John Neri1, Matthew F Koff1, Kevin M Koch2, and Ek Tan1
1Radiology and Imaging, Hospital for Special Surgery, New York, NY, United States, 2Medical College of Wisconsin, Wauwatosa, WI, United States
Synopsis
Diffusion-weighted MAVRIC is a novel pulse sequence that can obtain
quantitative diffusion values in regions with high susceptibility. The goal of
this work is to compare the quantitative accuracy of DWI-MAVRIC with sequences
currently available on clinical MR scanners, using a diffusion phantom and
images acquired at different orientation and spatial offsets in the scanner.
Overall, good linearity with diffusivity was observed, with a large positive
ADC bias at low diffusivity in both axial and coronal DWI-MAVRIC noted. Right
and left spatial offsets increased ADC errors that can be mitigated by gradient
nonlinearity correction.
Introduction
Diffusion weighted imaging (DWI) provides quantitative
measurement of random water displacement(1), as calculated
in an apparent diffusion coefficient (ADC) map. DWI is a promising tool for
evaluating musculoskeletal tissues(2–4)
but has had limited effectiveness in peri-prosthetic regions due to strong susceptibility
artifacts from metallic components frequently found in orthopedic hardware(5). One solution is to utilize multispectral
imaging (MSI) techniques, such as multi-acquisition with variable resonance
image combination (MAVRIC)(6), which
is an artifact-reducing technique that sums spectral bins to form a composite
image containing less metal artifact than 3D-FSE acquisitions alone. Diffusion-weighted
MAVRIC(7) is a
recently-developed modification of the 2D PROPELLER DUO diffusion-weighted pulse
sequence(8), which
acquires multiple spectral bins and DUO encoded PROPELLER blades, allowing quantitative diffusion values to be
obtained with low geometric distortion and improved signal in regions with high
susceptibility. Since diffusivity is sensitive to
alterations to tissue microstructure caused by disease(9), DWI-MAVRIC
may provide a novel means for evaluating peri-prosthetic joint infection and
avascular necrosis among other orthopedic conditions,. The
goal of this work is to compare the quantitative accuracy of DWI-MAVRIC with
standard EPI-readout-based DWI and diffusion tensor imaging (DTI) pulse
sequences currently available on clinical MR scanners. This analysis is important, as there are many
sources of quantitative bias such as spatial nonlinearity, imaging gradients, image
distortion, and eddy currents.Methods
Phantom scanning was performed using a Diffusion Standard Model 128 (CaliberMRI,
Boulder, CO). This phantom contains 13 vials of varying concentrations of the
diffusivity-modifier polyvinylpyrrolidone(PVP) (three of:
0% wt/wt, two of: 10%,20%,30%,40%,50%) with known
diffusivity values at 0oC(10).
All images were acquired on a 1.5T MRI scanner with an 8-channel cardiac coil (GE
Healthcare, Chicago, IL). The phantom was filled with crushed ice and stored in
a refrigerator overnight to establish steady-state temperature during apparent
diffusion coefficient (ADC) measurement. The phantom was first securely
positioned at iso-center. A total of 8 sequences were
acquired (six coronal, two axial), including: coronal conventional DWI-EPI (3
diffusion directions, repeated with 1 and 3 averages (NEX)), coronal DTI-EPI
(30 directions), coronal DWI-MAVRIC
(repeated with 3, 2, and 1 bins), axial DWI-EPI (NEX=3) and axial MAVRIC-DWI (3
bins) using a b-value of 1000 s/mm2. The phantom was rotated 90o
between the coronal and axial sequences to ensure the vial contents were
in-plane for all scans. Scans commenced 5 minutes after repositioning to reduce
residual flow effects. To evaluate effects of gradient
nonlinearity and off-isocenter positioning, the 8 sequences were then repeated
at two nominal spatial offsets of 10 cm left and right off iso-center. Left
and right positions were assigned from the perspective of a supine patient entering
the scanner headfirst. Images were processed with software (MATLAB, Mathworks, Natick, MA)
to generate ADC Maps with and without gradient nonlinearity correction (GNC)(11). Circular
region-of-interests of ~208 mm2
(coronal) and 102 mm2 (axial) were placed on the ADC maps using
ITK-SNAP(12) on
each vial for analysis.
Results
Fig. 1 compares ADC maps from DWI-EPI and DWI-MAVRIC,
showing less geometric distortion in the MAVRIC maps. Fig.
