Silu Han1, Chidi Patrick Ugonna1, Mahesh Bharath Keerthivasan2,3, and Nan-kuei Chen1,3
1Biomedical Engineering Department, The University of Arizona, Tucson, AZ, United States, 2Siemens Medical Solutions USA, New York, NY, United States, 3Medical Imaging Department, The University of Arizona, Tucson, AZ, United States
Synopsis
A novel two-dimensional (2D) coil-signature-based phase cycled correction
method has been developed for Nyquist artifact removal in echo planar imaging
(EPI). Our method uses already available
coil sensitivity information, without requiring extra reference scans, to
correct 2D phase errors and can be applied equally well to single-shot and
multi-shot EPI. Our results show that the developed method can effectively reduce
Nyquist artifacts in EPI data acquired using a variety of acceleration schemes,
such as through-plane Multi-band Imaging (MB) and in-plane parallel SENSitivity
Encoding Imaging (SENSE).
Introduction
EPI is a fast-imaging technique, which reduces scan time and dramatically
minimizes motion artifacts. However, EPI is more sensitive to artifacts due to the reversal of readout gradient
polarities between two consecutive echoes1. A unique artifact of
EPI, termed Nyquist artifact, is caused by inconsistency between positive-readout
and negative-readout ky lines.
Traditional techniques are designed to correct k-space trajectory shift
along frequency-encoding direction, referred to as one-dimensional (1D)
correction. However, residual artifacts after 1D correction usually remain significant,
particularly for oblique-plane EPI2 or in the presence of cross-term
eddy current3. Instead, 2D phase correction is more effective in
suppressing Nyquist artifacts. The 2D phase inconsistency map can be obtained
either directly from separately acquired reference scans4 or inherently
from the acquired image data itself using techniques like phase-cycled
reconstruction5. However, reference scans require additional scan time and may be susceptible to changes in subject
position4. Phase-cycled reconstruction requires a relatively large field
of view (FOV) so that artifacts in the image background can be measured5,
which may not be feasible for EPI scans under a small or tight FOV, which are
used to acquire high-resolution images of small regions (e.g. brainstem and
hippocampus) in brain.
To address these limitations, we investigated an inherent 2D
coil-signature-based phase-cycled correction technique for EPI Nyquist removal,
that can be implemented without acquiring additional reference scans and which is
effective for EPI with different acceleration approaches and under different FOVs.Methods
The correction algorithm is outlined below:
Step 1: The phase inconsistency map was determined by using phase-cycled
reconstruction and coil sensitivity profile (CSP).
We first performed phase-cycling along the frequency-encoding direction using Eq.1, in which $$$C_{0}$$$ was cycled
between $$$-\pi$$$ and $$$\pi$$$.
$$ \theta(x,y)=C_{0}(x)+C_{1}\times y \;\;\;\;\;[1] $$
Then Nyquist artifact-free complex image signals can be solved using Eq.2.
$$ I=Cg \;\;\;\;\;[2] $$
where $$$I$$$ represents the positive and negative readout images, $$$C$$$ represents phase inconsistency map with elements $$$\theta(x,y)$$$, and $$$g$$$ represents Nyquist artifact-free complex image signals.
With each possible phase $$$C_{0}$$$, we
generated corresponding CSPs from the corrected complex image $$$g$$$ and determined
the phase value with the minimum CSP difference compared to the
Nyquist artifact-free CSP. Phase cycling was
performed again but this time along the phase-encoding direction using Eq.1
and Eq.2, in which $$$C_{1}$$$
was also cycled between $$$-\pi$$$ and $$$\pi$$$.
Figure 1 shows an illustration of single-band
EPI (SB-EPI) correction. Figure 1(a) shows SB-EPI with strong Nyquist artifact and
Figure 1(b) shows the region used for comparing CSP (marked as green), which is
defined as the overlapping regions of Nyquist artifacts and parent image.
Figure 1(c) shows a series of corrected complex images of one coil. Figure 1(d) shows CSPs of one column from images in Figure 1(c) and Figure 1(e) shows the variation of CSP difference.
