Synopsis
Elastography in MRI (MRE) and ultrasound (USE)
are the most reliable imaging techniques for liver fibrosis detection. However,
there is no established, standardized material that mimics the elasticity and
viscous behavior of liver tissue and can be used as a phantom for MRE and USE. We
developed a phantom based on polyacrylamide which shows a viscoelastic dispersion
from 5-3000 Hz similar to that observed in the healthy human liver, which is measurable
by MRE and USE, stable over months and can be produced in a reproducible way.
The novel phantom facilitates technical improvements and cross-modality
comparisons in MRE and USE.
Introduction
Experiments on
standardized and realistic viscoelastic phantom materials are crucial for technical
development of elastography methods, improving data consistency and clinical
acceptance. However, current commercial phantoms (Elasticity QA Phantom 049, CIRS,
Inc. USA and MR Elastography Phantom, Resoundant, Inc. USA) cannot mimic realistic
tissue properties in terms of shear stiffness and viscous damping of shear
waves [1]. Literature data show that healthy human
liver has a marked shear wave speed (SWS) dispersion which is not mimicked by available
phantoms [2-8]. Especially ultrasound
elastography (USE) suffers from the fact that methods by different vendors
provide different SWS values. Quantitative Imaging Biomarkers Alliance (QIBA,
RSNA) has addressed this problem and recommended using phantoms with realistic
viscoelastic properties [9, 10]. Furthermore,
available phantoms are not well investigable by both magnetic resonance
elastography (MRE) and USE. We here address the need for a well-characterized phantom
which mimics viscoelastic properties of liver tissue, can be standardized manufactured
and is be suitable for MRE and USE.Methods
Phantom preparation
Three phantoms were manufactured consisting of $$$74.1\,\mathrm{wt}\%$$$ water,
$$$22\,\mathrm{wt}\%$$$ acrylamide, $$$1.7\,\mathrm{wt}\%$$$
ammonium persulfate and $$$1.7\,\mathrm{wt}\%$$$ tetramethylethylendiamin. Microcrystalline
cellulose ($$$0.5\,\mathrm{wt}\%$$$) was added providing scatters for
ultrasound. The material was filled into a two-liter cuboid container ($$$14.5\,\mathrm{cm}\,\times\,14.5\,\mathrm{cm}\,\times\,9.5\,\mathrm{cm}$$$)
and stored at room temperature.
Viscoelastic characterization
After one month ripening, one phantom was fully characterized over a wide
frequency range ($$$5\,-\,3000\,\mathrm{Hz}$$$) using rheomtery, MRE (clinical
MRI and compact tabletop MRI scanner) and USE. For each vibration frequency, SWS
as surrogate for elastic properties and penetration rate (PR) as a surrogate
for viscous properties were derived. The relation to the complex shear modulus $$$G^*=|G^*|\cdot\exp(\mathrm{i}\,\phi)$$$
is:
$$SWS=\sqrt{\frac{2\,|G^*|}{\rho\,(1+\cos(\phi))}}$$
$$PR=\sqrt{\frac{2\,|G^*|}{\rho\,(1-\cos(\phi))}}\frac{1}{2\,\pi}$$
Further imaging parameters of MRI (T1, T2, apparent diffusion coefficient
(ADC)) and ultrasound (ultrasound speed, ultrasound attenuation and
echogenicity) were obtained.
Rheomtery
Viscoelastic properties in the low frequency range ($$$5\,–\,60\,\mathrm{Hz}$$$)
were quantified using shear oscillatory rheomtery (Modular Compact Rheometer 301,
Anton Paar GmbH, Austria).
