Ruibin Liu1, Zihua Qian2, Zhe Wu3, Yi-Cheng Hsu4, Caixia Fu5, Yi Sun4, Dan Wu1, and Yi Zhang1
1Key Laboratory for Biomedical Engineering of Ministry of Education, Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China, 2Department of Radiology, The Children’s Hospital, Zhejiang University School of Medicine, National Clinical Research Center for Child Health, Hangzhou, China, 3Techna Institute, University Health Network, Toronto, ON, Canada, 4MR Collaboration, Siemens Healthcare Ltd., Shanghai, China, 5MR Application Development, Siemens Shenzhen Magnetic Resonance Ltd., Shenzhen, China
Synopsis
CEST imaging benefits from
longer saturation duration and a higher saturation duty cycle. Dielectric shading
effects occur when the RF wavelength approaches the object size. Here, we
proposed a parallel-transmission-based CEST (pTx-CEST) sequence to extend the
maximum saturation duration at 100% duty-cycle and mitigate shading effects.
The maximum saturation duration in pTx-CEST was lengthened to 2170, 3150, and 4130ms
compared to 1050ms in non-pTx-CEST at TR of 3s, 4s, and 5s, respectively, leading
to a significant sensitivity enhancement. Besides, the optimal amplitude ratio and phase
difference between RF channels, manifesting circular or elliptical polarization,
help reduce the dielectric shading effects.
Introduction
Chemical Exchange Saturation
Transfer (CEST) imaging is capable of detecting endogenous low-concentration
metabolites containing water-exchangeable protons 1-3. The CEST effects measured are partly
determined by the factor of (1-exp(-tsaturaion/T1water)) 3, indicating that the longer the saturation
duration, the more sensitive CEST imaging will be. However, clinical
applications of CEST imaging are typically limited by either inadequate saturation
duration 4,5 or low saturation duty cycles 6,7. In addition, dielectric shading artifacts
occur when the RF wavelength approaches the size of the object 8, making it an obstacle for CEST imaging in the
human abdomen. In this work, a parallel-transmission-based chemical exchange
saturation transfer (pTx-CEST) imaging sequence is proposed to achieve
arbitrarily long saturation duration and a 100% duty cycle simultaneously for enhanced
CEST sensitivity, as well as to mitigate dielectric shading artifacts in CEST
imaging. Methods
The study was performed on a 3T scanner (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany) using a 2-channel body coil for RF transmission. The two transmit
channels in the CEST saturation module can be driven with arbitrary amplitude
ratios R and phase
differences (θ1 - θ0),
independently, using freely-defined external RF shapes through the
vendor-provided pTx functionality (Fig. 1). Comparison between the proposed pTx-CEST
and the conventional non-pTx-CEST sequences was performed in a bovine serum
albumin (BSA) phantom consisted of a flask with 2% agarose gel and two test
tubes filled with 10% and 5% BSA dissolved in phosphate-buffered saline (PBS). The
RF pulse shape and peak transmit voltage in the pTx-CEST sequence were the same
as those in the non-pTx-CEST sequence. Optimizations of amplitude ratio and
phase difference settings for best B1 homogeneity measured with a pre-conditioning
RF method 9 were performed on a 140-mm-diameter (D140) cylindrical
phantom filled with NiSO4∙6H2O and NaCl solution, and a
240-mm-diameter (D240) spherical phantom filled with Marcol oil. The two
vendor-provided phantoms were chosen to mimic the size of the human brain and
abdomen. Scans on the D140 and D240 phantoms were performed using a 20-channel
head coil and an 18-channel flexible coil for signal reception, respectively. The
uniformity of the B1 field map was calculated with the metric: SD(|B1-1|),
where SD referred to standard deviation. The optimal amplitude and phase
setting yielding the best homogeneity in the phantoms was used in the human
study of pTx-CEST scans. The in vivo
study was approved by the local Institutional Review Board, and written consent
forms were obtained from the volunteers recruited. Results
Fig.
