Yu Chen1, Suhao Qiu1, Zhao He1, Fuhua Yan2, Ruokun Li2, and Yuan Feng1
1Institute for Medical Imaging Technology, School of Biomedical Engineering, Shanghai Jiao Tog University, Shanghai, China, 2Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China
Synopsis
Comparing viscoelastic properties of soft
tissues measured in vivo and ex vivo are of interests for basic
science and clinical translation. Magnetic resonance elastography (MRE) and
indentation are two commonly used techniques for measurement of tissue viscoelasticity.
In this study, we used a gelatin phantom to compare measurements by MRE and
indentation. Using 2nd-order Prony series, frequency response measured from
indentation was estimated. We observed apparent frequency-dependent behaviors
from MRE in the range of 50-100 Hz, but not from indentation. Results implied
that MRE and indentation measures at different frequency ranges.
Introduction
Viscoelastic properties of soft tissue were
traditional measured using indentation or rotational rheometry.1,2
Magnetic resonance elastography (MRE), as a non-invasive method, is clinically used
for in vivo measurement of
viscoelasticity.3,4 Many
comparative studies were conducted between rheometry and MRE for in vivo measurement verification. 5-7
However, few studies have compared measurement between MRE and indentation. In
this study, we used 2nd-order Prony series to analyze the frequency response of
a gelatin phantom based on indentation test. The results were compared with
that from a multi-frequency tests of MRE. Methods
Gelatin
phantoms mimicking soft tissues were measured using a ramp-hold indentation protocol
(Figure 1a).8
Viscoelastic properties were estimated by fitting the relaxation curve using 2nd-order
Prony series:
$$ G(t) = C_0 + C_1·e^{-\frac{t}{\tau_{1}}} + C_2·e^{-\frac{t}{\tau_{2}}}, $$
where
and
are viscoelastic
parameters.
A custom-built magnetic resonance
elastography (MRE) system was used for measuring the dynamic moduli of the
phantom 9.
The experiments were performed on a 3T MRI scanner (uMR 790, United Imaging
Healthcare, Shanghai, China). (Figure 1b). The measured frequencies
ranged from 50 to 100 Hz with a 10Hz step. A spin echo-EPI sequence was used
for imaging, with TR/TE=1000/48 ms,FOV=300x300 mm2, a
matrix size of 127x127, and a slice thickness of 5 mm. Dynamic shear modulus
measured from MRE were estimated by both local frequency estimation (LFE) and
direction inversion (DI).
To analyze the
frequency response of the model estimated from indentation, we convert the 2nd-order Prony series into a 5-parameter Maxwell model (Figure 2). The model parameters were
determined based on the physical interpretation of the Prony series:
$$ E_1 = 3C_0, E_2 = 3C_1, E_3 = 3C_{2}, $$
$$ F_2 = 3\tau_{1}C_1, F_3 = 3\tau_{2}C_{2}. $$
If
incompressibility was assumed, the dynamic shear modulus
can be determined by
.
For a 2nd-order linear viscoelastic model 10,
$$ E(i\omega) = \frac{-q_{2}\omega^{2} + iq_{1}\omega + q_{0}}{-p_{2}\omega^{2} + ip_{1}\omega + 1} $$
where
and
were linear viscoelastic model parameters.
Therefore, by comparing the Laplacian transformation of the constitutive
equation of the 5-parameter model with the analytical dynamic shear modulus,
the linear viscoelastic model parameters
and
can be determined:
$$ q_2 = 3\tau_{1}\tau_{2}(C_0 + C_1 + C_2), q_1 = 3C_0\tau_1 + 3C_0\tau_2 + 3C_1\tau_1 + 3C_2\tau_2, $$
$$ q_0 = 3C_{0}, p_2 = \tau_{1}·\tau_{2}, p_1 = \tau_{1} + \tau_{2}. $$Results and Discussion
The
typical ramp-hold relaxation curve from the indentation tests is shown in Figure 3a. The Wave propagation images
from the MRE tests with 6 different actuation frequencies are shown in Figure 3b. A quantitative comparison
of viscoelastic properties is shown in Figure 4. In the frequency range of
50-100 Hz, the result of MRE showed an apparent frequency-dependent behavior. The
measured shear moduli increased with the vibration frequency increased.
