Yuji Iwadate1, Atsushi Nozaki1, Shigeo Okuda2, Tetsuya Wakayama1, and Masahiro Jinzaki2
1Global MR Applications and Workflow, GE Healthcare Japan, Hino, Japan, 2Department of Radiology, Keio University School of Medicine, Tokyo, Japan
Synopsis
We propose a deep learning-based respiratory navigator (DLnav)
technique which uses a convolutional neuronal network (CNN) for respiratory motion
detection. The pencil-beam navigator signals were transferred to the real time
processing unit including a CNN module and a diaphragm position value was
calculated there. DLnav was incorporated into prospectively navigator-gated 3D
SPGR and its performance was evaluated in the volunteer scan. DLnav resulted in
good synchronization with actual respiratory motion and reduced motion-induced
blurring with two different tracker positions.
Introduction
The respiratory navigator echo technique is widely
used for free-breathing cardiac and abdominal MRI with minimal respiratory
motion artifacts. Navigator echo enables direct respiratory motion detection of
the diaphragm and can be used for more accurate respiratory triggering/gating
than other techniques with extra devices such as bellows. However, respiratory
motion detection with navigator can be degraded due to various causes (e.g.
inappropriate navigator tracker positioning), resulting in severe motion
artifacts. In this study, we propose a deep learning-based respiratory
navigator (DLnav) technique which uses a convolutional neuronal network (CNN)
for respiratory motion detection.Methods
Respiratory Gating with Deep Learning-Based Navigator Echo: The pulse sequence was based on the previously
reported navigator-gated 3D SPGR sequence1 (Figure 1a). As shown in Figure 1, a navigator
pulse sequence followed an imaging block for acquisition of pencil-beam
navigator echo signals. The multi-channel navigator signal data was then
transferred to the real time processing unit. The navigator data was used as an
input to the CNN module, which includes 4-layer CNN regression model
pre-trained with 1115 navigator data sets. The CNN module output a single
diaphragm position value (one of the respiratory waveform points in Figure 1), and
the value was passed to the pulse sequence and used for imaging data
acceptation/rejection in real time (prospective respiratory gating). The
navigator processing was performed in 24 ms including the navigator pulse
sequence and diaphragm position value calculation with the CNN module.
Data Acquisition: We performed experiments on a GE 3 T Pioneer imaging
system using floating anterior and embedded posterior coil arrays. The navigator-gated
3D SPGR scans were performed with a healthy volunteer using a 20-mm diameter
pencil-beam navigator. 3D data were
acquired in the axial orientation with imaging parameters: ARC acceleration
factor = 2 × 1, TR/TE = 3.2/1.5 ms, FOV = 40 × 40 cm2, matrix = 256
× 192, slice thickness = 4.4 mm, receiver bandwidth = ±83.3 kHz, flip angle = 10°.
The navigator-gated 3D SPGR scan
was performed twice, one with the normal tracker positioning (Figure 2a) and the
other with the tracker position intentionally shifted onto the heart (Figure
2b). Non-navigator-gated 3D SPGR scan was also conducted for comparison
purpose. Approximate scan time was 40 s and 12 s for navigator-gated and
non-gated 3D SPGR, respectively.
Data Analysis: For respiratory waveform evaluation, a
signal-to-noise ratio of respiratory-like frequencies2 (SNRR)
was measured. 10-s respiratory waveform data were Fourier transformed in the
time domain. The maximum value of the power spectrum within the frequency range
for respiration (0.1–0.4 Hz) was divided by the mean value for the frequency
range of 0.4–3.0 Hz. This evaluation was also performed for the
conventional navigator processing method using the same navigator data sets,
and the least-squares error method3 was chosen as the conventional
one. For image evaluation, sharpness4 of the left portal vein was
quantitatively calculated. Edge
values were calculated for both sides of the vessel using a first-order
derivative, and final sharpness was defined as the average score of both sides
divided by the center of vessel signal intensity.Results
Figure 3 shows the waveforms detected with the
conventional and DLnav methods and their SNRR values. The waveforms
were well synchronized with the actual respiratory motion and SNRR was
higher than 20 for both methods when the tracker was positioned normally. The
conventional method’s waveform was degraded (SNRR = 2.2) when the
tracker was on the heart due to undesired signals in the superior area, but
DLnav maintained a good synchronization with the respiratory motion (SNRR
= 12.8). Figure 4 shows 3D-SPGR images. DLnav images reduced motion artifacts
and improved the sharpness. Quantitative sharpness values of the portal vein
shown in Figure 4 were 0.42, 0.30 and 0.18 for DLnav with normal tracker, DLnav
with tracker on the heart and non-gating, respectively.Discussion
We have developed the DLnav technique and have shown
that navigator-gated 3D SPGR with DLnav is feasible. The respiratory waveform
had a high SNRR even when the tracker was positioned badly on the
heart. Though DLnav images had higher sharpness values for two different
tracker potions than the non-gated image, comparison between different tracker
positions shows that the tracker on the heart resulted in blurring even with
DLnav when compared to the normal tracker position. However, the motion
detection and resultant 3D SPGR image quality can be
improved further with increased number of training navigator signal data sets of
various scan conditions. Clinical evaluation with a large number of
subjects is required in future.Conclusion
The DLnav technique is feasible and improved
robustness of free-breathing abdominal MRI in the volunteer scan.Acknowledgements
No acknowledgement found.References
1. Vasanawala SS, Iwadate Y, Church
DG, et al. Navigated abdominal T1-W MRI permits
free-breathing image acquisition with less motion artifact. Pediatr Radiol.
2010;40:340–4.
2. Walker
MD, Morgan AJ, Bradley KM, et al. Evaluation of data-driven respiratory gating
waveforms for clinical PET imaging. EJNMMI Res. 2019;9:1.
3. Wang Y, Grimm RC, Felmlee JP,
et al. Algorithms for extracting motion information from navigator echoes. Magn
Reson Med 1996;36:117–123.
4. Botnar
RM, Stuber M, Danias PG
et al. Improved coronary artery definition with T2-weighted, free-breathing,
three-dimensional coronary MRA. Circulation. 1999;99:3139–3148.