Vanessa L. Franke1, Justyna Platek1, Philip S. Boyd1, Stephanie Laier2, Karin Mueller-Decker2, Andrey Glinka3, Mark E. Ladd1, Steffen Goerke1, Peter Bachert1, and Andreas Korzowski1
1Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Center for Preclinical Research, Core Facility Tumor Models, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Division of Molecular Embryology, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
31P MRSI allows for the
non-invasive investigation of energy metabolism in vivo and is therewith of interest for research on novel
therapies for cancer. The purpose of this study was to investigate suitable
acquisition strategies for 31P MRSI in tumor-bearing mice at B0=9.4T
in order to monitor in future studies effects of novel therapies
affecting energy metabolism and pH values.
31P MRSI datasets with a spatial resolution of (2.5x2.5x7.5)
mm³ obtained in 50 minutes enabled the quantification of signals in diseased
tissue, while maintaining acceptable separation to healthy tissue.
Introduction
Phosphorus
magnetic resonance spectroscopic imaging (31P MRSI) is a
non-invasive method for the investigation of energy metabolism in vivo. It enables the localized
detection of 31P metabolite signals, as well as mapping of the intracellular
pH value1,2. Therewith, 31P MRSI is of interest for research
on novel therapies for cancer.
Earlier studies using 31P MRS in animal studies using surface coils mainly focused on acquiring unlocalized spectra3 or applied
single voxel methods4,5, which only allows the investigation of a small
region of the measured volume. However, larger spatial coverage may be desirable
in order to compare diseased tissue to healthy control tissue within the same
measurement.
The purpose of this
study was to investigate suitable acquisition strategies for 31P
MRSI in tumor-bearing mice at B0=9.4T in order to monitor in
future studies effects of novel therapies affecting energy metabolism
and pH values.Methods
Nude
mice with DLD1 xenografts implanted subcutaneously on the flank were examined on
a 9.4T small animal MR scanner (Bruker) using a double-resonant 31P-1H volume
resonator (diameter 40mm). Various 2D and 3D
protocols with different matrix sizes
(from (10x10) to (14x14) points in-plane), spatial resolutions (from 3 to 6.75mm in-plane) and
acquisition durations (from 27 to 58min) were tested to assess detectability limits.Mice
examinations were approved by the local regulatory authorities under G284-15.
Two
exemplary protocols are presented in the following. In both protocols, 31P
MRSI data were acquired using an acquisition-weighted 2D CSI sequence with a
slice thickness of 7.5mm, one with an in-plane spatial resolution of (3×3)mm²
(Ttot=46min), and the other with (2.5x2.5)mm² (Ttot=50min). During 25% of the readout time, a 1H
decoupling train (Mlev16) was applied to enhance 31P signals.
Both MRSI datasets were processed by
one-fold spatial zerofilling and application of a Gaussian filter in the time
domain. Two regions of interest (ROIs) covering
muscle tissue and the marginal part of the tumor were defined on an overlay of the fitted PCr amplitude with an 1H image using MITK6 (figures 1,3). 31P
spectra from the ROIs were summed up after correcting zero-order phases and B0-offsets, and evaluated using a home-built Matlab (The
MathWorks) implementation of the AMARES algorithm7.
Intracellular
pH value was calculated using the chemical shift difference between inorganic phosphate (Pi)
and Phosphocreatine (PCr) according to the modified Henderson-Hasselbalch equation1.Results
With the presented protocols,
the obtained localized 31P MR spectra were of good quality where B0
homogeneity was sufficient (figures 2 and 4). In the region of the back muscle,
spectral quantification of all expected resonances (i.e. PCr, ATP, Pi)
could be achieved reliably, enabling the calculation of an intracellular pH
value of around 7.0 in the first dataset, and of around 7.1 in the second.
The spectra from the tumor ROIs show a significantly lower intensity for
the PCr and ATP resonances. In the first dataset, a sharp resonance from PC was
resolved at δ=6.3ppm (figure 2, right). Assuming the broad peak at δ=4.4ppm resulted
from compartmentalized contributions of Pi, a pH value of around 6.7
can be determined in the tumor ROI.
