Mark Gosselink1, Tijl van der Velden1, Hans Hoogduin1, Martijn Froeling1, and Dennis W. J. Klomp1
1University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
A 64 channel head coil receiver is presented
that operates with low noise coupling inside a continuously tuned 8-channel
transceiver coil. When used as a 72-channel receiver the setup outperforms the
default 32 channel receiver array in acceleration, demonstrated by high-resolution
MPRAGE images of the human brain.
Introduction
Ultra-high field MRI becomes more accessible by the rise in the clinical
grading of 7T MRI systems. The inherent gain in SNR (Signal-to-noise ratio) at
higher field strengths enables high-resolution imaging, but is more time
consuming and thus demands for acceleration techniques. Parallel imaging using
high-density phased array coils can increase the spatiotemporal resolution while
maintaining SNR by minimizing the g-factor. In this work, we show the
acceleration performance of a 64-channel receiver array in combination with an
8-channel pTx transceiver head coil when compared to a similarly designed 32-channel
headcoil.Methods
The coil setup consists of a high-density 64-channel head coil array
(WaveTronica BV, Utrecht, Netherlands) combined with an 8-channel parallel
transmit head-coil (Nova Medical, Wilmington MA, USA). A whole-body 7T MR
scanner (Philips Healthcare, Best, The Netherlands) standardly equipped with 32
receive channels and 8 parallel transmit (pTx) channels was upgraded using the
digital (dStream) architecture to expand the number of receivers to enable
reception with all 72-channels simultaneously.
In line with the 32-channel headcoil (Nova Medical, Wilmington MA, USA),
the high-density receiver array consists of coil strips shaped to fit around a
3D-printed helmet, 8 strips for the 32-channel array and 16 strips for the 64
channel array (Figure 1). Each coil strip of the 64-channel array has 24 x 30 mm
sized elements, which are decoupled using a predefined overlap in the axial
direction. Furthermore, preamplifier decoupling has been applied for all
elements by selecting the correct cable length. The interface box including 64
preamplifiers and driver boards is located at the end of the pTx head coil.
The output of the preamplifiers are connected via coaxial cables to
the digital receivers (ADC) located at the end of the bore.
Acceleration evaluation was performed based on acquired low flip angle
2D MS FFE scans with two dynamic acquisitions (flip angle: 10 degrees; TR =
5.5ms; 2 mm isotropic acquisition resolution; 240 x 240 x 240 cm3
field of view). Raw complex data per channel was exported for data
reconstruction and processing. Data was obtained once with the 32-channel coil setup
in the detunable transmitter array, and once with the 64 channel inside the
similar transmitter array albeit driven continuously in tuned mode, both loaded
with the same human head. Finally, high-resolution 3D MPRAGE sequence was
performed using all 72 channels of the 64 channel receiver array and the 8
channels of the transceiver array (flip angle: 6 degrees; TR = 9.0 ms; 0.8 mm
isotropic acquisition resolution; 250 x 250 x 200 cm3 field of view)
using SENSE factor 2.5 x 2.2 and Compressed SENSE factor 5.5.
Noise correlation and noise
covariance matrixes were
calculated from the second dynamic noise scan data. Thereafter,
1/g-factor maps were calculated by downscaling the fully sampled k-space. The
1/g-factor maps were obtained using the coil sensitivity maps and the noise
covariance matrixes according to Pruessmann et al [1].Results
Noise correlation and noise covariance matrixes
show low coupling (<0.3) between most of the local receive coils.
Furthermore, in the 64 channel setup, low coupling is observed between the continuously
tuned transmit elements and the high-density receiver array (Figure 2). With
the 64+8=72 channel receive setup, calculated 1/g-maps show a potential
acceleration up to 4 in one direction, and 4x3 (AP, LR) for two acceleration
directions while remaining a 1/g-factor threshold of >0.67 (Figure 3). Compared
to the 32-channel setup, this would be an acceleration of 1.7 for the same
threshold, (see Figure 4 for zoomed comparison between the coil setups). When
driven with 72-channels as receivers, high resolution MPRAGE could be obtained
from the human brain (Figure 5).Discussion
In line with a 64-channel receiver array using
fully overlapped elements in all directions [2], our increased channel counts
coil setup shows substantial improvements in imaging acceleration when compared
to a 32-channel array at 7Tesla. Moreover, as the transmit elements show low RF
coupling to the receiver array, they can be merged as receivers to facilitate a
72-channel array. The 0.8mm isotropic resolution MPRAGE obtained in 4:35
minutes illustrates the superb performance of the setup.Conclusion
We have shown the design of a 64-channel head
coil array that can be merged with an 8-channel transceiver to facilitate 72-channel
simultaneous signal reception at 7T. When compared to a similarly designed 32-channel
array, acceleration performance is increased. Moreover, with the 72-channel
coil, excellent MPRAGE images can be obtained from the human brain.Acknowledgements
No acknowledgement found.References
[1]: Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity encoding for fast MRI. Magn. Reson. Med. 1999;42:952–962 doi: 10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.3.co;2-j.
[2]: Uğurbil K, Auerbach E, Moeller S, et al. Brain imaging with improved acceleration and SNR at 7 Tesla obtained with 64‐ channel receive array. Magn Reson Med. 2019;82: 495–509.