Jeremiah Hess1, Marianne Black2, Feliks Kogan2, and Brian Hargreaves1,2,3
1Department of Bioengineering, Stanford University, Stanford, CA, United States, 2Department of Radiology, Stanford University, Stanford, CA, United States, 3Department of Electrical Engineering, Stanford University, Stanford, CA, United States
Synopsis
Flexible extremity coil-arrays for knee imaging in MRI can improve
coil positioning and comfort to patients in the scanner and wrap tightly around
the knee, placing coil elements close to anatomy of interest. Despite this,
flex coils have not been widely adapted due to perceived superiority of the SNR
of transmit-receive coils. In this study, we analyzed the SNR of flex coil-arrays
versus a rigid coil-array in phantom and in vivo to determine which
coil-arrays had improved SNR. Preliminary results suggest that flexible
coil-arrays show comparable or increased SNR and generally more uniform SNR
over rigid coil-arrays.
Introduction
MRI is commonly used in the clinic for knee evaluation and
diagnosis, with around 1.25 million clinical knee scans performed annually in
the US.1 Standard clinical knee imaging protocols for MRI call for
dedicated transmit and receive (T/R) knee coils to improve the image quality
for knee evaluation.2 Alternatively, flex coils in MRI can improve
coil positioning and comfort to patients in the scanner and allow for coil
elements to be placed closer to the anatomy of interest. Furthermore, such coil-arrays
also incorporate more elements, potentially allowing better conditioning of
g-factor maps for parallel imaging. However, these flexible coils have not been
widely adopted due to the perceived superiority of T/R coils for extremity
imaging.
In this study,
we compared pixel-wise SNR maps using a GRAPPA-based image reconstruction and pseudo
multiple replica approach3 for a state-of-the-art dedicated hard-shell
T/R knee coil and flexible extremity coil-arrays, in both phantoms and healthy
human subjects. Methods
Images were acquired on a 3.0T
scanner (GE Healthcare, Milwaukee, WI, USA) using a quantitative double-echo in
steady-state sequence (qDESS). A single large cylindrical phantom, a phantom
consisting of 6 fluid-filled falcon tubes, and the right knee of two healthy
subjects were imaged to assess SNR. Phantoms were scanned using 3 coils: a
medium 16-channel receive-only flexible extremity coil (NeoCoil, Pewaukee, WI,
USA), a large 16-channel receive-only flexible extremity coil (NeoCoil,
Pewaukee, WI, USA), and a rigid 18-channel transmit-receive dedicated knee coil
(Quality Electrodynamics, Cleveland, OH). Two human subjects were scanned with the
medium flex coil and rigid coil. Image reconstruction for both datasets
were performed in the GE Orchestra framework using complex coil combination,
with the in vivo dataset employing ARC reconstruction.
We
utilized a Monte Carlo-based, pseudo-multiple replica approach outlined in
Robson et al.3 for 200 repetitions to calculate SNR maps for both
coil arrays (Figure 1). This method uses a multi-channel noise covariance
matrix from the scans to estimate the correlated Gaussian noise present in the
k-space data. For each repetition, a random sample of this noise is added to
the acquired k-space data and reconstructed. A standard deviation is calculated
between all replicated images to estimate pixel-wise noise standard deviations,
and the SNR is estimated using the acquired k-space data as the signal.
For the
phantom dataset, average SNR was calculated in the cylindrical phantom and 6-tube
phantom. Masks were used to isolate the outer third of the cylindrical phantom
from the inner portion to investigate the average SNR differences at the
surface and the interior of the phantom. For the 6-tube phantom, masking was
done to isolate the top tube and the bottom tube in the phantom to compare
average SNR differences between opposite regions on the surface of the phantom.
For the in vivo datasets,
single-slice regions of interest were segmented: central slices for lateral
femoral cartilage, medial femoral cartilage, lateral meniscus, and medial
meniscus, and single slices for ACL, PCL, and patellar tendon (Figure 2). Data
was analyzed by comparing the mean and standard deviation of the tissue SNR
across coils and subjects. Results
The processed SNR maps for the phantom datasets are shown in
Figure 3, with central slices shown for each phantom and coil. Of note, the
medium flex coil appears to have slightly better SNR on the outer third of the cylindrical
phantom, reinforced by the mean SNR results shown in Figure 3c-d. The rigid
coil only has comparably high SNR at the bottom region of the cylindrical
phantom. In the 6-tube phantom, the medium flex coil had higher SNR in both the
top and bottom tubes compared to the rigid and large flex coils.
The SNR
averages computed from the tissue segmented in vivo are shown in Figure 4. Of note, the two subjects seemed to differ in which coil had
overall better SNR performance. For the first subject, the flex coil performed
the best for every tissue analyzed, but for the second subject, the rigid coil
performed the best, except in the patellar tendon.Discussion
We examined the effect of rigid and flex coils on SNR in
phantoms and knees in vivo. First, looking at the phantom data, the SNR
at the surfaces of the phantoms is more uniform when imaged with the medium
flex coil, as well as notably improving SNR at the surface compared to the
rigid coil. This overall improvement in surface SNR is most likely due to the
flex coil wrapped tightly around the phantom, as opposed to the rigid coil where
the phantom sits on the bottom surface of the coil.
In the in
vivo data, the differences in SNR between the two coils appear to differ
between subjects. One external factor that could explain this would be subject
size differences, potentially causing the substantial variations in SNR seen
here. Notably, the SNR of the patellar tendon was higher for the flex coil in
both subjects, which is probably due to the ability to wrap the flex coil
tightly on the anterior side of the knee.Conclusion
Preliminary phantom and in vivo knee data suggest that
flexible coil-arrays show comparable or increased SNR and generally more
uniform SNR over rigid coil-arrays.Acknowledgements
Research support provided by NIH R01-AR077604, NIH K24-AR062068, PHIND
Seed Grant, and GE Healthcare.References
1. Chaudhari A, Stevens K, Sveinsson B, Wood J, Beaulieu C, Oei
E.H., Rosenberg J, Kogan F, Alley M, Gold G, and Hargreaves B. Combined
5-Minute Double-Echo in Steady-State with Separated Echoes and 2-Minute
Proton-Density-Weighted 2D FSE Sequence for Comprehensive Whole-Joint Knee MRI
Assessment. Journal of Magnetic Resonance Imaging. (2019), 49(7), e183-e194.
2. Bolog, N.V., Andreisek, G. Reporting knee meniscal tears:
technical aspects, typical pitfalls and how to avoid them. Insights Imaging 7,
385–398 (2016).
3. Robson PM, Grant AK, Madhuranthakam AJ, Lattanzi R, Sodickson
DK, McKenzie CA. Comprehensive quantification of signal-to-noise ratio and
g-factor for image-based and k-space-based parallel imaging reconstructions.
Magn Reson Med. 2008;60(4):895-907.