2 demonstrates the
maximum saturation duration with 100% duty cycle was significantly lengthened
to 2170, 3150, and 4130 ms using the pTx-CEST sequence compared to 1050 ms in the
non-pTx-CEST sequence at the TR of 3s, 4s, and 5s, respectively. The amide-proton-transfer-weighted (APTw)
signals in the BSA tubes obtained from pTx-CEST were substantially higher than
those derived from non-pTx-CEST, indicating a remarkable sensitivity
enhancement by elongating the saturation duration. For the metric concerning B1
homogeneity, the optimal settings revealed essential circular polarization (CP)
with an amplitude ratio of 1 and a phase difference of 1000 between
channels in the D140 phantom (Fig. 3A-D), and elliptical polarization (EP) with
an amplitude ratio of 1.9 and phase difference of 1000 between
channels in the D240 phantom (Fig. 3E-H). Table 1 shows the comparison of
homogeneity metrics from circular and elliptical polarization, which was in
line with the previous studies 10,11. Consequently,
the EP mode was used in CEST imaging of the human abdomen because of its
similar size to the D240 phantom. The regions with abnormally high intensity in
muscle tissues of the image acquired at +3.5 ppm with non-pTx-CEST (Fig. 4A,
red arrow), possibly due to the dielectric shading effect, disappeared in pTx-CEST
(Fig. 4D, red arrow). Accordingly, the extremely low
in the APTw images of the non-pTx-CEST sequence vanished compared to that in the
pTx-CEST sequence (Fig. 4C, F, red arrows).Discussion & Conclusion
We proposed a pTx-CEST sequence to extend the maximum saturation
duration with a 100% duty cycle, which significantly enhanced the CEST imaging
sensitivity compared to the conventional non-pTx-CEST sequence. The optimal
setting for best B1 homogeneity of the two transmit channels exhibited
elliptical polarization when the object's size approached the RF wavelength. Experiments
regarding optimizations for B1 homogeneity were performed in phantoms in place
of human brains and abdomens because of the excessively long scan duration and
subject-level variations. It is worthwhile noting that we only switched the
CEST saturation module from the non-pTx mode to the pTx-mode without changing
the RF pulses in the readout sequence. Importantly, the maximum saturation
duration in the non-PTx mode was not able to be increased when the TR changed
from 3s to 5s. Thus, the limited maximum saturation
duration in the non-pTx mode was probably due to certain software restriction,
which was bypassed in pTx functionality. Nevertheless, running the pTx-CEST
sequence was still found to be monitored by the SAR safety watchdog in our
practical scans. In conclusion, this study demonstrated the feasibility of
arbitrarily long CEST saturation duration with 100% duty cycle using the pTx
functionality on the Siemens platform, and the RF settings for optimal B1
homogeneity varied depending on the target size.Acknowledgements
NSFC grant number: 81971605, 61801421. References
1. Ward K, Aletras A, Balaban RS. A
new class of contrast agents for MRI based on proton chemical exchange
dependent saturation transfer (CEST). J Magn Reson
2000;143(1):79-87.
2. Zhou J, van Zijl PC. Chemical exchange
saturation transfer imaging and spectroscopy. Prog Nucl Magn Reson Spectrosc
2006;48(2-3):109-136.
3. Van Zijl PC, Yadav NN. Chemical
exchange saturation transfer (CEST): what is in a name and what isn't? Magn
Reson Med 2011;65(4):927-948.
4. Zhou J, Blakeley JO, Hua J, et al. Practical data acquisition method for human
brain tumor amide proton transfer (APT) imaging. Magn Reson Med
2008;60(4):842-849.
5. Zhu H, Jones CK, van Zijl PC, et al. Fast 3D chemical exchange saturation transfer (CEST) imaging of the
human brain. Magn Reson Med 2010;64(3):638-644.
6. Schmitt B, Zaiß M, Zhou J, et al. Optimization of pulse train presaturation for CEST imaging in
clinical scanners. Magn Reson Med 2011;65(6):1620-1629.
7. Sun PZ, Lu J, Wu Y, et al. Evaluation
of the dependence of CEST-EPI measurement on repetition time, RF irradiation
duty cycle and imaging flip angle for enhanced pH sensitivity. Phys Med Biol 2013;58(17):N229.
8. Hoult DI. Sensitivity and power
deposition in a high‐field imaging experiment. J Magn Reson
Imaging 2000;12(1):46-67.
9. Chung S, Kim D, Breton E, et al. Rapid
B1+ mapping using a preconditioning RF pulse with TurboFLASH readout. Magn
Reson Med 2010;64(2):439-446.
10. Nistler J, Diehl D, Renz W, et al.
Homogeneity improvement using a 2 port birdcage coil. ISMRM 2007.
11. Murbach M, Neufeld E, Cabot E, et al. Virtual population‐based
assessment of the impact of 3 Tesla radiofrequency shimming and
thermoregulation on safety and B1+ uniformity. Magn Reson Med
2016;76(3):986-997.