Step 2: With the 2D phase map
from step 1 and the unaliased CSP from the
scanner, the unaliased parent-image intensity $$$P$$$
can be resolved using Eq.3 or equivalently using
Eq.4 and Eq.5.
$$ I=CKP\;\;\;\;\;[3] $$
where $$$I$$$ represents aliased positive or negative
read-out complex image intensity; $$$C$$$ is the 2D phase
map from step 1 and $$$K$$$ represents the unaliased CSP.
$$ I_{p,i,s_{l}}(x,y)=\frac{1}{2Acc_{in}}\sum_{z=1}^{Acc_{MB}}\sum_{n=0}^{2Acc-1}K_{z,i}(x,y+\frac{n\times FOV}{2Acc_{in}})P_{z}(x,y+\frac{n\times FOV}{2Acc_{in}})\;\;\;\;\;[4] $$
$$ I_{n,i,s_{l}}(x,y)=\frac{1}{2Acc_{in}}\sum_{z=1}^{Acc_{MB}}\sum_{n=0}^{2Acc-1}e^{i\theta(x,y+\frac{n\times FOV}{2Acc_{in}})}e^{i\theta_{n}}K_{z,i}(x,y+\frac{n\times FOV}{2Acc_{in}})P_{z}(x,y+\frac{n\times FOV}{2Acc_{in}})\;\;\;\;\;[5] $$
where $$$Acc_{MB}$$$ and $$$Acc_{in}$$$ represent
through-plane and in-plane acceleration factors, respectively. $$$K_{z,i}(x,y+\frac{n\times FOV}{2Acc_{in}})$$$ represents the
unaliased complex CSP intensity separated by $$$\frac{FOV}{2Acc_{in}}$$$ at slice $$$z$$$ of coil $$$i$$$; $$$P_{z}(x,y+\frac{n\times FOV}{2Acc_{in}})$$$ represents the
unaliased parent-image intensity separated by $$$\frac{FOV}{2Acc_{in}}$$$ at slice $$$z$$$; $$$\theta(x,y+\frac{n\times FOV}{2Acc_{in}})$$$ represents the
2D phase map from step 1, which is separated by $$$\frac{FOV}{2Acc_{in}}$$$ and $$$\theta_{n}$$$ is known as
priority, which is a phase term caused by the k-space acquisition trajectories.
Figure 2 shows a workflow of correction for single-shot EPI with $$$Acc_{MB}=2$$$ and $$$Acc_{in}=2$$$. If correction is performed for
SB-EPI, Nyquist artifact-free CSP in (F) is the same as unaliased CSP from
scanner in (J).Results and Discussion
Figure 3 shows the aliased images and (1D and 2D) corrected
images: (a) 1st-shot SB-EPI; (b) 2nd-shot SB-EPI; (c)
2-shot SB-EPI. The images in the left column were reconstructed
directly from k-space data, showing strong Nyquist and under-sampling aliasing artifacts.
The images in the middle column were reconstructed with 1D correction. Residual artifacts remain visible and may interfere with the
parent image, as indicated by arrows. The images in the right column were reconstructed with our proposed 2D phase correction technique. Nyquist artifacts are much better suppressed with 2D correction in
comparison to conventional 1D correction.
Figure 4 shows the aliased 2-shot MB-EPI ($$$Acc_{MB}=2$$$) and (1D and 2D) corrected images. Arrows in Figure 4(b-c) indicates significant residual
artifacts after 1D correction. After 2D correction, the Nyquist artifact are effectively eliminated,
shown in Figure 4(d-e).
Quantitative assessment of aliasing artifact, ghost-to-noise ratio (GNR), is shown in Table 1. For 2-shot SB-EPI, GNR is 3.01 after 1D correction
and reduces to 1.57 after 2D correction. For 2-shot MB-EPI, GNR for two
slices reduces to 1.69 and 1.03, respectively, after 2D correction.Conclusions
We have developed a novel 2D coil-signature-based phase-cycled
reconstruction method that can successfully reduce Nyquist artifacts in
single-shot and multi-shot EPI, that have been acquired using a variety of
in-plane and through-plane acceleration approaches. Acknowledgements
No acknowledgement found.References
1. Schmitt, F., Stehling, M., Turner, R. and Mansfield, P.
(2013). Echo-Planar Imaging. Berlin: Springer Berlin.
2. Reeder, S., Atalar, E., Faranesh, A. and McVeigh, E. (1999).
Referenceless interleaved echo-planar imaging. Magnetic Resonance in Medicine,
41(1), pp.87-94.
3. Grieve, S., Blamire,
A. and Styles, P. (2002). Elimination of Nyquist ghosting caused by read-out to
phase-encode gradient cross-terms in EPI. Magnetic Resonance in Medicine,
47(2), pp.337-343.
4.
Xu, D., King, K., Zur, Y. and Hinks, R. (2010). Robust 2D phase correction for
echo planar imaging under a tight field-of-view. Magnetic Resonance in
Medicine, 64(6), pp.1800-1813.
5. Chen, N., Avram, A. and Song, A. (2011). Two-dimensional phase cycled
reconstruction for inherent correction of echo-planar imaging nyquist
artifacts. Magnetic Resonance in Medicine, 66(4), pp.1057-1066.