Can Wu1,2 and Qi Peng3
1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Philips Healthcare, Andover, MA, United States, 3Department of Radiology, Albert Einstein College of Medicine and Montefiore Medical Center, Bronx, NY, United States
Synopsis
The 3D MAPSS T1ρ sequence
provides fast and accurate T1ρ quantification with high spatial fidelity using RF
phase cycling and variable flip angle in a MP-GRE sequence. However, the conventional
3D MAPSS sequence uses continuous-wave RF pulses for spin locking and is thus sensitive
to B0 and B1 field inhomogeneities. In this study, a 3D
MAPSS T1ρ sequence with adiabatic RF pulses
was implemented and was shown to be less sensitive to B0 frequency
offset on phantom compared to conventional 3D MAPSS T1ρ mapping. It was also successfully
applied to high-resolution 3D T1ρ mapping of knee cartilage and brain tissue.
INTRODUCTION
Quantitative
T1ρ mapping has gained increasing
attention due to its high sensitivity to low-frequency molecular motion process
that allows characterization of macromolecular composition and proton exchange
in tissue.1 Standard T1ρ preparation uses a composite continuous-wave (CW) RF
pulse for spin locking, but it is relatively sensitive to B0 and B1
imperfections. Self-compensating spin
lock pulses and 180° refocusing pulses can be used in the preparation to partially
compensate for B0 and B1 inhomogeneities to reduce
banding artifacts, but the off-center slices are still subject to quantification
inaccuracy. 3D Magnetization-Prepared Angle-Modulated Partitioned k-Space
Spoiled Gradient Echo Snapshots (3D MAPSS) is a fast MP-GRE technique using RF
phase cycling and a variable flip angle train to achieve fast T1ρ mapping with
high accuracy and spatial fidelity.2 Its quantification accuracy
however is compromised by the CW-T1ρ preparation. Instead, adiabatic RF pulses
with both amplitude and frequency modulations, such as the hyperbolic secant pulses
(HSn, n = 1, 2, 4, 8), have a wider bandwidth and are much less sensitive to B0
and B1 inhomogeneities, as shown in some previous studies.3,4
The purpose of this study was to implement a 3D MAPSS sequence with adiabatic T1ρ
preparation to achieve fast, accurate quantification with high spatial
resolution. The sequence was tested on phantom,
human knee and brain at 3T.METHODS
Figure 1 illustrates the 3D
MAPSS T1ρ sequences with continuous-wave (Figure 1a: CW-T1ρ) and adiabatic (Figure
1b: Adb-T1ρ) RF pulses used for spin locking. CW-T1ρ preparation was implemented
as a composite RF pulse consisting of 90°-TSL/2-180°-TSL/2-90°, where TSL is
the time of spin locking. Adiabatic full passage (AHP) RF pulses with HS8
amplitude and frequency modulations were used for Adb-T1ρ preparation. A group
of four AHP pulses following MLEV phase cycling were used for spin locking. For
CW-T1ρ, the last 90° hard pulse was used for phase cycling by reversing its
phase for two paired scans. For Adb-T1ρ, phase cycling was achieved by
reversing the frequency modulation for the second half of the last AFP pulse. Both
cases lead to a pair of scans with same TSL but opposite longitudinal
magnetization (Mz+ or Mz-) immediately before data acquisition. A complex
subtraction of data from the paired scans can eliminate the contaminating
signal from T1 relaxation.2 CW-T1ρ and Adb-T1ρ experiments were
performed on phantom (three pairs of tubes with 2%, 3% and 4% homogeneous agarose
concentrations), human knee and brain (male healthy volunteer, 45 years old) on
a 3T clinical scanner (Ingenia Elition, Philips Healthcare, The Netherlands). Table
1 summarizes the sequence parameters for each experiment. T1ρ maps were calculated
using a mono-exponential two-parameter fitting algorithm. The center slice
