Philip Kenneth Lee1,2, Yuxin Hu1,2, Catherine Judith Moran2, Bruce Lewis Daniel2, and Brian Andrew Hargreaves1,2,3
1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States, 3Biomedical Engineering, Stanford University, Stanford, CA, United States
Synopsis
In diffusion
weighted imaging, multi-shot Echo Planar Imaging (EPI) is preferred over single-shot
EPI due to reduced geometric distortion. Recent work has shown that low-rank
reconstructions can correct ghosts from shot-to-shot phase without explicitly
acquiring a phase navigator. These works have been limited to EPI sampling
trajectories with uniform ky sampling. 2D Cartesian Fast Spin Echo (FSE) is a
distortionless alternative to multi-shot EPI that has greater freedom in
k-space traversal and reduced chemical shift artifacts. Using FSE, we
demonstrate in simulation and in vivo that an intelligent choice of sampling
pattern greatly enhances image quality in multi-shot diffusion imaging.
Introduction
Multi-shot Echo
Planar Imaging (EPI) is often applied to reduce off-resonance induced geometric
distortion in diffusion weighted (DW) imaging. Ghosts from shot-to-shot phase caused
by motion sensitizing diffusion gradients can be resolved with phase correction
prior to shot combination1,2, or without explicit phase estimation
using low-rank regularizers in k-space3 or image domain4.
The application of low-rank regularizers for reconstructing multi-shot DW data has
been limited to 2D EPI data, which must be uniformly sampled to maintain
constant ky velocity.
In this work, we show
that the correct choice of Cartesian sampling trajectory can greatly enhance
image quality in an image domain locally low-rank reconstruction4,5.
Data is acquired with a distortionless, diffusion prepped, multi-shot 2D Fast
Spin Echo (FSE) sequence. Multi-shot FSE permits modification of sampling patterns
in ky-shot space and reduces T2 blurring effects compared to single-shot FSE.
First, we tested a random
sampling pattern that creates incoherent aliasing in y-shot space. Compressed
sensing reconstructions exploiting sparsity in certain dimensions are known to
benefit from noise-like aliasing6. We then tested a trajectory that
incorporates the center two k-space lines at the beginning of each randomly
sampled shot. This was inspired by calibrationless imaging work, which show improved
reconstruction quality with the addition of a few center lines7,8. Prospective
reconstructions were compared to shot combination using phase navigators
acquired at the end of each echo train. Sampling patterns are shown in Figure
1. We also compared to FSE-PROPELLER, which achieves distortionless DW images
using a hybrid radial-Cartesian trajectory at the cost of 1.5x more excitations
to fully cover k-space.Methods
Cartesian DW data
was acquired using a diffusion prepared, twice refocused M1 nulled 2D FSE
sequence shown in Figure 2a. Non-CPMG magnetization caused by random phase from
diffusion gradients was eliminated using a stimulated echo preparation9.
Simulated and
prospective multi-shot data were reconstructed using Shot Locally Low Rank
(SLLR), which frames the shot combination as a calibrationless reconstruction.
Shot images are treated as coil images, and varying phase between shots is
analogous to unknown coil sensitivities4,5.
SLLR minimizes
Equation (1), where $$$D_i$$$, $$$x_i$$$, and $$$y_i$$$ are respectively the sampling
operator, shot image to reconstruct, and acquired multi-channel data of the $$$i$$$th
shot; $$$F$$$ is the Fourier transform; $$$S$$$ is the sensitivity map; $$$\lambda$$$ is
the regularization parameter; and $$$\Omega$$$ Is the set of nonoverlapping blocks.
The $$$R_b$$$ operator collects a local patch of coil-combined shot images to form
a spatial-shot matrix, shown in Figure 2b.
$$\min_{x_{1\cdots N_{shots}}}\sum_{i=1}^{N_{shots}}\parallel{D_iFSx_i-y_i}\parallel^2_2+\lambda\sum_{k\in\Omega}\parallel{R_b(x_{1\cdots N_{shots}})}\parallel_*\qquad\qquad(1)$$
Uniform, random, and
center-oversampled patterns were tested in simulation using a T2-weighted
breast image, matrix 128$$$\times$$$128, $$$N_{shots}$$$ 12. Shot-to-shot phase maps from
phase navigators acquired during DW imaging of the subject were applied to
simulate motion-induced phase variations.