However, in the same frequency range, no such behavior was observed based on
the indentation test.
The differences in the frequency
response between the MRE and indentation tests indicated the considerable
influences of boundary conditions and loading conditions. Since the calculation
of shear modulus in indentation test was based on
,
the different frequency responses of shear modulus between MRE and indentation implied
that the Poisson’s ratio could be frequency-dependent, too.Conclusion
In this study, we analyzed and compared the
viscoelastic parameters estimated from indentation and MRE. Both Indentation
and MRE tests were used to measure the dynamic responses of viscoelastic soft
tissues. The former is more prone to capture the low-frequency range response,
while the latter is better for describing the response at a relatively higher
frequency range. The results could provide helpful information for the ex vivo and in vivo measurements of soft tissues.Acknowledgements
Funding support from grant 31870941 from National
Natural Science Foundation of China (NSFC) and grant 1944190700 from Shanghai
Science and Technology Committee (STCSM) are acknowledged. We thank the
technical support from Jiayu Zhu, Xiaomao Gong, Cong Zhang, Xiaodong Zhou, and
Qiang He from United Imaging Healthcare, Shanghai.References
1. Li Y, Deng J, Zhou
J, Li X. Elastic and viscoelastic mechanical properties of brain tissues on the
implanting trajectory of sub-thalamic nucleus stimulation. J Mater Sci Mater
Med 2016;27(11):163. doi: 10.1007/s10856-016-5775-5
2. Cirka
HA, Koehler SA, Farr WW, Billiar KL. Eccentric rheometry for viscoelastic
characterization of small, soft, anisotropic, and irregularly shaped biopolymer
gels and tissue biopsies. Ann Biomed Eng 2012;40(8):1654-1665. doi:
10.1007/s10439-012-0532-5
3. Bunevicius
A, Schregel K, Sinkus R, Golby A, Patz S. REVIEW: MR elastography of brain
tumors. NeuroImage: Clinical 2020;25:102109. doi: https://doi.org/10.1016/j.nicl.2019.102109
4. Garteiser
P, Doblas S, Van Beers BE. Magnetic resonance elastography of liver and spleen:
Methods and applications. NMR in Biomedicine 2018;31(10):e3891. doi: https://doi.org/10.1002/nbm.3891
5. Klatt
D, Friedrich C, Korth Y, Vogt R, Braun J, Sack I. Viscoelastic properties of
liver measured by oscillatory rheometry and multifrequency magnetic resonance
elastography. Biorheology 2010;47:133-141. doi: 10.3233/BIR-2010-0565
6. Dittmann
F, Hirsch S, Tzschätzsch H, Guo J, Braun J, Sack I. In vivo wideband
multifrequency MR elastography of the human brain and liver. Magnetic Resonance
in Medicine 2016;76(4):1116-1126. doi: https://doi.org/10.1002/mrm.26006
7. Vappou
J, Breton E, Choquet P, Goetz C, Willinger R, Constantinesco A. Magnetic
resonance elastography compared with rotational rheometry for in vitro brain
tissue viscoelasticity measurement. Magnetic Resonance Materials in Physics,
Biology and Medicine 2007;20(5):273. doi: 10.1007/s10334-007-0098-7
8. Chen
Y, Qiu S, Wang C, Li X, Tang Y, Feng Y. Measurement of viscoelastic properties
of injured mouse brain after controlled cortical impact. Biophysics Reports
2020. doi: 10.1007/s41048-020-00110-1
9. Feng
Y, Zhu M, Qiu S, Shen P, Ma S, Zhao X, Hu C-h, Guo L. A multi-purpose
electromagnetic actuator for magnetic resonance elastography. Magnetic
Resonance Imaging 2018;51:29-34. doi: https://doi.org/10.1016/j.mri.2018.04.008
10. Qiang
B, Greenleaf J, Oyen M, Zhang X. Estimating material elasticity by spherical
indentation load-relaxation tests on viscoelastic samples of finite thickness.
IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
2011;58(7):1418-1429. doi: 10.1109/TUFFC.2011.1961