Nora-Josefin Breutigam1, Daniel Christopher Hoinkiss1, Mareike Alicja Buck 1,2, Klaus Eickel1,2, Matthias Günther1,2, and David Porter3
1Imaging Physics, Fraunhofer MEVIS, Bremen, Germany, 2Faculty 01 (Physics/Electrical Engineering), University Bremen, Bremen, Germany, 3Imaging Centre of Excellence (ICE), University of Glasgow, Glasgow, Scotland
Synopsis
Simultaneous
multi-contrast imaging (SMC) can be used to combine acquisition of
diffusion-weighted (DW) and T2*-weighted (T2*W) images into a single scan.
Compared to conventional single-contrast imaging, SMC reduces the total scan
time and improves image registration. However, saturation effects can reduce
SNR and alter contrast. In this study, these effects are investigated in
simulations, in phantoms, and in vivo. By using the results of this study to
control saturation effects in SMC, the method enables rapid acquisition of
distortion-matched, high-quality, well-registered DW and T2*W imaging, which
could support rapid diagnosis and treatment of acute stroke.
Introduction
Simultaneous
multi-contrast (SMC) imaging1 can be used to combine the acquisition
of diffusion-weighted (DW) and T2*-weighted (T2*W) images
into a single scan. This contrast combination is of particular interest for the
diagnosis of acute stroke, where T2*W images can be used to detect
bleeding2-4 and DW scans are sensitive to acute infarcts in brain
tissue5. Compared to conventional single-contrast imaging, the overall
scan time with SMC is reduced, and image registration is improved, but saturation
effects can reduce SNR and modify contrast. In this study, these effects are investigated
in simulations, in phantoms, and in vivo.Methods
The
modified readout-segmented EPI sequence for SMC acquires DW and T2*W contrast from two slice
positions
at the same
time. The signals are separated using an in-house reconstruction at the
scanner based on the split slice-GRAPPA algorithm6,7.
Computer
Simulations:
The mutual
saturation effects of the two slices were simulated for a range of excitation
flip angles for the T2*W slice. To allow comparison with experimental data, T1
and B1 values determined from phantom scans and images in vivo were used for
the simulations.
Data
Acquisition:
Data were
acquired at 3T (MAGNETOM Skyra, Siemens Healthineers AG) using a standard 20-channel head coil.
For
saturation comparison with computer simulations, measurements were performed on
a spherical water-based phantom and a healthy subject at two slice positions in
each case. One slice was positioned outside the phantom or subject to prevent
the effects of potential cross-talk between slices. Acquisitions were performed
with both SMC and the corresponding single-contrast sequences. T1
maps were generated using a gradient-echo, slice-selective, inversion-recovery
sequence. B1 field maps were determined with a 2D EPI sequence using
the double-angle method (DAM) (α = 60°, 2α = 120°).
For the
second subject, a potential clinical protocol was used to acquire
trace-weighted and T2*W images simultaneously with the following parameters:
TR/TEDW/TET2*W= 4500/81.5/24.8 ms, voxel size 1.0x1.0x3.0mm3,
in-plane acceleration factor of 2, and acquisition time 3:09 min. For the DW scans, there
were two b-values (0 and 1000 s/mm2). For the T2*W
contrast the flip angle was 20°.
Data
Analysis:
The
saturation differences between scans with SMC and those using the equivalent
single-contrast sequences were compared. Data from the first healthy subject
were used to evaluate SMC saturation effects on grey and white matter separately.
Non-synchronous
in vivo scans were registered with an in-house registration library in MeVisLab8
with a 2D rigid transformation.Results
Figure 1
compares the measured and simulated saturation for DW and T2*W
contrast in the phantom with corresponding T1 and B1-field
maps. A detailed analysis of the measured and simulated saturation for both
contrast types in the phantom as a function of the T2*W excitation
flip angle is shown in figure 2. Experimental data are shown both with and
without the use of split-slice GRAPPA during image reconstruction. The image
data in figure 3 demonstrates the saturation effects seen with SMC in grey and
white matter compared to the single-contrast acquisitions. The figure also compares
these results with simulated values using measured T1 and B1-field
maps. Figure 4 shows the quantitative analysis for saturation effects in vivo.
