Mario Bacher1,2, Barbara Dornberger2, Jan Bollenbeck2, Matthias Stuber1, and Peter Speier2
1Department of Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 2Siemens Healthcare Magnetic Resonance, Erlangen, Germany
Synopsis
The Pilot Tone navigator is a novel electromagnetic motion sensing method capable of contactless respiratory and cardiac motion sensing. We demonstrate the physical processes underlying the Pilot Tone navigator using electromagnetic simulations on a realistic virtual human phantom and validate these simulation results in-vivo. These simulations can be used to investigate
how the PT signal is shaped by motion of the underlying anatomy with the aim of
using this information to aid in improving our processing pipeline.
Introduction
Pilot Tone (PT) is a novel electromagnetic motion sensing
method1 that enables contactless sensing of respiratory and cardiac motion in
an MR environment. It is based on inductive coupling between a PT signal
generator coil and the MR scanner’s receive coils via conductive tissues. The generator
loop, driven at a frequency close to the Larmor frequency, forms a magnetic
field $$$\mathbf{B}_{PT}$$$ which induces eddy currents in conductive tissues.
These eddy currents generate a secondary magnetic field $$$\mathbf{B}_{sec}$$$ and the sum $$$\mathbf{B}_{PT}+\mathbf{B}_{sec}$$$ is measured using the scanner’s receive coils.
Motion of these tissues then modulates the received signal. Each coil measures
a signal modulated by a combination of respiratory and cardiac motion and these
components can be separated by using independent component analysis (ICA2).
The resulting signals can be used in a variety of applications instead of ECG
or MR navigators3-7. We performed a simulation study aimed at understanding
how Pilot Tone works at a fundamental level and leveraging this knowledge to improve
our processing pipeline. We were able to corroborate our findings in-vivo and
present here first results aimed at improving the extraction of motion signals
from PT raw data: One of the remaining challenges in fully automating the
separation of cardiac and respiratory components has been the indeterminacy of
the extracted motion signal’s sign after performing ICA. In addition to that,
we have observed that the signs of motion signals differ between coils. Using
our simulation results, we can now describe and predict the relation of the
received signal’s sign to the underlying geometry.Methods
We simulated a realistic human torso (XCat v2 phantom8),
comprised of 30 different tissue types with their respective values for
conductivity and relative permittivity9 at
$$$f_{PT}=122.5\,MHz$$$ (simulating a $$$3\,T$$$ system) using the Sim4Life FDTD
(finite differences in the time-domain) solver (ZMT Zurich MedTech AG, Zurich,
Switzerland). The virtual phantom (Fig.1) was set to simulate a cardiac cycle
at 11 equally spaced timepoints. The PT generator coil was positioned anterior
to the heart of this phantom and driven with $$$i(t)=9\,mA*\sin(2 \pi f_{PT}t)$$$.
Simulation was performed using Sim4Life’s CUDA FDTD solver (Quadro M4000,
NVIDIA Corporation). The magnetic fields $$$\mathbf{B}(x,y,z,t_n)$$$ at each timepoint $$$t_n,n\in[1,11]$$$ were
extracted at an anterior plane $$$\Sigma_{ant}$$$ (where receive coils
would be positioned in-vivo, see Fig.1).
Receive coils were simulated by integrating the
magnetic field component normal to $$$\Sigma_{ant}$$$.
The resulting signal,
$$u_{t_n,ROI}\propto\int_{ROI}B_{y,t_n}dA_{ROI}$$
proportional to the magnetic flux through the
coil, is then evaluated along its $$$t_n$$$ direction, yielding the simulated
time-signal $$$u[t_n]$$$, shown for two virtual coils in Fig.2. To determine
whether the magnetic field time-signal at a point in $$$\Sigma_{ant}$$$
contains modulation due to cardiac motion, we determined the correlation
coefficient at each pixel of $$$\Sigma_{ant}$$$ to cardiac volume ground
truth, yielding the correlation map of $$$B_y[t_n]$$$ to $$$V_{card}$$$
in $$$\Sigma_{ant}$$$ (Fig.3, left). We then scanned a volunteer similar
in size to the virtual phantom at 3T using a 12-channel array-coil with
integrated PT generator (MAGNETOM Vida & Biomatrix Body12, Siemens
Healthcare, Erlangen, Germany) and compared the sign of the cardiac signal as
determined by temporal relation to ECG R-peak (see Fig.2a), in each receive
element to simulation results (Fig.3, right).Results
Figure 2 shows the simulated cardiac signal modulation in
percent of the average received signal in $$$L_{R1}$$$ (b) and $$$L_{R2}$$$ (c) compared to cardiac
volume ground truth (a). Note how the sign of the modulation signal differs in
these two coils. We found absolute modulation depths in $$$\Sigma_{ant}$$$ ranging
from 0 to 0.1 percent. Figure 3a shows a pixelwise correlation map for $$$\Sigma_{ant}$$$.
