Michael Herbst1
1Bruker BioSpin MRI GmbH, Ettlingen, Germany
Synopsis
Segmented
EPI enables high resolution DWI. However, phase differences between segments
can lead to severe artifacts. This work investigates an algorithm to enable reconstruction
of interleaved segmented acquisitions without the need of additional
calibration or navigator measurements. Given a limited number of interleaves, the
initial phase estimates can be calculated by a traditional parallel imaging
reconstruction, using the unweighted scan of the DWI measurement as a reference.
The ASeDiWA jointly reconstructs all segments of one DWI frame maintaining
their phase information. Therefore, the algorithm allows for an iterative
improvement of the phase estimates included in the joint reconstruction.
Introduction
For high resolution diffusion weighted (DW) EPI, interleaved
segmentation provides substantially improved image quality. One drawback of
such segmentation is the vulnerability to phase fluctuations.
Multiplexed sensitivity encoding1 enables robust
high-resolution DWI by formulating a joint SENSE reconstruction. Such joint
reconstruction was already applied to k-space based algorithms2,3. However, these required
navigators to estimate the segments’ phases.
In this work, an operator for joint reconstruction of interleaved
segmented DW data in k-space without phase navigation is investigated.Methods
Figure 1 visualizes the general idea of the ASeDiWA operator.
First, (fig.1a) an ordinary interleaved combination is displayed, leading to
substantial ghosting artifacts.
Figure 1b shows the GRAPPA reconstruction of each individual
segment. Missing samples are reconstructed according to:
$$x_{is0}(r)=\sum_j g_{rji}^*(\widetilde{R}_{rs}x_{js}),\color{white}{.....}[1]$$
where for each segment ($$$s$$$) the nonacquired k-space
value at a position $$$r$$$ in the $$$i$$$-th coil $$$x_{is0}(r)$$$ is calculated from the acquired data of all
coils of this segment selected by an operator $$$\widetilde{R}_{rs}$$$ about the position $$$r$$$. $$$g_{rji}^*$$$ denotes a set of
weights calculated from reference data obtained from the unweighted (B0) scans.
In fig.1c, the ASeDiWA operator reconstructs missing samples
in one segment, using data from all segments. For this operation, new reference
data are generated by combining the phase information of each segment reconstructed
in Eq.1 with the absolute value of the original reference data in image space:
$$x'_{isn}=IFFT\left\{abs\left\{FFT\left\{x_i\right\}\right\}*phase\left\{FFT\left\{x_{isn} \right\}\right\}\right\},\color{white}{.....}[2]$$
where $$$x'_{isn}$$$ is the modified reference data of a segment $$$s$$$
in the $$$i$$$-th coil.
Given the augmented reference data ($$$x'_{isn}$$$) a second calibration is
formulated:
$$\min_{g_{risn}}\sum_{p \epsilon Calib}\left\| \sum_t \sum_j g^*_{rjitsn}(\widetilde{R}_{pt}x'_{jtn}) - x'_{isn}(p) \right\|^2,\color{white}{.....}[3]$$
where the operator $$$\widetilde{R}_{pt}$$$ selects sampled k-space locations from all
segments and coils about the positions $$$p$$$.
This set of weights ($$$g^*_{rjitsn}$$$) is used for the second
reconstruction step:
$$x_{is(n+1)}(r)=\sum_t\sum_j g_{rjitsn}^*(\widetilde{R}_{rt}x_{jt})\color{white}{.....}[4]$$
To reconstruct a k-space location in one
segment $$$x_{is(n+1)}(r)$$$, data from all
segments are used.
ASeDiWA reconstructs a full k-space dataset for each segment
and coil ($$$x_{is(n+1)}(r)$$$), including information of
the object phase. Thus, the steps described in Eq. 2-4 can be iterated as
denoted by the index n. Note, that the dataset $$$x_{is(n+1)}(r)$$$ is still reconstructed from the acquired data.Measurements and Simulations
DWI data with 4 interleaves were acquired on a 3T system with
32-channel head coil (Tim Trio, Siemens Healthcare). Settings: TE/TR =
107/4000ms, FOV = 220×220×3.0mm3, Matrix = 256×260, 1 B0 scan, 3
b-values (500, 1000, and 2000 s/mm2). Data were reconstructed using
the three algorithms visualized in fig.1.