Finally, figure 5 shows the results from a high-resolution, trace-weighted
acquisition with and without SMC.Discussion & Conclusion
SMC imaging
can provide inherently registered ADC maps and T2*W images in around
three minutes. There are however additional saturation effects, which have been
investigated in this study.
The signal
changes introduced by the SMC method depend on a number of factors, including
the time between DW and T2*W excitation of the same slice, the
repetition time (TR), and the flip angle of the T2*W excitation. It
has been demonstrated that the slice-GRAPPA reconstruction allows good
separation of the SMC data (fig. 5) and does not have a major influence on
signal level compared to a standard reconstruction (fig. 2). Simulated data predict
higher saturation effects than the measured case. One potential reason for this
is that slice-profile effects were not considered during the simulation. Despite
image registration, the results from the measured data may also have been
affected by an uncorrected frequency drift during data acquisition, and motion
during the scans in vivo. Frequency drift can be corrected in future work.
As shown in
figure 4 and figure 5, an appropriate choice of T2*W excitation
pulse minimizes SNR loss and contrast changes in the DW images. SNR loss in the
T2*W case is higher, but this is more than offset by four-fold averaging
of T2*W data during the acquisition of data for multiple diffusion-gradient
directions and b-values. The main effect on the T2*W images is the
change in contrast due to suppression of CSF signals. However, this is unlikely
to affect the diagnostic capability of the sequence to detect hemorrhage.
By using
the results from this study to control saturation effects in SMC, the method
allows the rapid acquisition of distortion-matched, high-quality,
well-registered DW and T2*W images, which could support the prompt
diagnosis and treatment of acute stroke. In future work, SMC could be combined
with SMS for reduced scan time or increased slice coverage. The multiple T2*W
images during the acquisition might also be useful for motion-correction
purposes.
Acknowledgements
No acknowledgement found.References
1. Breutigam
NJ, Frost R, Eickel K, Porter D. Simultaneous Multi-Contrast Imaging with
Readout-Segmented EPI. Proceedings of the 25th Annual Meeting of ISMRM. 0520. 2017.
2. M.
Köhrmann, P.D. Schellinger. Acute stroke triage to intravenous thrombolysis and
other therapies with advanced CT or MR imaging: pro MR imaging. Radiology, 2009;
251, pp. 627-633, 10.1148/radiol.2513081074
3. C.S.
Kidwell, J.A. Chalela, J.L. Saver, S. Starkman, M.D. Hill, A.M. Demchuk, et al.
Comparison of MRI and CT for detection of acute intracerebral hemorrhage. JAMA
2004;292, pp. 1823-1830, 10.1001/jama.292.15.1823
4. J.B.
Fiebach. Stroke magnetic resonance imaging is accurate in hyperacute
intracerebral hemorrhage: a multicenter study on the validity of stroke imaging.
Stroke 2004; 35, pp. 502-506, 10.1161/01.STR.0000114203.75678.88
5. P.D.
Schellinger, R.N. Bryan, L.R. Caplan, J.A. Detre, R.R. Edelman, C. Jaigobin, et
al. Evidence-based guideline: the role of diffusion and perfusion MRI for the
diagnosis of acute ischemic stroke: report of the Therapeutics and Technology
Assessment Subcommittee of the American Academy of Neurology. Neurology 2010; 75,
pp. 177-185, 10.1212/WNL.0b013e3181e7c9dd
6.
Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled
aliasing in parallel imaging for simultaneous multislice echo planar imaging
with reduced g-factor penalty. Magn Reson Med 2012;67:1210-1224.
7. Cauley
SF, Polimeni JR, Bhat H, Wald LL, Setsompop K. Interslice leakage artifact
reduction technique for simultaneous multislice acquisitions. 2014;72(1):93-102.
8. Ritter F, Boskamp T, Homeyer A, Laue H, Schwier M, Link
F,
Peitgen H-O. Medical
image analysis. IEEE Pulse 2011;2:60–70.