Liyuan LIANG1, Chim-Lee Cheung2, Ge Fang2, Justin Di-Lang Ho2, Chun-Jung Juan3,4,5, Hsiao-Wen Chung6, Ka-Wai Kwok2, and Hing-Chiu Chang1
1Department of Diagnostic Radiology, The University of Hong Kong, Hong Kong, Hong Kong, 2Department of Mechanical Engineering, The University of Hong Kong, Hong Kong, Hong Kong, 3Department of Medical Imaging, China Medical University Hsinchu Hospital, Hsinchu, Taiwan, 4Department of Radiology, School of Medicine, College of Medicine, China Medical University, Taichung, Taiwan, 5Department of Medical Imaging, China Medical University Hospital, Taichung, Taiwan, 6Department of Electrical Engineering, National Taiwan University, Taipei, Taiwan
Synopsis
Head
motion is a significant problem for the challenging populations, and the wireless
tracking coils has previously been proposed to enable prospective motion
correction in MRI. In this study, we evaluated the tracking performance of a novel
tiny wireless tracking marker by using a linear motion phantom, and tested the feasibility
in omnidirectional 3D head motion tracking using three tiny wireless tracking markers.
Both phantom and in-vivo results suggest that the novel tiny wireless tracking
markers can provide good fidelity in 3D position tracking, with improved subject
comfort and better flexibility in fixation of markers.
Introduction
Head
motion is a significant problem during MRI scan, and the induced image artifacts
can confound the interpretation for diagnosis. Therefore, the prospective
motion correction has been proposed to address head motion problem during data
acquisition, relying on real-time head position tracking using image-based navigators,
optical tracking devices, or NMR markers1-3. The wireless tracking markers (or semi-active
markers)2, 3 have been developed
for tracking head position using MRI projection signals, without any needs of cable
connections and additional receiver coils. Thus, wireless tracking markers are
advantageous for prospective motion correction. However, the tracking accuracy
and fidelity are limited by the size of wireless tracking marker3. Recently, a novel tiny wireless tracking
marker4 has been developed for localizing interventional
instruments within MRI. In this work, we first evaluated the tracking
performance of this novel tiny wireless tracking marker by using a linear
motion phantom, and then tested its feasibility in omnidirectional 3D head
motion tracking.Method
System setup
All
experiments were performed on a 1.5T MRI scanner (Explorer, GE Healthcare) using
an 8-channel head coil. Figure 1c shows the tiny wireless tracking marker composed
of 1) wireless multi-layer tracking coils (6.7×1.5×0.3mm3), and 2) a cylindrical tube filled
with 10mM Gd solution, which is much smaller than the design proposed by Ooi
et al.2.
Figure 2a shows the fast-tracking sequence used for acquiring three orthogonal
projection signals from wireless tracking markers. A flip angle of 1° was used to minimize the
excitation to brain tissue, and additional dephasing gradients were used to
suppress residual background signal from brain tissue. Scan parameters included:
TR=6.7ms/projection, TE=minimum,
FOV=240mm, bandwidth=20kHz, sampling points=240. The intensity linear
interpolation (ILI) method5 was employed to achieve more accurate extraction of marker
positions from the projection signals.
Phantom
experiments
A motion phantom made
of MR-compatible hydraulic motor6 was used to simulate periodic linear motion and
evaluate tracking performance, with 3 different motion ranges (±50 mm, ±30 mm
and ±10 mm) and 3 different moving speeds (high, medium, low) (Figs. 1a and 1b).
Three tiny markers were stuck on a wooden rod, and then attached on the moving plate
of motion phantom. A
bottle of saline water was also placed inside head coil to simulate
background signal. Projection data were acquired from 9 different motion
settings (3 ranges × 3
speeds) with continuous
sampling of 6 motion cycles for each setting. Discrepancies (mean ± std)
between measured movement and known movement were calculated for assessing the
accuracy of motion tracking. Standard
deviations of measured periods from each cycle were also calculated.
In-vivo
experiments
Marker
fixation
Three
wireless tracking markers were placed in plastic holders and then attached on a
homemade head strap (Figs. 1d-1f), with careful design to avoid marker overlapping
on three projection signals.
Evaluation of in-vivo tracking
precision
To
assess head tracking precision, 6000 repeated tracking scans were performed
while the volunteer remained stationary. Although the respirations might induce
slight head motion,
tracking precision were roughly estimated using the stand deviation of 6000
measured positions.