(slice 24) and two off-center slices (slices 8 and 41) of the phantom were
selected for quantification using a custom analysis tool developed in IDL
(Exelis VIS, Boulder, CO). The study was approved by local institutional review
board and consent was received from the participating subject. RESULTS
Figure 2 shows T1ρ maps of the phantom at the center slice (slice
24) and two off-center slices (slices 8 and 41) where frequency offset is
present. The T1ρ maps of center slice look similar and are free
of artifacts (Figure 2, top row) for both CW-T1ρ
and Adb-T1ρ. Severe artifacts can be observed in the two off-center slices for CW-T1ρ
(middle and bottom left), indicating that Adb-T1ρ is less sensitive to B0
field inhomogeneities. Quantitative results further confirmed that Adb-T1ρ
provides more consistent T1ρ measurements in the off-center slices compared to
CW-T1ρ. For the center slice (slice 24), both CW-T1ρ and Adb-T1ρ provided very
consistent T1ρ values (largest difference < 2%) for tubes with the same
agarose concentration. However, for slice 8, CW-T1ρ values were (38.21±1.28 ms,
34.33±1.19 ms), (53.23±2.28 ms, 43.59±1.25 ms), (83.41±7.85 ms, 65.62±3.17 ms) for
the three pairs of tubes, indicating significant variation of the T1ρ values (largest
difference = 27%) for tubes with same agarose concentration. In contrast, the T1ρ
values from Adb-T1ρ were more consistent as follows, (49.00±1.72 ms, 51.45±1.96
ms), (67.55±2.45 ms, 63.16±2.09 ms), (100.32±5.22 ms, 95.89±2.95 ms) for three
pair of tubes (largest difference = 7%). Similar observations were found for slice
41. Figure 3 shows good delineation of the knee cartilage for both CW-T1ρ and
Adb-T1ρ maps without apparent artifacts or boundary blurring. Figure 4
illustrates example CW-T1ρ and Adb-T1ρ maps with isotropic resolution of the
brain in sagittal (top row), coronal (middle row), and axial (bottom row) views.DISCUSSION
3D MAPSS T1ρ mapping with
adiabatic RF pulses was successfully implemented and the phantom experiment demonstrated
that Adb-T1ρ is less sensitive to frequency offset compared to CW-T1ρ, suggesting
Adb-T1ρ may provide more accurate T1ρ measurement in conditions where magnetic
field is inhomogeneous. The Adb-T1ρ sequence also provided good image quality for
human knee and brain T1ρ mapping. Future study is warranted to quantitatively
compare the performance of Adb-T1ρ to conventional CW-T1ρ in the presence of B0
and/or B1 inhomogeneities in human knee and brain applications. CONCLUSION
3D MAPSS Adb-T1ρ offers fast and high
resolution T1ρ mapping as 3D MAPSS CW-T1ρ and is more robust to B0 inhomogeneities to potentially obtain more accurate
T1ρ quantification.Acknowledgements
None.References
1. Wang
Y, Zhang Q, Li X, Chen W, Ahuja A, Yuan J. T1ρ magnetic resonance: basic physics principles and applications in knee
and intervertebral disc imaging. Quant Imaging Med Surg 2015; 5(6): 858-885.
2. Li
X, Han ET, Busse RF, Majumdar S. In vivo T1ρ mapping in cartilage using 3D
magnetization-prepared angle modulated partitioned k-space spoiled gradient
echo snapshots (3D MAPSS). Magn Reson Med. 2008; 59(2):298-307.
3. Schuenke P, Koehler C, Korzowski A, Windschuh J,
Bachert P, Ladd ME, Mundiyanapurath S, Paech D, Bickelhaupt S, Bonekamp D,
Schlemmer H, Radbruch A, Zaiss M. Adiabatically prepared spin-lock approach for
T1ρ-based dynamic glucose enhanced MRI at
aultrahigh fields. Magn Reson Med 2017; 78(1):215-225.
4. Okuaki T, Takayama Y, Nishie A, Ogino T, Obara M,
Honda H, Miyati T, Van Cauteren M. T1ρ mapping improvement using stretched-type
adiabatic locking pulses for assessment of human liver function at 3T. Magn
Reson Imaging 2017; 40:17-23.