Helge Herthum1, Hugo Carrillo2, Axel Osses2, Sergio Uribe3, Ingolf Sack1, and Cristóbal Bertoglio4
1Experimentelle Radiologie, Charité Universitätsmedizin Berlin, Berlin, Germany, 2Center for Mathematical Modeling, Universidad de Chile, Santiago, Chile, 3Department of Radiology, School of Medicine, Pontificia Universidad Católica de Chile, Santiago, Chile, 4Bernoulli Institute, University of Groningen, Groningen, Netherlands
Synopsis
Optimal multiple motion encoding (OMME) for phase-contrast MRI was
developed with application to magnetic resonance elastography for unwrapping
motion images. OMME is formulated as a least-squares problem using multiple
phase-contrast measurements with different motion encoding gradients (MEG).
OMME is applied to phantom and in vivo human brain experiments. The wrap-free
motion images are further used to reconstruct shear-wave-speed maps and compared
to conventional phase unwrapping methods. OMME generates wrap free
phase-contrast images with the wrap limit determined by the smallest MEG used, while
maintaining the signal-to-noise ratio(SNR) of the largest MEG, which makes OMME
especially suitable for high SNR applications.
Introduction
In magnetic resonance elastography(MRE), harmonic motion
is encoded in the phase of the complex MR signal proportional to the encoding
efficiency of the motion encoding gradient (MEG)[1-3].
Because the phase can only be measured in the half-open interval [-pi,pi), phase
wraps occur if the encoded phase exceeds this limit. Unfortunately, selecting a
large dynamic range leads to poor quality images since the
“motion-to-noise-ratio"(MNR) in the phase image is inversely proportional
to the dynamic range.
Phase wraps can be removed by unwrapping algorithms,
which may fail if the wrapped regions are highly heterogeneous or noisy. The
true motion cannot be recovered[4].
Another possibility is voxelwise motion reconstruction
using dual-encoding strategies, which are based on unwrapping low dynamic-range
data by exploiting high-dynamic range data[5-7]
at the cost of additional measurements. We extended the mathematical framework
of optimal dual-encoding to optimal multiple motion encoding(OMME). The basic
idea of OMME is to include additional measurements with larger MEGs while
keeping the dynamic range of the motion-sensitive phase image large. OMME
presents a superior performance with respect to noise compared to standard dual
encoding unwrapping and succeeds where other methods fail.
Applying the OMME method to MRE allows obtaining high
motion-to-noise ratio wrap-free phase-contrast images. Reconstructed shear wave
speed maps show better detail resolution by overcoming the unwrapping problem.Theory and methods
OMME is formulated as a least-squares problem for the motion
using an arbitrary number of phase-contrast measurements N with different motion
encoding gradients (MEG) with corresponding dynamic range , i.e.
the inverse of the encoding efficiency. The global minimization of the cost
functional J gives the multiple motion encoding reconstruction u*.
$$J_N (u) = \sum_{j=1}^N \left( 1 - \cos\left( \frac{\pi}{d_j} (u_j - u)\right) \right),$$
OMME’s performance was assessed on MRE data from
heparin phantom experiments and two vivo human brain experiments for the most
robust MEG combination obtained by the theory. The imaging parameters for the
single-shot, spin-echo EPI sequence at the 3T Siemens Magnetom Lumina scanner
are given in Table 1. OMME offers wrap-free phase images with MNR corresponding
to the largest MEG used for OMME phase recovery. To calculate MNR for
experimental data we used the blind noise estimation method of Donoho et al.[8].
The unwrapped wave images were further used to
reconstruct shear-wave-speed(SWS) maps based on phase-gradient wavenumber
recovery k-MDEV[9]
and compared to the ones obtained using conventional phase unwrapping methods
like Flynn[10] and Laplacian[11] unwrapping. Original k-MDEV was adapted to the brain
by replacing the linear radial filter in the spatial frequency domain by a
radial bandpass Butterworth filter (third order, highpass threshold:15 1/m, lowpass
threshold:250 1/m).Results
Figure 1 shows the results of the noise analysis for
the phantom and in vivo data for the wrap-free phase images recovered from
Laplacian, Flynn and OMME based unwrapping. MNR increases with increasing MEG
as noise levels decrease as predicted by theory. OMME based unwrapping outperforms
the other methods in terms of MNR.
Figure 2 presents the waves for the phase contrast cases
and OMME. The dynamic range for the phase contrast images lowers as MEG
increases, resulting in more wraps. The OMME methods show the dynamic range of
the smallest MEG and are fully unwrapped.
Figure 3 presents SWS maps obtained by single and OMME
reconstructions compared with Laplace and Flynn unwrapping applied to the
single encoding images. The SWS maps are more homogeneous towards the center of
the phantom and noisier far from the probe. Increasing the MEG reduces the noise
in the SWS maps. Both conventional phase unwrapping algorithms exhibit different
drawbacks. Laplacian based unwrapping smooths image details as it can be seen
from the disappeared air inclusions in the reconstructed SWS map for the
phantom data. The inclusions also disappear with Flynn since it cannot solve
correctly the phase jump at the air-heparin boundary.