In-vivo
motion tests
The
volunteer was instructed to perform three different head motions: head shaking,
head nodding, and tracking out a “figure of eight”7 with nose. Each motion was repeated during a 40-second
scanning period. In addition, GRE and T2-FLAIR images were acquired when the
volunteer remained stationary for evaluating the influence of wireless tracking
marker to routine MRI imaging.Results
Phantom experiments: Figure 3 shows the
measured traces for 3 markers with ±30mm movement range at medium moving speed,
and calculated discrepancies and periods for different motion settings. The mean
and standard derivation of medium discrepancy were 0.1586mm and 0.0618mm,
respectively.
In-vivo experiments: Figure 2 shows successful
suppression of background signal from brain tissue using dephasing gradients. Precision
of motion tracking for the markers along LR, SI and AP directions were
0.1259mm(pixel), 0.0962mm(pixel), 0.0899mm(pixel), respectively. Figure 4a.
shows measured traces for the motion of head shaking. Three markers positions
at two selected time points were shown in Figure 4b. Traces of the other two
motions were measured successfully with no marker overlapping. This can be
explained by the small size of markers and improved flexibility of marker
placements with the used of head strap. Figure 5 shows the routine MRI images
without any influence from the markers (red arrows).Discussion and Conclusion
Both
phantom and in-vivo experiment results show good fidelity in motion tracking
using the proposed tiny wireless tracking markers. The measured tracking
accuracy and precision were satisfactory because the small size of marker can
produce sharper peak signal for improving tracking accuracy. Localization of
peak signal from the projection signal highly relies on the suppression of
background signal using dephasing gradients. It is because dephasing gradients
can produce substantial phase shift within a large volume while still keeping
signal from small structure8. Therefore, the proposed tiny wireless tracking markers is
less affected by the dephasing gradient and can provide better marker-to-background
signal ratio. In conclusion, the proposed tiny wireless marker can provide good
fidelity in omnidirectional 3D position tracking for prospective motion
correction, with improved subject comfort and better flexibility in fixation of
markers.Acknowledgements
The work was in
part supported by grants from Hong Kong Research Grant Council (GRFs
HKU17121517 and HKU17106820) and Hong Kong Innovation and Technology Commission
(ITS/403/18).References
1. M.
B. Ooi, S. Krueger, W. J. Thomas, S. V Swaminathan, and T. R. Brown, “Prospective
real‐time correction for arbitrary head motion using active markers,” Magn.
Reson. Med. An Off. J. Int. Soc. Magn. Reson. Med., vol. 62, no. 4, pp. 943–954,
2009.
2. M.
B. Ooi, M. Aksoy, J. Maclaren, R. D. Watkins, and R. Bammer, “Prospective
motion correction using inductively coupled wireless RF coils,” Magn. Reson.
Med., vol. 70, no. 3, pp. 639–647, 2013.
3. S.
Sengupta, S. Tadanki, J. C. Gore, and E. B. Welch, “Prospective real‐time head
motion correction using inductively coupled wireless NMR probes,” Magn.
Reson. Med., vol. 72, no. 4, pp. 971–985, 2014.
4. C.-L.
Cheung, J. D.-L. Ho, V. Vardhanabhuti, H.-C. Chang, and K.-W. Kwok, “Design and
Fabrication of Wireless Multilayer Tracking Marker for Intraoperative
MRI-Guided Interventions,” IEEE/ASME Trans. Mechatronics, vol. 25, no.
2, pp. 1016–1025, 2020.
5. M.
Rea, D. McRobbie, H. Elhawary, T. H. Zion, M. Lamperth, and I. Young, “Sub-pixel
localisation of passive micro-coil fiducial markers in interventional MRI,” Magn.
Reson. Mater. Physics, Biol. Med., vol. 22, no. 2, pp. 71–76, 2009.
6. Z.
Dong et al., “High-performance continuous hydraulic motor for MR safe
robotic teleoperation,” IEEE Robot. Autom. Lett., vol. 4, no. 2, pp.
1964–1971, 2019.
7. M.
Herbst, J. Maclaren, M. Weigel, J. Korvink, J. Hennig, and M. Zaitsev, “Prospective
motion correction with continuous gradient updates in diffusion weighted
imaging,” Magn. Reson. Med., vol. 67, no. 2, pp. 326–338, 2012.
8. C.
L. Dumoulin, R. P. Mallozzi, R. D. Darrow, and E. J. Schmidt, “Phase‐field
dithering for active catheter tracking,” Magn. Reson. Med., vol. 63, no.
5, pp. 1398–1403, 2010.