Cian Scannell1, Torben Schneider2, Ebraham Alskaf1, Filippo Bosio1, Joao Tourais3, Javier Sanchez-Gonzalez4, Mariya Doneva5, Christophe Schuelke5, Jakob Meineke5, Jochen Keupp5, Jouke Smink6, Marcel Breeuwer7, Amedeo Chiribiri1, Markus Henningsson8,9, and Teresa M Correia1
1School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2Philips Healthcare, Guildford, United Kingdom, 3Magnetic Resonance Systems Lab, Delft University of Technology, Delft, Netherlands, 4Philips Healthcare Iberia, Madrid, Spain, 5Philips Research, Hamburg, Germany, 6MR Clinical Science, Philips Healthcare, Best, Netherlands, 7Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands, 8Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden, 9Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden
Synopsis
First-pass perfusion cardiac MR
(FP-CMR) is an essential tool for characterizing ischemic
heart disease. Moreover, automated quantitative methods enable a reliable and
operator-independent assessment of myocardial blood flow. Conventional FP-CMR
has limited spatial resolution and is performed under breath-hold. Therefore,
diagnostic accuracy is compromised by respiratory induced motion artifacts and
false-positives due to dark-rim artifacts. To overcome these challenges,
free-breathing Fat-water separation for mOtion-corrected Spatio-TEmporally accelerated myocardial peRfuSion
(FOSTERS) has been proposed. Here, we implement and evaluate a
high-resolution dual-echo Dixon version of FOSTERS and compare its quantitative
performance to standard-resolution FP-CMR in patients with suspected
cardiovascular disease.
Introduction
First-pass perfusion cardiac MR
(FP-CMR) enables the non-invasive detection of ischemic heart disease.
Conventional FP-CMR is performed under breath-hold to reduce respiratory
motion, which can be challenging for patients.1,2 Moreover,
low-resolution FP-CMR images are prone to dark-rim artifacts, which mimic
perfusion defects and affect diagnostic accuracy. Pixel-wise quantification
of myocardial blood flow (MBF) can be compromised by these artifacts
and by signal loss caused by T2* decay at high contrast concentrations.1-4 Recently, the FOSTERS4,5 framework has been proposed
to address respiratory motion, T2*-related signal loss (which causes MBF
overestimation) and the limited spatial resolution. FOSTERS provides fat-only images, which are used to
estimate respiratory motion, while motion-corrected water-only images are used
for quantifying MBF. FOSTERS combines dynamically varying undersampling with a
motion-corrected reconstruction using low-rank and sparsity constraints to
achieve high-resolution FP-CMR images. The high-resolution acquisition is
interleaved with a low-resolution image with low T1-sensitivity for
estimating the arterial input function (AIF).6 Additionally, the
multi-echo images are used for correcting the AIF for T2*-related signal losses
to further improve MBF quantification. In this work, we propose an accelerated
dual-echo version of FOSTERS to enable compatibility with stress perfusion. The
performance of the free-breathing high-resolution dual-echo Dixon FP-CMR scheme
is compared to the corresponding clinical low-resolution breath-hold FP-CMR in
8 patients with suspected cardiovascular disease. Methods
Eight patients (4 females, 36-78 years
old) were scanned (ethically approved; informed written consent obtained) during
rest with 8-fold k-t accelerated FOSTERS during the first-pass of a contrast bolus
injection (0.075mmol/kg of Gadobutrol at 4ml/s
followed by 25ml saline flush) on a 3T MRI scanner (Achieva, Philips, NL).
FOSTERS consisted of an ECG-triggered dual-saturation, 2D dual-echo gradient
echo sequence with the following parameters: FOV=320×300mm2, three
short-axis slices (basal, mid, and apical); in-plane resolution=1.6×1.6mm2, slice
thickness=10mm, TR/TE1/TE2=2.8/1.1/1.9ms, saturation
delay=100ms, flip angle=15°, acquisition window=70ms and 54-87 dynamic frames.
For comparison, patients were scanned with a clinical breath-hold
low-resolution 2D FP-CMR acquisition6 with parameters: FOV=320×320mm2,
three short-axis slices, in-plane resolution=2.6×2.6mm2, slice
thickness=10mm, TR/TE=2.2/1ms, SENSE=2 and partial Fourier=0.75. To assess the
motion correction performance of the dual-echo FOSTERS, an additional patient
was scanned with FOSTERS in free-breathing and breath-hold, during the same CMR
examination. The total acquisition time for all scans was
60s.
The
schematic
of the free-breathing FOSTERS framework is shown in Fig.