In the tumor ROI of the second dataset (figure 4, right), broad PE and
PC resonances were resolved, and a sharp Pi resonance at δ=5ppm.
Therewith a pH of around 7.2 was determined.
In the tested 3D MRSI protocols, the spectral quality was poorer
compared to the 2D acquisitions, due to the more inhomogeneous B0
obtained after shimming. In the other tested 2D MRSI protocols, similar
spectral quality could be obtained as demonstrated in figures 2 and 4, as long
as the B0 homogeneity was comparable.Discussion
In comparison to earlier studies using surface coils3-5, the
use of a volume resonator with a diameter of 40mm in order to ensure a larger
spatial coverage posed a significant challenge concerning sufficient SNR in this
study. In addition, due
to the position of the tumor in the upper flank relatively close to the
abdomen, B0 shimming was challenging and only sufficient in a limited
region of the measured volume. Therefore, the tested 3D CSI protocols with low
resolution ((5-6.75)mm in-plane) did not yield spectra of sufficient quality. However,
2D protocols with higher resolution ((2.5–3.5)mm in-plane) yielded localized 31P
spectra of good quality in a reasonable measurement time (30-50min), in the
region where B0 homogeneity was sufficient.
The therewith-acquired 31P
spectra from the back muscle are in line with expectations (i.e. clear
resonances of PCr, ATP and Pi, pH around 7).
However, spectra from tumor
ROIs had not in all cases pure signatures from tumor tissue (e.g. increased
Phosphomonoester and -diester signals, higher pH value). This is due to signal contamination from the high signal from adjacent muscle tissue. The broad signal around δ=4.4ppm in the
first dataset (figure 2, right) might result from
compartmentalized contributions of Pi, which would be in line with
the finding of partial necrosis within the tumor, which was verified in a
subsequent resection of the tumor.Conclusion
31P MRSI with large spatial coverage
in mice at B0=9.4T is feasible, despite several challenges. Datasets
acquired with a spatial resolution of (2.5x2.5x7.5)mm³ in 50 minutes enabled
the quantification of signals in diseased tissue, while maintaining acceptable
separation to healthy tissue.Acknowledgements
No acknowledgement found.References
1. de Graaf RA. In Vivo NMR Spectroscopy:
Principles and Techniques: 2nd Edition.; 2007.
2. Korzowski, A, Weinfurtner, N, Mueller, S, et al.
Volumetric mapping of intra‐ and
extracellular pH in the human brain using 31P MRSI at 7T. Magn
Reson Med. 2020; 84: 1707– 1723.
3. Zhou R, Bansal N,
Leeper DB, Glickson JD. Intracellular acidification of human melanoma
xenografts by the respiratory inhibitor m-iodobenzylguanidine plus
hyperglycemia: a 31P magnetic resonance spectroscopy study. Cancer Res. 2000 Jul 1;60(13):3532-6.
PMID: 10910065.
4. Rackayova, V.,
Braissant, O., McLin, V.A. et al. 1H and 31P
magnetic resonance spectroscopy in a rat model of chronic hepatic
encephalopathy: in vivo longitudinal measurements of brain energy metabolism. Metab
Brain Dis 31, 1303–1314 (2016).
5. Bakermans, AJ, Abdurrachim,
D, van Nierop, BJ, Koeman,
A, van der Kroon, I, Baartscheer,
A, Schumacher, CA, Strijkers,
GJ, Houten, SM, Zuurbier,
CJ, Nicolay, K, and Prompers,
JJ (2015), In
vivo mouse myocardial 31P MRS using three‐dimensional image‐selected in
vivo spectroscopy (3D ISIS): technical considerations and biochemical
validations. NMR Biomed., 28, 1218– 1227.
6. Nolden M, Zelzer S, Seitel A, et al. The
medical imaging interaction toolkit: Challenges and advances: 10 years of
open-source development. Int J Comput Assist Radiol Surg.
2013;8(4):607-620.
7. Vanhamme L, Van Huffel S. AMARES:
Advanced Method for Accurate, Robust and Efficient Spectral fitting of MRS data
with use of prior knowledge. J Magn Reson. 1997;43(129):1-2.