruction with low-rank and subspace modeling. Magn. Reson. Med.
2018;79:933–942.
9. Liang Z. Spatiotemporal
Imagingwith Partially Separable Functions. In: 2007 4th IEEE International
Symposium on Biomedical Imaging: From Nano to Macro. ; 2007. pp. 988–991..
10. Cao X, Ye H, Liao C, Li Q, He
H, Zhong J. Fast 3D brain MR fingerprinting based on multi‐axis spiral
projection trajectory. Magn. Reson. Med. 2019;82:289–301.
11. Weigel M. Extended phase
graphs: Dephasing, RF pulses, and echoes - Pure and simple. J. Magn. Reson.
Imaging 2015;41:266–295.
12. Uecker M, Ong F, Tamir J, et al. Berkeley advanced reconstruction toolbox. ISMRM 2015 p.2486.
13. Hamilton JI, Jiang Y, Ma D, et
al. Investigating and reducing the effects of confounding factors for robust T
1 and T 2 mapping with cardiac MR fingerprinting. Magn. Reson. Imaging
2018;53:40–51.