Sen Jia1, Zhilang Qiu1,2, Lei Zhang1, Haifeng Wang1, Xin Liu1, Hairong Zheng1, and Dong Liang1
1Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Shenzhen, China, 2University of Chinese Academy of Sciences, Beijing, China
Synopsis
Equipping the existing parallel imaging methods
such as ESPIRiT and SPIRiT with a full field-of-view (FOV) calibration could
resolve the fold-over artefacts induced by reducing the imaging FOV to be smaller
than the object size. Full-FOV images could be reconstructed by accurately
resolving the aliased components in image space, or by reconstructing the kspace at a finer sampling interval corresponding to full-FOV. Both approaches requires a separate full-FOV
calibration data which could be acquired efficiently. Reduced FOV Parallel
imaging methods with full-FOV calibration may provide an alternative approach
to treat the common FOV aliasing problem in practice.
Introduction
The Nyquist-Shannon
sampling theorem demands the MR imaging field-of-view (FOV) be larger than the imaged
object to avoid fold-over artifacts1,2,3. In
practice, the imaging FOV may be reduced intentionally to be smaller than the object
size to increase imaging speed without scarifying imaging resolution, e.g., in isotropic
high-resolution whole-brain imaging4. Fold-over/aliasing artifacts
would occur at the image boundary, further limit the choice of subsequent parallel
imaging methods and their acceleration performance, such as increased noise
amplification2 and potential central aliasing artifacts3.
This work proposes extensions to the existing parallel imaging algorithms5,6 to resolve the boundary fold-over artifacts and reconstruct full-FOV images for
reduced FOV parallel imaging.Methods
ESPIRiT5 models
the FOV fold-over of reduced FOV parallel imaging by using multiple coil
sensitivity maps (CSM) correspond to multiple eigenvectors for eigenvalue “=1”.
This model could reconstruct reduced FOV images without central aliasing
artifacts but still suffers from residual boundary fold-over artifacts5.
We propose to firstly estimate a full-FOV ESPIRiT map from calibration data
(ACS) acquired separately at a finer sampling interval than the imaging scan;
then create multiple maps according to the folder-over process of reduced FOV
imaging (Figure 1). Finally, soft-SENSE reconstruction 5 using the
new multiple maps could resolve the aliasing components and reconstruct a
full-FOV image. The proposed ESPIRiT reconstruction with full-FOV calibration
has a similar scan and reconstruction efficiency as the original ESPIRiT. Full-FOV
calibration could also apply to kspace parallel imaging methods such as SPIRiT
and GRAPPA as illustrated in Figure 2. GRAPPA or SPIRiT kernels estimated from
full-FOV ACS could reconstruct the full-FOV kspace from reduced FOV imaging
data by direct convolution or iterative optimization.
In-vivo experiments were IRB approved with informed
consent obtained.All scans were performed on a 3T scanner (UIH uMR 780, China)
with a 32-channel head coil. (1)
High-resolution whole-brain imaging was performed on a 26 years old male
healthy volunteer using a T1
weighted 3D MATRIX sequence (sagittal,
non-selective excitation, 0.6 mm3 isotropic resolution, TE/TR = 8.8/800 ms). The
imaging FOV in the phase and partition encoding directions were
intentionally reduced to 178 (AP, anterior-posterior)
x 152 (LR,
left-right) mm2. Separate ACS (24x24) was acquired with full FOV being
220x206 mm2 for whole-brain coverage. The image size reduced from 366x342 to 296x252 would
benefit both the scan and reconstruction efficiency, especially when iterative reconstruction was utilized. The scan
was further accelerated by 2x2 uniform undersampling and took 4 minutes. (2) Another two 3D scans were performed on a 30 years
old female healthy volunteer at an isotropic resolution of 0.6 mm3. The FOV along the LR direction was reduced from 210 mm
to 144 mm. The two scans were accelerated by 2x2 uniform and 5-fold variable
density Poisson-disc undersampling, respectively. Full-FOV ACS of size 24x24
was acquired with FOV being 220x210 mm2. (3) A head-shaped phantom was scanned by 3D gradient
echo sequence twice using axial imaging orientation (FOV=240(LR)x240(AP)x64(HF) mm3, resolution=1 mm3 , TR/TE=11.8/4.5 ms), one with 100% slice oversampling (TA=6 min), while
the other without oversampling (TA=3 min). A separate full-FOV ACS calibration of size 24x24 was acquired with 100% slice oversampling and took about 2 sec. All accelerated datasets were reconstructed by ESPIRiT/SPIRiT
with reduced- or full-FOV calibration respectively. All iterative reconstruction utilized sparsity regularization with L1 weights manually optimized and 120 iterations. All algorithms were implemented in MATLAB and run on a
workstation with two 40-cores CPUs with 256GB memory.Results
Figure 3 compared the proposed four maps created from the full-FOV ESPIRiT map with the four maps estimated directly from
reduced FOV ACS by ESPIRiT with the eigenvalue larger than 0.95. Soft-SENSE
reconstruction using the multiple maps created from full-FOV CSM could accurately
resolve the aliased pixels and reconstruct a full-FOV image without boundary fold-over
artifacts.
