Shu Zhang1, Xinzeng Wang2, F. William Schuler1, R. Marc Lebel3, Mitsuharu Miyoshi4, Ersin Bayram2, Elena Vinogradov5, Jason Michael Johnson6, Jingfei Ma7, and Mark David Pagel1,7
1Cancer Systems Imaging, MD Anderson Cancer Center, Houston, TX, United States, 2Global MR Applications & Workflow, GE Healthcare, Houston, TX, United States, 3Global MR Applications & Workflow, GE Healthcare, Calgary, AB, Canada, 4Global MR Applications & Workflow, GE Healthcare Japan, Tokyo, Japan, 5Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 6Neuroradiology, MD Anderson Cancer Center, Houston, TX, United States, 7Imaging Physics, MD Anderson Cancer Center, Houston, TX, United States
Synopsis
Chemical exchange saturation transfer (CEST) measurements
can be compromised by a low signal-to-noise ratio (SNR) due to the small CEST
contrast in vivo. Deep learning-based image reconstruction (DL Recon) can
enhance image SNR without losing image resolution or altering the image
contrast, hence has the potential to improve quantitative CEST measurements. In
this study, we investigated the improvement to CEST quantitation by DL Recon in
glioma patients. We found that DL Recon substantially reduced the noise in the
MTRasym maps and improved the lesion conspicuity.
Introduction
In vivo chemical
exchange saturation transfer (CEST) contrast is usually small, and its
quantitation can be compromised by a low image signal-to-noise ratio (SNR).1
Deep learning-based image reconstruction (DL Recon) enhances image SNR without decreasing
image resolution or altering the image contrast. In a previous study, we have demonstrated
that DL Recon can compensate for low SNR in CEST imaging, as was
quantified using phantoms and qualitatively in 3 patients.2 The SNR
gain enables CEST imaging with higher spatial resolution without or with only a
mild increase in scan time. Alternatively, the SNR gain by DL Recon allows
faster CEST acquisition through parallel imaging with a higher acceleration
factor.
In this study, we investigated the improvements of CEST quantitation by DL
Recon in 10 glioma patients.Methods
Ten patients were enrolled in this IRB-approved study. 8 patients were scanned
on a 3T human scanner (GE Healthcare, Discovery 750) using an 8-channel head
coil. 2 patients were scanner on a PET/MR human scanner (GE Healthcare, Signa)
using an 8-channel head coil. 29 CEST images with saturation from -7 to 7 ppm with
equal intervals were acquired using a saturation power of 2.0 μT and a
saturation time of 2 sec.3 A reference image was acquired without
CEST saturation. 11 WASSR images from -1.88 to 1.88 ppm with equal
intervals were acquired in the same scan for B0 inhomogeneity correction.4 All the images were acquired using a single-shot fast spin echo sequence
with a matrix size of 256 and parallel imaging (ASSET = 2). MTRasym
was calculated for CEST quantitation. In addition to the standard image
reconstruction (standard recon), a vendor supplied DL Recon was used to
reconstruct images using the same CEST raw data.5 The DL Recon network
was based on a deep convolutional residual encoder pre-trained to reduce image
noise without a loss in spatial resolution. The reconstructed CEST images were
registered to the reference image to reduce motion artifacts. A tumor ROI and a
background ROI containing homogeneous brain tissue were manually drawn on the
reference images. The standard deviation of the MTRasym within the background ROI was used to characterize
the noise level in the MTRasym
maps. The DL Recon’s SNR improvements were calculated by dividing the MTRasym
within the tumor ROI by the standard deviation of the background ROI. The tumor
ROI-averaged MTRasym and the SNR values using the two reconstruction methods were evaluated using Wilcoxon
matched-pairs signed rank test. P < 0.05 was considered statistically
significant.Results
Fig. 1 compares the reference images, MTRasym
maps at 3.5 ppm, and MTRasym maps averaged between 3.0-4.0 ppm using
standard recon and DL Recon. The images using DL Recon had lower noise and were
sharper than those using standard recon (Fig. 1a vs. 1d). The images using DL
Recon generated MTRasym maps at both frequency ranges had less noise,
and therefore there was greater tumor conspicuity (Fig. 1b vs. 1e and 1c vs.
1f). Fig. 2 compares the MTRasym at 3.5 ppm and averaged between 3.0-4.0
ppm within the tumor ROI using standard recon and DL Recon. The tumor ROI-averaged
MTRasym at both frequency ranges using the two reconstruction
methods was similar for all patients (Fig. 2a, p = 0.322; 2b, p = 0.492), indicating that DL
Recon did not alter the CEST contrast. The tumor ROI-averaged MTRasym
divided by the standard deviation of the background ROI showed that DL Recon had
significantly improved SNR of the CEST measurements (Fig. 3a, p = 0.002; 3b, p
= 0.016).Discussion and Conclusion
Our study showed that the DL Recon can reduce the noise in the
CEST images and the calculated MTRasym maps, leading to better lesion
conspicuity. With the SNR improvement, DL Recon can potentially improve a voxel-wise
analysis within clinically feasible scan times. The results demonstrated that
DL Recon was applicable to CEST imaging and may improve its clinical
applications.Acknowledgements
This work was supported in part by the Odyssey Program and
Cockrell Foundation Award for Scientific Achievement at The University of Texas
MD Anderson Cancer Center (S.Z.).References
- Zhou J,
Blakeley JO, Hua J, et al. Practical data acquisition method for human brain
tumor amide proton transfer (APT) imaging. Magn Reson Med. 2008; 60(4): 842-849.
-
Zhang S, Wang
X, Schuler FW, et al. Deep Learning Reconstruction improves CEST MRI. Proc. Of
ISMRM 2020, 3100.
-
Miyoshi M,
Matsuda T, Kabasawa H. CEST imaging with phase cycled rectangular RF
preparation pulse: analytical solution, simulation and phantom study.
Proceedings of ISMRM 2014, 3299.
-
Kim M,
Gillen J, Landman BA, et al. Water saturation shift referencing (WASSR) for
chemical exchange saturation transfer (CEST) experiments. Magn Reson Med. 2009;
61(6): 1441-1450.
-
Lebel RM. Performance characterization of a novel deep
learning-based MR image reconstruction pipeline. arXiv
preprint arXiv:2008.06559 2020.