Navid PourramzanGandji1, Christopher T. Sica2, Gary W. Yang3, Hannes Wiesner4, Soo Han Soon4, Xiao-Hong Zhu4, Michael Lanagan5, Wei Chen4, and Qing X. Yang1
1Neurosurgery, Pennsylvania State University, Hershey, PA, United States, 2Radiology, Pennsylvania State University, Hershey, PA, United States, 3Pennsylvania State University, Hershey, PA, United States, 4Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, MN, United States, 5Materials Science and Engineering, Pennsylvania State University, State College, PA, United States
Synopsis
Prior work has
examined the effects of uHDC material upon phased array coils. Several
beneficial effects have been observed: reduction of per-channel noise and
inter-channel noise correlations, and enhancement of SNR. Due to the high dielectric
constant the electric field will be confined to the dielectric material, which
leads to a reduction of the primary noise source, the conservative electric
field. As a result the non-conservative field becomes more dominant within the
sample, with a resulting increase of B1 field intensity and more
substantial SNR enhancement.
Introduction
It has been shown
that the usage of ultra-high dielectric materials (uHDC) with RF coils can
enhance the B1 field and signal-to-noise
ratio (SNR) across a range of human magnets (3T - 10.5T) due to the strong
displacement current induced in the uHDC material(1,2). In addition, a significant reduction of noise by the uHDC
material has also been observed as a contributing factor to the SNR enhancement(3,4). This de-noising effect is thought to be induced by the reduction
of the conservative electric field within the sample by uHDC materials(3). The goal of this investigation is to determine this
effect upon receive array coils commonly used for clinical MRI at 3T.Methods
Theory: The
electric field created by the RF coil consists of two
components, a conservative and a non-conservative field:
$$ \overrightarrow{E}_{total} = \overrightarrow{E}_{Conservative} + \overrightarrow{E}_{NonConservative} = -\overrightarrow{\triangledown\phi} - \frac{\partial \overrightarrow{A}}{\partial t} (1)$$
where $$$\phi$$$ is the electric scalar potential, $$$\overrightarrow{A}$$$ is the vector potential. Both E-field components
produce currents that induce B field
and losses in the tissue samples during transmission and reception. However, the
conservative E-field is undesirable
because it does not interact with the magnetization of the nuclear spin system and
only generates noise. We have shown as a proof of concept that using uHDC
material, the conservative E-field in the sample can be reduced, leading to a
noise reduction and ultimate SNR improvement with a surface coil. The goal of
this study is to demonstrate and apply this effect for a commonly used 3T 4
channel surface phased array coil.
Experiment: A 4 channel Flex array
with each coil dimension 20 cm x 11 cm was used on a 3T system (PrismaFit, Siemens
Healthineers, Erlangen, Germany). Data were acquired both without and with uHDC
ceramic blocks with εr~880 (Figure 1). The dimension of the uHDC blocks was 8.7
x 10.8 x 2.7 (W x L x H cm3). Four blocks were placed in the middle of each
loop. A 2D interleaved spoiled GRE sequence (2 x 2 x 4 mm3, FA = 2 degrees) with
integrated noise pre-scans was acquired to calculate SNR maps. Flip angle maps
were acquired with a 2D pre-saturated TurboFlash sequence(5) (Saturation Angle = 60 degrees), with scan
geometry matched to that of the GRE scan. The noise standard deviation and
noise correlation matrices were calculated from the noise pre-scans for both
baseline and uHDC blocks. The SNR was calculated with the SNR Units Method
divided by the B1+ map to
remove the flip angle effects(6).Results and Discussion
As can be seen in Figure 2, in the presence of the uHDC the Noise Standard
Deviation (Std) for each individual coil element was reduced. The noise
correlation between adjacent coil elements showed an even more significant
reduction with uHDC material (Figure 3). This effect of the uHDC material is
due to the shielding of the conservative E
component by the uHDC material (3). The induced E field
not only decreases the conservative component and associated noise but also create
a strong displacement current that increases the B1 field within the sample. Therefore, the SNR increases
synergistically. As shown in Figure 4 the SNR was increased up to 4-fold next
to the surface of the phantom. However, an SNR drop around the edges of the
dielectric material was also observed.Conclusion
We
demonstrated experimentally that employing the uHDC materials in a standard clinical
phase array coil could reduce the total noise as well as enhance the B1 field and thereby increase the SNR substantially.
We showed that proper placement of uHDC materials with respect to the phase
array coil could effectively reduce the noise correlation by decreasing the
conservative E component. Acknowledgements
This work is supported by NIH grant U01EB026978. References
1. Sica
CT, Rupprecht S, Hou RJ, Lanagan MT, Gandji NP, Lanagan MT, Yang QX. Toward
whole-cortex enhancement with a ultrahigh dielectric constant helmet at 3T.
Magnetic Resonance in Medicine 2019;0(0).
2. Navid P. Gandji, Myung-Kyun Woo,
Sebastian Rupprecht, Michael Lanagan, Bei Zhang, Riccardo Lattanzi, Russell L.
Lagore, Jerahmie Radder, Gregor Adriany, Kamil Ugurbil, Brian Rutt a, Yang QX.
The displacement current wave on a high dielectric constant (HDC) helmet and
its effect on B1 Field at 10.5 T (447 MHz). Virtual 2020.
3. Chen W, Lee BY, Zhu XH, Wiesner HM,
Sarkarat M, Gandji NP, Rupprecht S, Yang QX, Lanagan MT. Tunable Ultrahigh
Dielectric Constant (tuHDC) Ceramic Technique to Largely Improve RF Coil
Efficiency and MR Imaging Performance. IEEE Transactions on Medical Imaging
2020;39(10):3187-3197.
4. Navid Pourramzangandji , Christopher
Sica , Byeong-Yeul Lee , Hannes Wiesner , Maryam Sarkarat , Michael Lanagan ,
Xiao-Hong Zhu , Wei Chen , Yang Q. Large denoising effect of ultrahigh
dielectric constant (uHDC) materials. ISMRM. Virtual2020.
5. Chung S, Kim D, Breton E, Axel L.
Rapid B1+ mapping using a preconditioning RF pulse with TurboFLASH readout.
Magn Reson Med 2010;64(2):439-446.
6. Kellman P, McVeigh ER. Image
reconstruction in SNR units: A general method for SNR measurement†. Magnetic
Resonance in Medicine 2005;54(6):1439-1447.