Eddy Solomon1,2, Syed Saad Siddiq1,2, Daniel K Sodickson1,2, Hersh Chandarana1,2, and Leeor Alon1,2
1Radiology, New York University School of Medicine, New York, NY, United States, 2New York University Grossman School of Medicine, New York, NY, United States
Synopsis
MRI scans are often under continues involuntary motion which weakens their reliability and diagnostic utility for
examining the chest and abdomen. Acquisitions using traditional external
sensors (e.g. respiratory belt) and self-gated techniques tend to be highly
sensitive to patient position and setup, and on MR sequence parameters. Here,
we demonstrate the use of accelerometer sensors for detecting respiratory
signals. We show how the use of this simple sensor, with its relatively small
dimensions, high sampling rate capability, and low cost, can produce motion
corrected-images under free-breathing conditions.
INTRODUCTION
The relatively slow acquisition timing of MRI challenges its applicability
in many types of anatomical exams. Long acquisition durations are not only a
burden for patients but also can lead to motion artifacts caused by physiological
movements such as cardiac and respiratory motion (1). A common mitigation strategy in clinical routines include the use of
external sensors (2) such as respiratory belts. This technique is not always
robust and tends to be time-consuming to achieve the best motion sensitivity
signal. Alternatively, self-gated (3-4) techniques directly
extract motion from k-space, allowing tracking of motion via MR signal changes of
the complete object. However, this approach tends to depend strongly on the
slice orientation and imaging parameters (e.g. TR, number of slices). In this work, we demonstrate an alternative approach for detecting
respiratory motion using external MR-compatible accelerometer sensors. Following
our recent work on motion detection using a wireless RF transmitter (5-6), here
we explore a simple, open-source, and low-cost (<$300) option for using
wired accelerometers sensors. The accelerometer approach was tested with a radial
MR sequence and was compared to conventional methods for motion correction,
demonstrating comparable results.METHODS
Sensor setup: Sensors were controlled by a Raspberry Pi computer run on Debian
Linux and placed outside the scanner room (Fig. 1). Equipped by shielded
cables, the accelerometer (BNO-055, Adafruit Industries LLC.) and RF detector
were connected from the control room into the scanner room, through the RF
penetration panel. The accelerometer was placed on top of the abdomen of the subject
to ensure proper monitoring of the breathing motion. In this study, we have
chosen to explore only the abdomen sensor (Fig. 2, circled in red) since this
sensor was expected to have the most dominant motion signal. The RF detector
circuit includes a pick-up loop connected to an RF envelope detector that
senses the RF pulses and triggers the recording of the accelerometer sensor.
Both the RF detector and accelerometer are controlled by a dedicated python
script, enabling a sampling rate of 50ms. Once executed, the accelerometer measurements are presented in a graphical user
interface (GUI) plotting the acceleration in x-, y-, and z-directions and saving
the data to a text file. We have also explored
the use of an RF transmitter device for the detection of motion, termed ‘Pilot-Tone’. The Pilot-Tone is a small wireless device placed outside the MR
bore with no direct contact with the patient that has been shown to be
effective for motion tracking (5-6).
Imaging: Images were acquired on a 3T Prisma system (Siemens Healthcare,
Erlangen, Germany) using a body coil array. The study protocol included a
free-breathing radial stack-of-stars 3D GRE (RAVE) sequence with golden-angle
acquisition. RAVE imaging parameters included TR/TE=5.0/1.7ms, BW=500Hz/pixel.
Scans were conducted while the patient was instructed to follow different breathing
conditions: 1) Deep breathing, 2) Breath-holds. 3) Hyperventilation (shallow
and rapid breathing), 4) Normal physiological breathing. Fig. 3 (conditions 1-3) and Fig. 4 (condition
4) were acquired with 24-48 slices, 400 radial views, 1.4 mm in-plane
resolution and 3 mm slice thickness. Fig. 5 (condition 4) was acquired with 120 slices and 1.3 mm isotropic resolution.
