Maxim Terekhov1, Theresa Reiter2, Michael Hock1, David Lohr1, Wolfgang Bauer1,2, and Laura M. Schreiber1
1Chair of Molecular and Cellular Imaging, University Hospital Würzburg, Comprehensive Heart Failure Center, Wuerzburg, Germany, 2Department of Internal Medicine I, Cardiology, University Hospital Würzburg, Wuerzburg, Germany
Synopsis
7T cardiac MRI is an emerging technique with
the potential to increase the spatial resolution and functional informativity
of the cardiac images. T2*-quantification becomes a more sensitive MRI
method at 7T compared to the lower B0-fields and may provide
valuable information on cardiac tissue remodeling. However, in practice, the
value of T2* at 7T as a marker of tissue alterations can be
diminished by the suboptimal B0-shimming and problems of acquisition and analysis. We present the results of the
reproducibility and quality assurance study with 10 healthy volunteers for the
optimization of T2*-measurement protocol for a future clinical
application.
Introduction
7T cardiac MRI (CMR) is an emerging methodology with the potential to
increase the spatial resolution and functional informativity of the cardiac images.
In particular, at 7T a T2*-quantification becomes essentially more
sensitive compared to clinical B0-fields (1.5/3T) and may provide
valuable information on cardiac tissue structure remodeling both in acute and
chronic cases. However, in practice, at 7T the value of T2* as a
marker of tissue alterations can be diminished by the suboptimal B0-shimming and problems of acquisition and analysis. The
regions suffered from B0-inhomogeneity may show a significant
variation in T2*-maps reconstructed on pixel-basis even when data
were acquired from the same subject in different breath-holds. Elaborating the
robust methodology of acquisition and analysis being able to ensure
reproducible T2*-values and to correct for B0-shimming
imperfections is a prerequisite of using
T2*-contrast-based techniques in clinical routine 7T cMRI. In
this study, we present results of the reproducibility and quality assurance
study with healthy volunteers for the optimization of T2*-measurement
protocol for a future clinical application.
Materials and Methods
All measurements have been done with the approval
of the Local Ethics Committee using Magnetom™ “Terra” 7T MR-scanner (Siemens
Healthineers, Erlangen) 8Tx/16Rx cardiac array with the B1+-shim-presets
adjusted to the subjects cohorts was used [1]. T2*-measurements were
performed using cardiac triggered mGRE with 9 TE-times distributed in the range
[1.1..14.6]ms. Image matrix was 256x205 at FOV=300x300mm. B0-shimming
was performed using the integrated system including coils up to III-d order [2]. The shimming adjustment volume
covering the whole heart was used. Repetitive T2*-scans (5 to
7) were performed with each of the 10 healthy volunteers (m/f) with varying
parallel imaging acceleration factor (R=2,3), triggering methods (acoustic
system and ECG), and trigger delay. Magnitude images were used for T2*
quantification and phase images for the B0-map reconstruction.
Segmenting of the left ventricle was performed by an experienced MR-cardiologists.
Comparison of the reproducibility of T2*-curves fitting was done for
two evaluation methods: pixel-based and ROI-based. Coefficient of determination
and confidence bounds were used to evaluate the goodness of T2*-curves
fitting. Phase unwrapping and reconstruction of B0-maps were done
using ROMEO algorithm [3]. One of the examples of the volunteer's scans which revealed
the strong influence of B0-inhomogeneity on T2* in the
posterior wall area was chosen for demonstration in this work. Results
Figure 1 shows frames of two echo trains
acquired in subsequent scans and used for T2*-analysis. Local B0-inhomogeneities
can lead to artifacts in pixel-based T2*-maps varied from scan to
scan for the same subject. Figure 2 shows an example of a segmentation of the LV-myocardium
used for the validation of T2*-analysis reproducibility. Allocation
of segments A1-A6 was performed according to AHA-scheme. Figure
3 shows an example of „goodness-of-fit“ analysis for pixel and ROI-based
fitting of T2*-curves. The maps of the pixel-based coefficient of
determination (R2) show
areas with R2<0.9 values in up to 40% of all pixels. Segment-based analysis ensures
R2>0.9 even for a fitting of drastically shortened and presumably
non-mono-exponential T2*-curves in the areas with strongly
inhomogeneous B0. Figure 4 (a,b) shows T2*-maps done on
pixel- and segment-basis for the six repetitive scans of the same volunteer.
Examples of susceptibility artifacts strongly influencing T2* values
in pixel maps and leading to data misinterpretation are labeled. Panel (c) summarizes the
confidence bounds for T2* in each segment determined over 6 scans.
Figure 5(a) shows the reproducibility of the B0-map reconstruction
from the unwrapped phase maps acquired during T2* measurements
scans. Panel (b) demonstrates the gradient of B0 in the myocardium
segments contributing additively to the measured relaxation rate (R2*=1/T2*)
and significantly changing in the posterior wall compared to the typical T2*
of healthy tissue (15-20ms) measured in the septum region. Discussion
The reproducibility analysis shows that in
typical conditions of 7T CMR in terms of
B0 and B1-heterogeneities
the pixel-based T2*-curve fitting procedure may produce both
small and large scale artifacts not related to tissue property, thus,
misleading the data interpretation. The segment-based fitting provides essentially more reproducible
confidence bounds of T2* values compared to the confidence bounds
for the same segments averaged from pixel-based maps. The analysis
confirms that inhomogeneous B0 conditions of the 7T CMR may essentially influence the reliability of T2*
as the marker of myocardial tissue malformations. The significant susceptibility-induced B0-gradients
still persists in the myocardium after the shimming performed using adjustment
volume allocated to cover the whole heart. The effective B0
gradient computed from 2D B0–maps allows to explain a significant
part of the increased relaxation rate (R2*) due to B0-heterogeneity
(even for an extreme reduction of T2* in the posterior/lateral wall
segments as in the shown data). The
refinement introduced by the T2* values correction procedure should
be based on the 3D B0 -maps data to take into account the gradient
of B0 in the direction transversal to the slice plane. This will be provided by acquiring a 3D
contiguous stack of T2*-data with individually set B0-shimming
volume for each slice.Conclusion
The acquired experience of measurements and
preliminary analysis of the volunteer's data allows for a comprehensive
approach towards T2* measurements in patients to assess alterations of
the myocardial microstructure. Acknowledgements
Financial
support: German Ministry of Education and Research (BMBF, grants: 01EO1004,
01E1O1504).References
[1]Terekhov, M, Elabyad, I, Kögler, C, et al. „Customized B1+-Shaping using
multi-Channel transceiver array prototype for 7 T cardiac MRI with central
elements symmetry“. Proc 28th Intl. Soc. Mag. Reson. Med. (Virtual
meeting, 2020)
[2] Hock, M, Terekhov, M, Stefanescu, MR, et al. B0 shimming of the human heart at 7T. Magn Reson Med. 2020; 85: 182– 196. https://doi.org/10.1002/mrm.28423
[3]
Dymerska, B, Eckstein, K, Bachrata, B, et al. Phase unwrapping
with a rapid opensource minimum spanning tree algorithm. Magn Reson.
Med. 2020; 00: 1– 15. doi: 10.1002/mrm.28563