Sampling patterns
were applied prospectively in the pelvis of two female volunteers with informed
consent and IRB approval. Acquisition
parameters were: matrix 128$$$\times$$$128, 38 cm FOV, 8 shots, ETL 16 + (4 for navigator,
2 if center oversampled), slice thickness 5 mm, b-value 500 s/mm2, TR 8000 ms, ESP 6
ms, preparation TE 50 ms, refocusing flip angle 120°, readout BW ±50 kHz, 3T
(GE Signa Premier). DW FSE-PROPELLER with identical matrix size and matching
scan time was collected for comparison. A higher resolution 12-shot scan with
matrix 192$$$\times$$$192 was also performed.
Coil sensitives were
estimated with ESPIRIT10 from b0 images. The $$$\lambda$$$ that provided
the best reconstruction quality without block-like artifacts was
$$$\lambda=0.001$$$ for uniform density, and $$$\lambda=0.01$$$ for center oversampled
data with a 5$$$\times$$$5 window.
Results
Simulated SLLR
reconstructions in high SNR data are shown in Figure 3. Uniformly and randomly
sampled data do not recover all details and have altered contrast. SLLR
reconstruction in center oversampled data corrects these artifacts and matches
the reference.
Figure 4 compares a
uniformly sampled 8-shot SLLR reconstruction with a time matched DW
FSE-PROPELLER. SLLR has improved detail compared to FSE-PROPELLER, which has
noticeable streaking. SLLR reconstruction has improved ghost reduction compared
to the navigator combination, which has residual ghosts from subcutaneous fat.
Figure 5 compares
different sampling patterns in a 12-shot acquisition. Uniform sampling has loss
of structure in low SNR regions, which is partially corrected with random
sampling and further improved with center oversampling. Uniform and random sampling
have reduced organ contrast, which is corrected with center oversampling.Discussion
We have shown that FSE sampling patterns can be modified to
complement low-rank reconstructions. Random ky sampling improves detail in low
SNR areas, and center oversampling preserves contrast in acquisitions with a
high number of shots, which are required to mitigate T2 blurring effects
prominent in single-shot FSE. This may be due to uniform sampling missing the contrast-rich k-space center when the number of shots is high. Although the acquisition
time of DW FSE is 1.5-2$$$\times$$$ longer than EPI due to gradient heating, FSE eliminates
distortion in the phase encode direction. This can be applied to off-resonance
environments where multi-shot EPI would fail, such as DW imaging near metallic breast
biopsy clips, hip prostheses, and gas filled bowel loops.Conclusion
Sampling patterns in Cartesian DW FSE can be modified to
improve low-rank reconstructions. Cartesian DW FSE is more time efficient than
FSE-PROPELLER and has less geometric distortion than EPI.Acknowledgements
R01 EB009055. GE Healthcare.
Karolinska Neuro MR Physics group for pulse programming assistance.References
1. N Chen et al. A robust multi-shot scan strategy
for high-resolution diffusion weighted MRI enabled by multiplexed
sensitivity-encoding (MUSE). Neuroimage. 2013;72:41-47.
2.
Z Zhang et al. Multishot cartesian turbo
spin-echo diffusion imaging using iterative POCSMUSE Reconstruction. JMRI.
2017;6(1):167-174.
3.
M Mani et al. Multi-shot sensitivity-encoded
diffusion data recovery using structured low-rank matrix completion (MUSSELS).
MRM. 2016;78(2):494-507.
4.
Y Hu et al. Motion-robust reconstruction of
multishot diffusion-weighted images without phase estimation through locally
low-rank regularization. MRM. 2019;81(2):1181-1190.
5. JD Trzasko et al. Calibrationless parallel MRI
using CLEAR. IEEE ASILOMAR. 2011. pp. 75-79.
6.
M Lustig
et al. Sparse MRI: The application of compressed sensing for rapid MR imaging.
MRM. 2007;58:1182-1195.
7. JP
Haldar et al. P-LORAKS: Low-rank modeling of local k-space neighborhoods with
parallel imaging data. MRM. 2016;75(4):1499-1514.
8. PJ Shin
et al. Calibrationless parallel imaging reconstruction based on structured
low-rank matrix completion. MRM. 2014;72(4):959-970.
9. Q Zhang
et al. Diffusion-prepared stimulated-echo turbo spin echo (DPsti-TSE). NMR in
Biomedicine. 2017;30(7).
10. M Uecker
et al. ESPIRiT—an eigenvalue approach to autocalibrating parallel MRI: where
SENSE meets GRAPPA. MRM. 2014;71(3):990-1001.