Ming lu1,2, Bei Zhang3, John C. Gore1,2, and Xinqiang Yan1,2
1Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN, United States, 2Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States, 3Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States
Synopsis
Preamplifier
decoupling is a standard way of suppressing inter-element inductive coupling in
MRI receive arrays. A well-known method to evaluate and optimize RF coils before practical fabrication is to use circuit-level and electromagnetic (EM) field-level simulations. However, to be best of our knowledge, currently there is not a solid method to evaluate the preamp decoupling effect with circuitry and EM field analysis. In this work, we used RF circuit and EM co-simulation4 to evaluate the preamplifier decoupling effect and tested it on the conventional coil (LIC) and the recently introduced high impedance coil (HIC).
Purpose:
Preamplifier
decoupling is a standard way of suppressing inter-element crosstalk in MRI
receive arrays. The original preamplifier decoupling was mainly used to reduce
the coupling among non-adjacent elements in which the geometrically-overlap
decoupling1 is not feasible. Recently, a strong trend has been
arising to develop advanced preamp decoupling methods for flexible coils.
Unlike the original method in Roemer et al1, advanced preamp
decoupling methods are aiming at robust decoupling/noise suppression so the
overlapping area is not crucial2,3. A well-known method to evaluate
and optimize RF coils before practical fabrication is to use circuit-level and
electromagnetic (EM) field-level simulations. However, to be best of our
knowledge, currently there is not a solid method to evaluate the preamp
decoupling effect with circuitry and EM field analysis. In this work, we used RF
circuit and EM co-simulation4 to evaluate the preamplifier
decoupling effect and tested it on the conventional coil (LIC)1 and
the recently introduced high impedance coil (HIC)2.Methods:
Figure 1 showed the schematic
drawings of the LIC and HIC. For the LIC, we evaluated two matching strategies,
one is the parallel matching strategy with a single matching capacitor Cm, and the
other one is the series matching strategy with two series capacitors (Cms) and an
intentional small capacitor Cm (referred to as small-Cm LIC). The small Cm is
expected to improve the quality factor of the blocking circuit consisting Cm
and the preamplifier load5-7. In each case, the coil has a diameter of
78 mm, and was tuned at 128 MHz and matched to 50 Ω.
Figure 2 depicted the
procedures of the proposed co-simulation approach. First, coils were modeled in
an EM simulation (Ansys HFSS) with all lumped elements replaced with 50 Ω.
Second, values of tuning and matching capacitors were optimized by minimizing coils’
reflection coefficient (S11) in co-simulation environment (Ansys Designer).
Third, the imaginary part of the preamplifier's input impedance was optimized
to maximize the preamplifier decoupling. Similar to the practical case8,
this was obtained by minimizing the transmission coefficient (S21) of a well-decoupled
double-pick-up probe that was placed several centimeters above the coil. By
now, the preamplifier decoupling ability can be evaluated in the RF circuit
simulator. To further evaluate the SNR performance, the optimized imaginary
part and the previously set Re[Zin] were converted to equivalent parallel
resistance (Req) and parallel reactance (usually inductance, Leq) and imported
into EM simulation. Finally, individual B1- fields and electric (E-) fields were
calculated and exported to MATLAB for SNR calculation. Unlike the conventional
simulations assuming other coils opened, this simulation method will account
for additional noise correlation and SNR loss caused by residual coupling.Results and Discussions:
Figures 3 showed the preamplifier
decoupling effect of the three coils, with Re[Zin] set with 0, 2, 5 and 10
Ω. Similar to the practical case, the
preamplifier decoupling effect was evaluated as the resonate response
difference (ΔS21) between the 50 Ω termination and the preamplifier input
impedance termination8. For
Re[Zin] = 0 Ω, almost perfect decoupling can be achieved, with ΔS21 of ~58 dB.
As Re[Zin] increases, ΔS21 of LIC coil with normal Cm dropped dramatically,
with ΔS21=25/18/15 dB when Re[Zin] =2/5/10 Ω respectively. On the contrary,
both the HIC coil and the small-Cm LIC varied much less when the Re[Zin] changed.
For the HIC, the ΔS21=50/45/43 dB when Re[Zin] =2/5/10 Ω. For the small-Cm LIC
coil, the ΔS21=56/53/50 dB when Re[Zin] =2/5/10 Ω. It should be noted that choosing
extremely small Cm in the series matching circuit of the LIC will induce a
dipole mode to the loop coil, making the current distribution along conductors
less uniform than that of the typical LIC coil.
Figure 4 showed the B1-
field distortion due to non-perfect preamplifier decoupling. For each case, two
coils were placed side-by-side with 2 mm apart. The coil on the right was
driven with 1-watt power while the coil on the left was terminated with Req and
Leq. These B1- fields were also compared to the baseline B1-
fields from a single coil (without the left coil, left column in Figure 4). The
B1- results were consistent with the S21 in Figure 3. For
the typical LIC coil, the B1- distortion was highly
dependent on the Re[Zin]. For the HIC and small-Cm LIC coils, however, the
original B1- field can be maintained even with Re[Zin] =
10 Ω.
Figure 5A showed the
normalized SNR maps from three side-by-side coils. Figure 5B plotted the
average noise correlation of neighbor coils (i.e., left-middle and right-middle
coils). Figure 5C plotted normalized SNR values in an 40-mm-deep region that is
directly under the middle coil. The SNR result was also consistent with simulated
decoupling (ΔS21) and B1- results.Conclusion:
We proposed an RF
circuit and EM field co-simulation method that can evaluate the preamplifier
decoupling in terms of the resonate response difference, receive sensitivity (B1-), and array SNR. Acknowledgements
No acknowledgement found.References
1. Roemer
PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array.
Magnetic resonance in medicine. 1990 Nov;16(2):192-225.
2. Zhang
B, Sodickson DK, Cloos MA. A high-impedance detector-array glove for magnetic
resonance imaging of the hand. Nature biomedical engineering. 2018
Aug;2(8):570-7.
3. Vasanawala
SS, Stormont R, Lindsay S, Grafendorfer T, Cheng JY, Pauly JM, Scott G, Guzman
JX, Taracila V, Chirayath D, Robb F. Development and clinical implementation of
next generation very light weight and extremely flexible receiver arrays for
pediatric mri. arXiv preprint arXiv:1705.00224. 2017 Apr 29.
4. Kozlov
M, Turner R. Fast MRI coil analysis based on 3-D electromagnetic and RF circuit
co-simulation. Journal of magnetic resonance. 2009 Sep 1;200(1):147-52.
5. Zhu
H. Imris Inc. Coil decoupling for an RF coil array. United States patent US
8,138,762. 2012 Mar 20.
6. Wiggins
GC, Polimeni JR, Potthast A, Schmitt M, Alagappan V, Wald LL. 96‐Channel receive‐only head coil for 3 Tesla: design
optimization and evaluation. Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine. 2009
Sep;62(3):754-62.
7. Hergt
M, Oppelt R, Vester M, Reykowski A, Huber KM, Jahns K, Fischer H. Low noise
preamplifier with integrated cable trap. In: Proceedings of the 15th Annual
Meeting of ISMRM, Berlin, Germany, 2007 (Abstract 1037).
8. Reykowski
A, Wright SM, Porter JR. Design of matching networks for low noise
preamplifiers. Magnetic resonance in medicine. 1995 Jun;33(6):848-52.