2 compares the mean ADC at all PVP concentrations against the reference values(10). Excellent (negative) correlation between
ADC and concentration was obtained with all EPI (r=-0.987 to -0.991) and MAVRIC
scans (r=-0.978 to -0.992). However, we observed a pattern of bias at the high
PVP vials (40-50%) with low diffusivities in the DWI-MAVRIC scans (Fig. 3). The
bias was analyzed more closely in the error plot of Fig.4, which showed the
lowest error in axial DWI-EPI, a small positive bias in the 50% vials with
coronal DWI-EPI, a large positive bias in the 50% vials with axial DWI-MAVRIC,
and large biases in both 40% and 50% vials in coronal DWI-MAVRIC. Fig. 5 shows
the effects of positioning, whereby the lowest ADC errors were found at
isocenter, as compared to the left and right positions. Gradient nonlinearity
correction reduced the absolute error in all eight sequences by 0.31 %,
3.63%,0.65% for the center, left, and right positions respectively.Discussion
Validating diffusion MAVRIC as a clinical imaging tool
requires evaluating the differences between measured ADC values and ground
truth. The negative ADC bias associated
with lower PVP is likely associated to noise and can be addressed using various
denoising techniques. The positive ADC bias due to high PVP was attributed to
suboptimal b-value, which can be mitigated by a multi b-value acquisition. Finally,
the large positive bias identified in data from MAVRIC images remains a point
of investigation but is suspected to be caused by a
combination of residual CPMG violations and reconstruction algorithm
vulnerabilities of the PROPELLER DUO approach. Acknowledgements
Research reported in this abstract was supported by
NIH/NIAMS under award number R01AR064848. The content is solely the
responsibility of the authors and does not necessarily represent the official
views of the NIH.References
1. Bihan DL,
Turner R, MacFall JR. Effects of intravoxel incoherent motions (IVIM) in
steady-state free precession (SSFP) imaging: application to molecular diffusion
imaging. Magnetic Resonance in Medicine 1989;10:324 337.
2. Raya JG,
Melkus G, Adam-Neumair S, et al. Change of diffusion tensor imaging parameters
in articular cartilage with progressive proteoglycan extraction. Invest Radiol
2011;46:401–9 doi: 10.1097/rli.0b013e3182145aa8.
3. Hooijmans MT,
Damon BM, Froeling M, et al. Evaluation of skeletal muscle DTI in patients with
duchenne muscular dystrophy. Nmr Biomed 2015;28:1589–1597 doi:
10.1002/nbm.3427.
4. Wengler K,
Tank D, Fukuda T, et al. Diffusion tensor imaging of human Achilles tendon by
stimulated echo readout‐segmented EPI (ste‐RS‐EPI). Magnet Reson Med
2018;80:2464–2474 doi: 10.1002/mrm.27220.
5. Farahani K,
Sinha U, Sinha S, Chiu LC-L, Lufkin RB. Effect of field strength on
susceptibility artifacts in magnetic resonance imaging. Comput Med Imag Grap
1990;14:409–413 doi: 10.1016/0895-6111(90)90040-i.
6. Koch KM,
Lorbiecki JE, Hinks RS, King KF. A multispectral three‐dimensional acquisition
technique for imaging near metal implants. Magnet Reson Med 2009;61:381–390
doi: 10.1002/mrm.21856.
7. Koch KM,
Bhave S, Gaddipati A, et al. Multispectral diffusion‐weighted imaging near
metal implants. Magnet Reson Med 2018;79:987–993 doi:
10.1002/mrm.26737.
8. Pipe JG,
Farthing VG, Forbes KP. Multishot diffusion‐weighted FSE using PROPELLER MRI.
Magnet Reson Med 2002;47:42–52 doi: 10.1002/mrm.10014.
9. Razek AA,
Nada N. Role of diffusion-weighted MRI in differentiation of masticator space
malignancy from infection. Dentomaxillofac Rad 2013;42:20120183 doi:
10.1259/dmfr.20120183.
10. Palacios EM,
Martin AJ, Boss MA, et al. Toward Precision and Reproducibility of Diffusion
Tensor Imaging: A Multicenter Diffusion Phantom and Traveling Volunteer Study.
Am J Neuroradiol 2017;38:537 545 doi: 10.3174/ajnr.a5025.
11. Tan ET,
Marinelli L, Slavens ZW, King KF, Hardy CJ. Improved correction for gradient
nonlinearity effects in diffusion-weighted imaging. J Magn Reson Imaging
2012;38:448–53 doi: 10.1002/jmri.23942.
12. Yushkevich
PA, Piven J, Hazlett HC, et al. User-guided 3D active contour segmentation of
anatomical structures: Significantly improved efficiency and reliability.
Neuroimage 2006;31:1116–1128 doi: 10.1016/j.neuroimage.2006.01.015.