Zhongliang Zu1, Fatemeh Adelnia1, Kevin Harkins1, Feng Wang1, and John Gore1
1Vanderbilt University Medical Center, Nashville, TN, United States
Synopsis
Spin-lock imaging at low locking amplitudes are sensitive to the effects of water
diffusion in intrinsic gradients and may provide information on tissue
microvasculature. However, although composite pulse preparations have been used
to reduce artifacts due to B0 and B1 inhomogeneities,
there are still residual errors. Here, we analyze the source of these errors and developed an
approximate theoretical analysis to correct these errors.
Simulations and experiments on
a healthy human subject show that these errors are mainly due to B0 shifts,
and can be reduced by this correction approach when the B0 shift is
relatively small compared with w1.
Purpose
Spin-lattice relaxation rates in the rotating frame (R1ρ=1/T1ρ)
measured by spin-lock methods at low locking amplitudes (w1) are sensitive to the effects of water
diffusion in intrinsic gradients and may provide information on tissue
microvasculature1,2, but accurate estimates are challenging in the
presence of B0 and B1 inhomogeneities. Although composite
pulse preparations (90x-τy/2-180y-τ-y/2-90x)
have been developed to compensate for non-uniform fields3, there are residual image artifacts and errors in quantification of R1ρ
and their dispersion at low w1. Here, we analyzed the source of these errors
and developed an approximate theoretical analysis that provides a means to
correct them.Methods
When ω1
is comparable to a local B0 shift (Δωoff), the
magnetization nutates about the effective field. When there is also a B1
imperfection, the nominal 90° excitation pulse rotates the magnetization to a zenith angle of
only α = β·π/2, in which β is the ratio of the actual flip angle to the nominal
flip angle. Fig. 1 shows the nutation of the magnetization about the
effective field at angle = θ to the effective field. The effective field itself
has a zenith angle of φ and an amplitude of ω1_eff,
$$φ=tan^{-1}(\omega_1/\Delta\omega_{off})$$ (1)
$$\omega_{1\_eff}=\sqrt{\omega_1^2+\Delta\omega_{off}^2}$$ (2)
In spin-lock
imaging after the composite preparation cluster, the magnetization nutates back
into the transverse plane, in which case image artifacts
are reduced. However, a portion of the magnetization is not in the transverse
plane during the preparation period, and so does not experience R1ρ decay,
and subsequently causes errors in the quantification of R1ρ. Here,
we simply separate the magnetization after the first 90x excitation
pulse into two components: a transverse component which experiences R1ρ
decay, and a residual longitudinal component. The transverse component (f⏊)
can be calculated as the average of the projection of magnetization onto the
transverse plane with θ varying from 0 to θτ/2=2π·τ/2·ω1_eff
(τ = the spin-lock time) and assuming no R1ρ decay,
$$ f_⏊(\theta)=\frac{\int_{0}^{\theta_{\tau/2}}\sqrt{(\sin(\alpha-φ)\cdot\sin(\theta))^2+(\mid\cos(\alpha-φ)\sin(φ)\mid+\mid\sin(\alpha-φ)\cos(φ)\mid\cos(\theta))^2}d\theta}{\theta_{\tau/2}}$$ (3)
The residual
longitudinal component will experience relatively slow R1 recovery. If
there are no R1ρ or R1 decays, the spin-lock signal after
the composite pulse preparation would approximately be the sum of signals from
these two components. Assuming the magnetization is 1, the residual
longitudinal component is then 1 - f⏊. When there are R1ρ
and R1 decays, the spin-lock signal after the composite pulse
preparation is approximately these two components modified by relaxation. By
assuming R1ρ>>R1 and rewriting f⏊(θ)
with f⏊(τ), the normalized spin-lock signal in
the presence of B0 and B1 shifts (Soff) can then be described by,
$$\frac{S_{off}(\tau)}{S_0}\approx f_⏊(\tau)exp(-R_{1\rho}\tau)+1-f_⏊(\tau)$$
(4)
R1ρ can
be obtained by fitting Soff/S0 to Eq. (4).
We evaluated this correction approach
through both simulations of the Bloch equations and experiments on a healthy human
subject. Simulations were performed of both a single-pool and a two-pool model
of tissue water. The two-pool system was used to mimic self-diffusion with pool
A representing water molecules that do not experience any intrinsic gradients
and pool B representing water molecules that diffuse through intrinsic
gradients. Self-diffusion through intrinsic gradients alters the signal
similarly to proton exchange between environments with different resonance frequencies.
Spin-lock images were acquired at
3T with a TSE readout and with τ of 2, 12, 22, 32, 43, 52, 62, and 72ms, and w1 of 0, 20, 50 and 100Hz.
A B0 map was obtained by two images acquired with a TFE sequence with
TEs of 2.14ms and 5.34ms, TR of 7.35ms, and flip angle 30º. A B1 map was obtained by two images acquired with TSE sequence
with excitation flip angles of 80º and 160º, TR of 677ms, and TE of 10ms.
All images were
acquired with a 32-channel head coil on a 3T Philips scanner
with an image resolution of 1×1×4mm.Results
Fig. 2
shows the simulated R1ρ dispersions with a B0
shift, a B1 error, and both B0 and B1 shifts, with
and without correction. Note in Fig. 2e and 2f that the artifact can be reduced by using the correction
approach. Also note that there are artifacts in the presence of a B0
shift, but not a B1 shift, suggesting the source of the artifact is mainly
B0 offsets. Fig. 3 shows images acquired from a healthy human
subject. Note in Fig. 3b that there are severe B0 shifts in the
region indicated by the red arrow. Note in Fig. 3d that the artifact from this
region is not corrected, especially at low w1 (i.e. 20 and 50 Hz). In contrast, the artifact from ROIs
with small B0 shift in Fig. 3e can be corrected at low w1. Although the R1ρ dispersion from the
region of large B0 shift cannot be corrected with ω1 of
20 and 50 Hz, it is corrected at 100 Hz. Fig. 3f and 3g show maps of ΔR1ρ, the subtraction
of two ΔR1ρ images acquired with
ω1 of 0 and 100Hz, without and with correction, respectively. Note
that the artifact in Fig. 3f (red arrow) is reduced in Fig. 3g.Discussion and Conclusion
The
R1ρ artifact at low w1 can be reduced using
this correction approach when there are small B0 shifts.Acknowledgements
No acknowledgement found.References
1.
Spear JT, Gore JC. Effect of diffusion in magnetically inhomogeneous media on
rotating frame spin-lattice relaxation. J Magn Reson 2014;249:80-87
2.
Zu ZL, Janve V, Gore JC. Spin-lock imaging of intrinsic susceptibility
gradients in tumors. Magn Reson Med 2020;83(5):1587-1595
3. Witschey WR, 2nd, Borthakur A, Elliott
MA, et al. Artifacts in T1 rho-weighted imaging: compensation for B(1) and B(0)
field imperfections. J Magn Reson 2007;186(1):75-85.