Clinical MRE and USE
A vibration plate was immersed in the phantom material in order to induce
plane shear waves in the frequency range from $$$20$$$ to $$$200\,\mathrm{Hz}$$$
(Figure 1). For clinical MRE, experiments were performed on a 3-Tesla MRI
scanner (Siemens Magnetom Lumina, Erlangen, Germany) using a single-shot,
spin-echo echo-planar imaging sequence. Further imaging parameters were: echo
time $$$(TE)\,=\,112\,\mathrm{ms}$$$, repetition time $$$(TR)\,=\,2000\,\mathrm{ms}$$$,
motion encoding gradient (MEG) amplitude$$$\,=\,34\,\mathrm{mT}/\mathrm{m}$$$,
MEG duration$$$\,=\,47\,\mathrm{ms}$$$ with fractional encoding. For USE, we
used a SonixMD (Utrasonix, USA) scanner with a convex probe (C5-2) and a frame
rate of $$$125\,\mathrm{Hz}$$$ [6]. Analysis for
each frequency was done by drawing manual profiles perpendicular to the wave
fronts and fitting by the analytical solution of a plane shear wave.
Tabletop MRE
For high frequency characterization ($$$300\,-\,3000\,\mathrm{Hz}$$$) a customized
0.5-Tesla tabletop MRI scanner (Mag Spec, Pure Devices GmbH, Germany) with a
permanent magnet was used [11]. The phantom was filled in a glass tube ($$$7\,\mathrm{mm}$$$ diameter)
and sealed. Due to the cylindrical geometry, the shear wave field could be
described and analyzed by analytical Bessel function. A spin-echo imaging
sequence was used with the following parameters: $$$TE\,=\,27\,\mathrm{ms}$$$, $$$TR\,=\,1100\,\mathrm{ms}$$$, MEG amplitude$$$\,=\,200\,\mathrm{mT}/\mathrm{m}$$$, MEG duration$$$\,=\,20\,\mathrm{ms}$$$.
Diffusion weighted imaging (DWI) was used with $$$7\,b$$$-factors, linearly
spaced between $$$50\,-\,800\,\mathrm{s}/\mathrm{mm}^2$$$
to compute ADC.
Long-term stability and reproducibility
All three phantoms were characterized after one month ripening by
tabletop MRE and rheomtery ensuring reproducibility. One phantom was measured
over four months by tabletop MRE assessing long-term stability.Results
SWS and PR values
measured by Rheometry, clinical and tabletop MRE and USE were consistent over a
wide frequency range (Figure 2). SWS and PR
followed the fractional Maxwell-model (Springpot serial to dashpot; $$$\alpha_{\mathrm{SP}}\,=\,0.345\,\pm\,0.002$$$,
$$$\mu_{\mathrm{SP}}\,=\,8.45\,\pm\,0.04\,\mathrm{kPa}$$$, $$$\eta_{\mathrm{SP}}$$$
set to $$$1\,\mathrm{Pa}\,\mathrm{s}$$$ and $$$\eta_{\mathrm{DP}}\,=\,55\,\pm\,5\,\mathrm{Pa}\,\mathrm{s}$$$).
For reproducibility, we calculated for each frequency the range between the
minimum and maximum values of SWS and PR over all three phantoms which was, on
average, $$$0.07\,\mathrm{m}/\mathrm{s}$$$ for the two parameters. For the long-term
behavior, we observed a significant increase of SWS ($$$0.14\,\pm\,0.06\,\mathrm{m}/\mathrm{s}$$$,
$$$p\,=\,7\cdot10^{-13}$$$) and penetration rate ($$$0.25\,\pm\,0.09\,\mathrm{m}/\mathrm{s}$$$,
$$$p\,=\,2\cdot10^{-14}$$$) during the first month. From one to four months we observed
a further weak but significant increase of SWS ($$$0.05\,\pm\,0.06$$$, $$$p\,=\,9\cdot10^{-4}$$$)
and of PR ($$$0.07\,\pm\,0.10$$$, $$$p\,=\,7\cdot10^{-4}$$$). Parameters
measured by tabletop MRI at $$$0.5\,\mathrm{T}$$$ were $$$T1\,=\,923\,\pm\,191\,\mathrm{ms}$$$,
$$$T2\,=\,686\,\pm\,9\,\mathrm{ms}$$$ and $$$ADC\,=\,1439\,\pm\,149\,\mathrm{µm}^2/\mathrm{s}$$$.