Figure 4 shows SWS maps for volunteer 1 and 2,
reconstructed from OMME phase images using four MEGs, and Laplacian based
unwrapping as well as Flynn unwrapping with the strongest MEG. MRE magnitude
images are shown for anatomical reference. Red arrows indicate areas where OMME
based reconstruction shows higher level of details than unwrapping procedures
by fully recovering fluid/tissue boundaries between brain tissue and either
ventricles or gyri. The transition between the skull and the brain tissue is
also properly reconstructed, while the unwrapping methods smooth that region
leading to spurious stiffness values.Discussion
OMME allowed to successfully combine three and four
MRE wave images with different dynamic ranges in the phantom and volunteer
data, respectively. This leads to improved motion-to-noise ratio (MNR) in the
measured waves and therefore to SWS maps with greater resolution of details
than obtained with conventional unwrapping methods. While conventional
unwrapping methods perform well for a lower amount of phase wraps, thus low MNR,
OMME extends the dynamic range in phase contrast MRI towards high-MNR wrap-free
motion encoding. Conclusion
The proposed OMME method allows for an effective
increase in the dynamic range of phase-contrast images with respect to the
number of MEGs while maintaining the MNR of the image with the lowest dynamic
range. The method may be especially suitable for applications where high-resolution
MRE images with high MNR are needed.Acknowledgements
H.H. and I.S. acknowledge the funding from the German Research Foundation (GRK 2260 BIOQIC, SFB1340 Matrix inVision) and from the European Union’s Horizon 2020 Program (ID 668039, EU FORCE – Imaging the Force of Cancer). C.B. acknowledges the funding from the European Research Council (ERC) under the European Union’s Horizon 2020research and innovation programme (grant agreement No 852544 - CardioZoom). A.O. acknowledge the funding of Conicyt Basal Program PFB-03, Fondecyt 1151512 and Fondap CR2-1511009. A.O and S.U. acknowledge funding from ANID Millennium Science Initiative Program – NCN17–129.References
1 Hirsch, S., Braun, J. & Sack,
I. Magnetic Resonance Elastography:
Physical Background And Medical Applications. (Wiley-VCH, 2017).
2 Singh, S., Venkatesh, S. K., Wang,
Z., Miller, F. H., Motosugi, U., Low, R. N., Hassanein, T., Asbach, P.,
Godfrey, E. M., Yin, M., Chen, J., Keaveny, A. P., Bridges, M., Bohte, A.,
Murad, M. H., Lomas, D. J., Talwalkar, J. A. & Ehman, R. L. Diagnostic Performance
of Magnetic Resonance Elastography in Staging Liver Fibrosis: A Systematic
Review and Meta-analysis of Individual Participant Data. Clin Gastroenterol Hepatol (2014).
3 Klatt, D., Hamhaber, U., Asbach,
P., Braun, J. & Sack, I. Noninvasive assessment of the rheological behavior
of human organs using multifrequency MR elastography: A study of brain and
liver viscoelasticity. Phys Med Biol 52, 7281–7294 (2007).
4 Barnhill, E., Kennedy, P.,
Johnson, C. L., Mada, M. & Roberts, N. Real-time 4D phase unwrapping
applied to magnetic resonance elastography. Magn
Reson Med, [Epub ahead of print] doi: 10.1002/mrm.25332 (2014).
5 Carrillo, H., Osses, A., Uribe, S.
& Bertoglio, C. Optimal Dual-VENC Unwrapping in Phase-Contrast MRI. IEEE Trans Med Imaging 38, 1263-1270,
doi:10.1109/tmi.2018.2882553 (2019).
6 Lee, A. T., Pike, G. B. &
Pelc, N. J. Three-point phase-contrast velocity measurements with increased
velocity-to-noise ratio. Magn Reson Med
33, 122-126,
doi:10.1002/mrm.1910330119 (1995).
7 Yin, Z., Sui, Y., Trzasko, J. D.,
Rossman, P. J., Manduca, A., Ehman, R. L. & Huston, J., 3rd. In vivo
characterization of 3D skull and brain motion during dynamic head vibration
using magnetic resonance elastography. Magn
Reson Med 80, 2573-2585,
doi:10.1002/mrm.27347 (2018).
8 Donoho, D. L., Johnstone, I. M.,
Kerkyacharian, G. & Picard, D. Wavelet shrinkage: asymptopia? Journal of the Royal Statistical Society:
Series B (Methodological) 57,
301-337 (1995).
9 Tzschatzsch, H., Guo, J.,
Dittmann, F., Hirsch, S., Barnhill, E., Johrens, K., Braun, J. & Sack, I.
Tomoelastography by multifrequency wave number recovery from time-harmonic
propagating shear waves. Med Image Anal
30, 1-10,
doi:10.1016/j.media.2016.01.001 (2016).
10 Flynn, T. J. Two-dimensional phase unwrapping
with minimum weighted discontinuity. JOSA
A 14, 2692-2701 (1997).
11 Schofield, M. A. & Zhu, Y. Fast
phase unwrapping algorithm for interferometric applications. Optics letters 28, 1194-1196 (2003).