Equipping the image-domain ESPIRiT and k-space
SPIRiT with full-FOV calibration could achieve the full-FOV reconstruction of
reduced FOV imaging data, as demonstrated in Figure 4. Moreover, the proposed full-FOV
reconstruction scheme was compatible with arbitrary sampling schemes and
sparsity regularization. The computational complexity of the proposed extension
to ESPIRiT was similar to the original ESPIRiT (24sec vs. 28sec). However, the computational
burden of the proposed extension to SPIRiT would increase since the matrix size
of the unknown increased from reduced FOV to full-FOV (80sec vs. 140sec).
In Figure 5, the
axial 3D scan without slice oversampling suffered from fold-over artifacts due
to the imperfect slice-selective excitation profile. Using the proposed GRAPPA
or ESPIRiT reconstruction with full-FOV calibration, the fold-over artifacts
could be resolved. This might provide an alternative approach to avoid
fold-over artifacts for slice-selective 3D imaging, if acquiring separate full-FOV
ACS is more efficient than slice oversampling.Discussion
By equipping the existing parallel imaging
methods with separately acquired full-FOV calibration data, the FOV aliasing from
reduced FOV parallel imaging could be accurately resolved, and full-FOV images
could be reconstructed with similar scan and reconstruction efficiency. Reduced FOV parallel
imaging with full-FOV calibration may provide an alternative approach to deal
with the common FOV aliasing problem in practice7. One potential limitation of the proposed
methods is that inconsistency may occur between imaging scan and separate full-FOV
calibration scan due to motion.Acknowledgements
This work is supported by the State Key
Program of the National Natural Science Foundation of China (Grant No.
81830056), the National Natural Science Foundation of China (Grant No. 81801691), the National Key R&D
Program of China (2017YFC0108802 and 2017YFC0112903), and Key Laboratory for Magnetic Resonance and
Multimodality Imaging of Guangdong Province.References
- Hamilton J, Franson D, Seiberlich N. Recent
advances in parallel imaging for MRI. Progress
in Nuclear Magnetic Resonance Spectroscopy 2017,
101:71-95.
- Griswold MA, Kannengiesser S, Heidemann RM,
Wang J and Jakob PM. Field‐of‐view limitations in parallel
imaging. Magn Reson Med 2004, 52:
1118-1126.
- Goldfarb JW. The
SENSE ghost: Field‐of‐view restrictions for SENSE
imaging. J Magn Reson Imaging 2004, 20:
1046-1051.
- Jia S,
Zhang L, Ren L, etc. Joint Intracranial and Carotid Vessel Wall Imaging in 5
minutes using Compressed Sensing accelerated DANTE-SPACE. Eur Radiol 2019, 30:119-127.
- Uecker
M, Lai P, Murphy MJ, etc. ESPIRiT—an eigenvalue approach to autocalibrating
parallel MRI: Where SENSE meets GRAPPA. Magn Reson Med 2014, 71: 990-1001.
- Lustig M and
Pauly JM.
SPIRiT: Iterative self‐consistent parallel imaging reconstruction from
arbitrary k‐space. Magn Reson Med 2010, 64: 457-471.
-
Sandino,
CM, Lai, P, Vasanawala, SS, Cheng, JY. Accelerating cardiac
cine MRI using a deep learning-based ESPIRiT reconstruction. Magn Reson
Med. 2020; 85: 152– 167.