Parameters derived by ultrasound were speed
of sound$$$\,=\,1606\,\pm\,8\,\mathrm{m}/\mathrm{s}$$$ and ultrasound
attenuation$$$\,=\,0.4\,\pm\,0.01\,\mathrm{dB}/\mathrm{cm}/\mathrm{MHz}$$$. Echogenicity
was visually similar to that of healthy human liver. Discussion
Our proposed phantom shows SWS dispersion very
similar to in vivo healthy human
liver (Figure 3). Furthermore, the viscosity quantified by PR better
reflects the behavior of in vivo
human liver than other phantoms, which have a higher penetration rate [8]. The good
reproducibility of the phantom material properties permits a straightforward comparison
of techniques in MRE and USE across modalities and platforms. Over time we
observed a continuous decreasing change of SWS and PR, indicating that the
phantom becomes stable over time. However, the long-term stability needs to be
further investigated. MRI and ultrasound parameters are roughly comparable to in vivo liver.Conclusion
We have introduced a novel phantom for MRE and USE which has viscoelastic
properties close to healthy human liver within a wide frequency range. The
phantom is measurable by MRI and ultrasound and can thus be used for a
straightforward comparison of elastography techniques across modalities and
platforms.Acknowledgements
Funding from the
German Research Foundation (GRK 2260 BIOQIC, SFB1340 Matrix in Vision) is
gratefully acknowledged.References
1. Felix
Schrank, H.T., Angela Ariza de Schellenberger,
Jürgen Braun, and I. Sack, Heparin as MRE phantom material with
viscoelastic powerlaw properties similar to soft biological tissues. 2017,
Charité-Universitätsmedizin Berlin, Berlin, Germany
2. Chen, S., et al., Assessment
of liver viscoelasticity by using shear waves induced by ultrasound radiation
force. Radiology, 2013. 266(3):
p. 964-70.
3. Muller, M., et al., Quantitative
viscoelasticity mapping of human liver using supersonic shear imaging: preliminary
in vivo feasibility study. Ultrasound Med Biol, 2009. 35(2): p. 219-29.
4. Nightingale, K.R., et al., Derivation and analysis of viscoelastic properties in human liver:
impact of frequency on fibrosis and steatosis staging. IEEE Trans Ultrason Ferroelectr
Freq Control, 2015. 62(1): p.
165-75.
5. Dittmann, F., et al., In
vivo wideband multifrequency MR elastography of the human brain and liver.
Magn Reson Med, 2016. 76(4): p.
1116-26.
6. Tzschätzsch, H., et al., Two-Dimensional Time-Harmonic Elastography of the Human Liver and
Spleen. Ultrasound Med Biol, 2016. 42(11):
p. 2562-2571.
7. Tzschätzsch, H., et al., Tomoelastography by multifrequency wave number recovery from
time-harmonic propagating shear waves. Med Image Anal, 2016. 30: p. 1-10.
8. Hudert, C.A., et al., Tomoelastography
for the Evaluation of Pediatric Nonalcoholic Fatty Liver Disease.
Investigative radiology, 2019. 54(4):
p. 198-203.
9. Hall, T.J., et al. RSNA/QIBA:
Shear wave speed as a biomarker for liver fibrosis staging. in 2013 IEEE International Ultrasonics
Symposium (IUS). 2013.
10. Palmeri, M., et al. RSNA
QIBA ultrasound shear wave speed Phase II phantom study in viscoelastic media.
in 2015 IEEE International Ultrasonics
Symposium (IUS). 2015.
11. Braun, J., et al., A
compact 0.5 T MR elastography device and its application for studying
viscoelasticity changes in biological tissues during progressive formalin
fixation. Magnetic resonance in medicine, 2018. 79(1): p. 470-478.