Changes in the sign of the received signal can be seen along a well-defined
boundary surrounding the heart and extending into the abdomen (P<0.001
except on sign boundary). In our volunteer (Fig.3b), the sign of the received
signal follows a similar pattern: negatively correlated channels (like Fig.2c) over the heart, and positively correlated channels in the periphery.Discussion
Our simulations show qualitatively a similar signal behavior
as that observed in-vivo, with channels close to the heart having negative
correlation to cardiac volume, while more distant elements are positively
correlated. Quantitatively, the simulated modulation depth close to the
generator is lower than that typically observed in-vivo (0-0.1% vs. 0-1%) and
approaches in-vivo results in the periphery. We believe that this is due to
additional capacitive effects, i.e. capacitive coupling between generator,
tissue and receive coils, and suppression of the primary field by geometric
decoupling not considered in our simulation. The sign distribution in the
correlation maps agree well with in-vivo results and knowledge of the location-dependent
signal behavior could aid in the aforementioned problem of automatically
determining the sign of motion components. The utilization of existing receive
hardware makes PT a very attractive technique for obtaining physiological
motion signals, but this also means that the placement of receive coils is
constrained by requirements of the actual MR imaging. It is therefore
imperative to understand how the PT signal behaves at these positions and
electromagnetic simulations offer a useful tool for further research in this
direction. The XCat phantom can simulate a wide variety of anatomies and future
research could be focused on investigating how anatomical variation impacts
signal behavior.Acknowledgements
No acknowledgement found.References
1] P. Speier, M. Fenchel, and R. Rehner,
“PT-Nav: a novel respiratory navigation method for continuous acquisitions
based on modulation of a pilot tone in the MR-receiver,” in Magnetic
Resonance Materials in Physics, Biology and Medicine, 2015, vol. 28, no. Book
of Abstracts, pp. 97–98, doi: 10.1007/s10334-015-0487-2.
[2] A. Hyvärinen, J. Karhunen, and E. Oja,
“Independent Component Analysis,” Appl. Comput. Harmon. Anal., vol. 21,
no. 1, pp. 135–144, 2001, doi: 10.1002/0471221317.
[3] M. Bacher, P. Speier, J. Bollenbeck, M.
Fenchel, and M. Stuber, “Pilot Tone Navigation Enables Contactless Prospective
Cardiac Triggering : Initial Volunteer Results for Prospective Cine,” in Joint
Annual Meeting of the ISMRM & ESMRMB, Paris, 2018, p. 4798.
[4] M. Bacher, P. Speier, J. Bollenbeck, M.
Fenchel, and R. Stollberger, “Retrospective Evaluation of Pilot Tone Based
Cardiac Trigger Quality In A Volunteer Cohort,” in Magnetic Resonance
Materials in Physics, Biology and Medicine, 2017, vol. 30, no. Book of
Abstracts, pp. 360–361, doi: 10.1007/s10334-017-0634-z.
[5] M. Bacher et al., “Performance
Analysis of a Fully Automatic Real-Time Capable Pilot Tone Cardiac Triggering
Framework,” Proc. from 23rd Annu. SCMR Sci. Sess., vol. 63, pp. 412–414,
2020.
[6] T. Vahle et al., “Respiratory
Motion Detection and Correction for MR Using the Pilot Tone,” Invest.
Radiol., vol. 55, no. 3, pp. 153–159, Mar. 2020, doi:
10.1097/RLI.0000000000000619.
[7] M. B. L. Falcão et al., “5D Flow
– A quantitative in vivo comparison between Self Gating and Pilot Tone Gating,”
vol. 74, no. 5, pp. 2019–2021, 2019.
[8] W. P. Segars, G. Sturgeon, S. Mendonca,
J. Grimes, and B. M. W. Tsui, “4D XCAT phantom for multimodality imaging
research,” Med. Phys., vol. 37, no. 9, pp. 4902–4915, 2010, doi:
10.1118/1.3480985.
[9] P. Hasgall et al., “IT’IS
Database for thermal and electromagnetic parameters of biological tissues,
Version 4.0, May 15, 2018.” [Online]. Available: itis.swiss/database.