To evaluate the parallel imaging algorithms, g-factor maps were generated.
For different segmentation factors (2, 3, and 4), data were
reconstructed with GRAPPA, ASeDiWA, and ASeDiWAn (n denoting the
number of additional iterations). Simulations were performed on fully sampled
FLASH data, acquired on a 7T system with a 4-channel coil (BioSpec, Bruker
BioSpin). Settings: TE/TR = 2.576/500ms, FOV = 20×20×1.0mm3, Matrix
= 128×128.
DTI processing was applied on a preclinical dataset with
different reconstruction settings. Data were acquired on a 9.4T system with a four-channel
cryogenic coil (BioSpec , Bruker BioSpin). Settings: TE/TR = 22/2000ms, FOV =
25×25×0.8mm3, Matrix = 128×128×5, 4 segments, 5 B0 scans, b-value
650 s/mm2, 30 directions, with phase navigation.Results
Figure 2 compares the FFT reconstruction of the human data
with the results of the GRAPPA reconstruction and ASeDiWA. For all b-values,
the FFT reconstruction shows ghosting. Ghosting is corrected by GRAPPA as well
as ASeDiWA. The comparison between GRAPPA and ASeDiWA shows improved results of
the ASeDiWA reconstruction.
Figure 3 shows results of the g-factor simulations. The left
side of each image contains the magnitude image, the right side displays the
g-factor maps. From left to right each reconstruction establishes the phase
estimate for the following reconstruction step, improving reconstruction
results over iterations.
Figure
4 shows fractional anisotropy and corresponding color maps of the preclinical measurement.
The first column (a) shows the standard FFT including the phase information of
the navigator. In Figure 4b the navigator information was discarded, resulting
in ghosting. Figure 4c displays the GRAPPA reconstruction. While artifacts from
ghosting are reduced, a substantial loss in SNR can be observed. In fig.4d, ASeDiWA
shows improved quality of the DTI maps, compared to GRAPPA. Further improvement
can be observed in fig.4e, displaying ASeDiWA2 with two additional iterations.Discussion
An algorithm for joint reconstruction of segmented DWI
without navigators was presented. Phase estimates of each segment were
calculated using a kernel calculated from an unweighted scan of the same
experiment. Particularly an iterative reconstruction was shown to improve the
g-factor and result in robust DWI, even with a small number of receivers. While
the number of receivers limits the approach to a few segments, this constraint is
relaxed by iteratively improving phase estimates. The method was applied to
several in-vivo datasets, rodents as well as human.Conclusion
The presented ASeDiWA algorithm allows for robust
reconstruction of segmented DWI data without additional navigation or
calibration measurements. In comparison to GRAPPA reconstruction approaches,
the method reduces parallel imaging artifacts and improves SNR. An iterative
variant was shown to further improve image quality and to provide robust results
especially in the case of a small number of receiver channels.Acknowledgements
No acknowledgement found.References
- Chen N, Guidon A,
Chang H-C, Song AW. A robust multi-shot scan strategy for high-resolution
diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE). NeuroImage.
2013;72:41-47. doi:10.1016/j.neuroimage.2013.01.038
2.
- Dong
Z, Wang F, Ma X, et al. Interleaved EPI diffusion imaging using SPIRiT-based
reconstruction with virtual coil compression: Compression Accelerated SPIRiT
for iEPI DWI. Magn Reson Med. 2018;79(3):1525-1531. doi:10.1002/mrm.26768
3.
- Liu
W, Zhao X, Ma Y, Tang X, Gao J-H. DWI using navigated interleaved multishot EPI
with realigned GRAPPA reconstruction: Multishot DWI Using Realigned GRAPPA
Reconstruction. Magn Reson Med. 2016;75(1):280-286.
doi:10.1002/mrm.25586