Post-processing: After the accelerometer measurements were recorded and saved, the
signals were smoothed by a Savitzky–Golay filter and
normalized. The respiratory signal was processed by eXtra-Dimensional
(XD) reconstruction pipeline (4), which bins the continuously acquired radial data
into different respiratory states and use compressed-sensing to exploit
correlations between the motion states.RESULTS AND DISCUSSION
When comparing respiratory signals from the k-space self-navigation signal
(Fig. 3, blue), and the accelerometer (Fig. 3, red), similar patterns were seen for the deep breathing (Fig. 3a,
cond. 1), breath-holds (Fig. 3b, cond. 2), and hyperventilation (Fig. 3c, cond.
3). Interestingly, the signal recorded by the accelerometer included
small changes in breathing patterns (Fig. 3a, see white arrows) which implies
its high precision capability. However, during hyper-ventilation, signals
recorded by the accelerometer (without correction) seemed susceptible to drift artifacts
(Fig. 3c, see while arrows). Next, the MR data was sorted and binned to four
respiratory states from end-inspiration to end-expiration (Fig. 4a). Data binned by the k-space center signals
(Fig. 4b, blue) and by the accelerometer signal
(Fig. 4b, red) showed comparable
anatomical details and sharper liver tip in the end-expiration state (indicated
by yellow arrows). Moreover, a slightly more stable respiratory signal provided
by k-space center (Fig. 4a, blue) resulted in less streak artifacts, compared
to data binned by the accelerometer. In a 1.3 mm isotropic 3D view (Fig. 5), the
use of the accelerometer signal resulted in fine
tissue boundaries (indicated by green arrow) relative to the k-space center
method. Additionally, in this scan we compared our recent Pilot-Tone RF
transmitter (right column) to k-space center (left column) and accelerometer signals (middle column). Due to its high sampling rate (TR,
5ms), data binned using Pilot-Tone showed finer liver
anatomical details (indicated by yellow arrows).CONCLUSION
An MR-compatible accelerometer sensor showed reliable
results in tracking motion when compared to conventional k-space
self-navigation and Pilot-Tone. Its small dimensions, flexible high sampling
rate, and low-cost makes it a good solution for tracking breathing motion in
the MRI environment.Acknowledgements
We acknowledge support from NIH grant P41
EB0171813 and R01 5R01EB018308.References
1.
Havsteen I, Ohlhues A, Madsen KH,
Nybing JD, Christensen H, Christensen A. Are Movement Artifacts in Magnetic
Resonance Imaging a Real Problem?-A Narrative Review. Front Neurol 2017;8:232.
2.
Zaitsev M, Maclaren J, Herbst M. Motion
artifacts in MRI: A complex problem with many partial solutions. J Magn Reson
Imaging 2015;42(4):887-901.
3.
Zhang T, Cheng JY, Chen YX, Nishimura DG, Pauly
JM, Vasanawala SS. Robust Self-Navigated Body MRI Using Dense Coil Arrays.
Magnetic Resonance in Medicine 2016;76(1):197-205.
4.
Feng L, Axel L, Chandarana H, Block
KT, Sodickson DK, Otazo R. XD-GRASP: Golden-angle radial MRI with
reconstruction of extra motion-state dimensions using compressed sensing. Magn
Reson Med. 2016 Feb;75(2):775-88.
5.
Solomon E, Rigie DS, Vahle T, Paška
J, Bollenbeck J, Sodickson DK, Boada FE, Block KT, Chandarana H. Free-breathing
radial imaging using a pilot-tone radiofrequency transmitter for detection of
respiratory motion. Magn Reson Med. 2020 Dec 11. doi: 10.1002/mrm.28616.
6. Speier P, Fenchel M, Rehner R, PT-Nav: a novel respiratory navigation
method for continuous acquisitions based on modulation of a pilot tone in the
MR-receiverProc. ESMRMB 2015, 128: 97-98.