1. Image reconstruction was implemented on the scanner with modified software
and includes the following steps: 1) Dixon reconstruction with low-rank and sparsity
constraints7 which generates water and fat images from k-t
undersampled data after 5 iterations; 2) Fat-only
images are used to estimate the frame-by-frame respiratory motion; 3)
Translational motion correction is performed in k-space by applying a linear
phase shift to the dual-echo images; 4) Motion-corrected water-only images are
generated using a reconstruction with low-rank and sparsity constraints after
50 iterations. The echo images are used for estimating the AIF and T2*-related
signal loss. Finally, FP-CMR
images were automatically segmented using a deep learning-based method and MBF
maps were generated using a Bayesian inference method.8 Dynamic
images were non-rigidly
motion corrected before quantification. Results and Discussion
Representative fat and water dynamic
images from one patient obtained using FOSTERS are shown in Fig. 2. The
fat-only images contain enough structural information for estimating respiratory
motion, to then generate motion-corrected diagnostic water-only images. Fig. 3
shows images from another patient where both free-breathing and breath-hold
FOSTERS was acquired, to evaluate the motion correction performance. More
uniform MBF maps were obtained using the free-breathing approach. This is
likely because free-breathing acquisitions encourage more regular and shallower
breathing, which is easier to correct, and reduces the risk for large amounts
of motion that can occur at the start or end of a long breath-hold. Moreover,
non-rigid motion correction performs better when correcting for small residual
motion than in the presence of large motion. FOSTERS provides high-quality dynamic
images with reduced dark-rim artefacts compared to the standard low-resolution FP-CMR
images (Fig. 4). Moreover, FOSTERS provides uniform quantitative perfusion maps
(Fig. 4-5), as expected in the absence of ischemia and scar under resting
conditions, whereas the presence of motion introduces artifacts (larger
standard deviation) in the standard low-resolution MBF maps. Patient 5 suffered
from frequent ectopic heartbeats, and thus, cardiac motion is the likely cause
of the inhomogeneous MBF maps. The mean MBF (± standard deviation, SD) values were 0.90 (± 0.31) and
1.72 (± 0.35) mL/min/g for FOSTERS and standard sequence, respectively. The
corresponding mean (± SD) standard deviation of the MBF maps was 0.21 (± 0.21)
and 0.36 (± 0.22) mL/min/g. Significant differences (p<0.05) in MBF
were found between the two methods. This is probably in part due to the FOSTERS
T2* correction of the AIF, and therefore more accurate MBF estimates, but also
the denoising nature of the reconstruction, and smaller residual respiratory motion
artifacts.Conclusion
FOSTERS, a k-t accelerated
dual-saturation dual-echo Dixon FP-CMR framework, enables free-breathing and
high-resolution FP-CMR and improves MBF quantification. Here,
FOSTERS was compared to standard low-resolution breath-hold FP-CMR and provided
higher quality diagnostic images with minimal dark-rim artifact. Furthermore,
FOSTERS corrects for respiratory motion and T2*-related signal losses and
hence, provides improved MBF maps. Future studies will aim to test FOSTERS in
patients during stress.Acknowledgements
This work was supported by the Wellcome/EPSRC Centre for Medical Engineering
[WT 203148/Z/16/Z] and The Swedish Research Council [grant 2018-04164]. References
1.
Kellman P, Arai AE. Imaging sequences for first pass perfusion - a review. J Cardiovasc Magn
Reson. 2007;9(3):525-37.
2. Jerosh-Herold M.
Quantification of myocardial perfusion by cardiovascular magnetic resonance. J
Cardiovasc Magn Reson. 2010;12(57):12-57.
3. Kellman P, Hansen MS, Nielles-Vallespin S, et al. Myocardial perfusion
cardiovascular magnetic resonance: optimized dual sequence and reconstruction
for quantification. J Cardiovasc Magn Reson. 2017;19(1):43.
4. Scannell C, Correia T, Villa A, et al. Feasibility of free-breathing quantitative
myocardial perfusion using multi-echo Dixon magnetic resonance imaging. Sci Rep. 2020;10:12684.
5. Tourais J, Schneider T, Scannell C, et al. High-Resolution
Free-Breathing Quantitative Myocardial Perfusion MRI Using Multi-Echo Dixon. ISMRM 2020;1315.
6. Sanchez-Gonzalez J, Fernandez-Jimenez R, Nothnagel N, et al. Optimization of dual-saturation
single bolus acquisition for quantitative cardiac perfusion and myocardial flow
maps. J Cardiovasc Magn Reson. 2015; 7:21.
7. Otazo R, Kim D, Axel L, Sodickson D. Combination of compressed
sensing and parallel imaging for highly accelerated first-pass cardiac
perfusion MRI. Magn Reson Med. 2010;64(3):767-776.
8. Scannell C, Veta M, Villa A, et al. Deep‐Learning‐Based Preprocessing
for Quantitative Myocardial Perfusion MRI. J Magn Reson Imaging. 2020;51:1689-1696.