Samuel Perron1, Matthew S. Fox1,2, Hacene Serrai1, and Alexei Ouriadov1,2,3
1Physics and Astronomy, The University of Western Ontario, London, ON, Canada, 2Lawson Health Research Institute, London, ON, Canada, 3School of Biomedical Engineering, The University of Western Ontario, London, ON, Canada
Synopsis
To produce high-SNR
images, MRI systems usually employ many signal averages, expensive high
field-strength coils, and/or expensive contrast agents. The proposed sampling
and reconstruction method, requiring no hardware modifications, would produce
images with higher SNR without increasing scan times. Nine under-sampled images
are averaged for every combination of images, and the resulting SNR vs.
averages function is fitted according to the Stretched-Exponential Model (SEM).
The resulting reconstructed image yielded higher SNR than the original image
for all three imaging schemes (FGRE, x-Centric, FE-Sectoral), demonstrating the
feasibility of the proposed method for the first time.
Introduction
The MRI modality is generally a low sensitivity method1 due to the nature of the MRI-signal
formation governed by a very small number of the nuclear spins that contribute
to the resulting signal. It is well
known that the resolution of any imaging modality always depends on the
signal-to-noise ratio (SNR). For MRI,
this limitation is mostly solved only for high field clinical 1H MRI
measurements. In this particular case, the
low sensitivity can be overcome by very high concentrations of water content in
body tissue and the high gyromagnetic ratio of 1H. Unfortunately, this is the only example when
the image-quality and image-resolution are acceptable. In cases of other nuclei with low
gyromagnetic ratios (e.g., 13C/23Na/129Xe)2-4,
low isotopic natural-abundance (2Deuterium)5, low-density (1H/3He/19F
thermally polarized gases)6-8,
or low field MRI9, image quality remains a longstanding
scientific problem which can presently be solved by using expensive MRI-hardware
or expensive enriched isotopes.
Alternatively, higher SNR can be achieved by performing many averages
during, for example, a breath-hold as in the case of the 19F lung
MRI.10
We propose a new method that does not
require any extra MRI-hardware or signal averaging to significantly improve the
quality of MRI images. We will use a
significant k-space under-sampling acquisition method where only 10% of the
k-space points will be acquired per image, resulting in an accelerated type of
k-space data acquisition with acceleration factor (AF) of 10. Thus, the time necessary to acquire one
fully-sampled image should be equal to the time needed to acquire ten
under-sampled images. The MRI image SNR
is directly proportional to the spin-density and √(signal averages), so one can average ten different
under-sampled k-spaces together using a specific averaging-pattern, generating 14
under-sampled k-spaces with different SNR levels. It can be assumed that this SNR difference reflects
the spin-density difference rather than the actual signal level
difference. In this case, a density
decay curve can be fitted and reconstructed using the Stretched-Exponential-Model
(SEM) combined with Compressed Sensing (CS).11,12
We
hypothesize that the SEM-equation
can be adapted for fitting the SNR dependence of the MR-signal similar to
fitting time/b-value dependences.13,14Methods
1H MR was performed on a resolution-phantom at 3.0T
(MR750, GEHC, WI) using a clinical gradient coil set and commercial head RF coil. The
following parameters were used: FOV=10x10cm2, Matrix=128x128, TE/TR=1ms/3sec,
BW=32kHz, 90°.
Nine 2D single-averaged fully-sampled k-spaces
were acquired. We have obtained five k-spaces
using 9-averages, 8-averages, 7-averages, 6-averages, and 5-averages (1_k-space
each); two k-spaces using 4-averages (2_combinations); three separate k-spaces
using 3-averages (3_ combinations); and four separate k-spaces using 2-averages
(4_ combinations). This resulted in 14 k-spaces total.
Three
Cartesian sampling schemes (FGRE, x-Centric15 and 8-sector FE_Sectoral16) were used. 14 fully-sampled k-spaces were retroactively under-sampled (AF=7/10/14) in
the wash-out or SNR attenuation direction. Figure_1 shows the under-sampling patterns for
three sampling schemes/AF. The signal-intensity
of the under-sampled k-spaces were represented as a function of the image
number (Figure_3) and then fit following Abascal method.12 Values of $$$\boldsymbol{n}$$$ =1, …,14 were used
to fit the under-sampled data and generate the pseudo mean fractional-ventilation
maps.
For simplicity, one can assume that the SNR attenuation
reflects a decrease of the fluorinated-gas in the lung after delivery of the oxygen
wash-out breaths.7 Clearly, each new wash-out breath of oxygen replaces
some volume of the fluorinated-gas
in lung, so the signal-intensity of the resulting images were gradually
attenuated. The following equation can
be fitted to the wash-out data:17 $$$\boldsymbol{S(n)} = \boldsymbol{S_0(1-r)^n}$$$, where S0 is the initial signal, n is the breath number, S(n)
is the signal intensity after the nth
wash-out breath and r is the
fractional-ventilation parameter.7,17 r
can be expressed as the fraction of fresh gas entering the lung and the total
volume of gas within the lung ($$$V_{total}$$$):7,17
$$$r = V_{new}/V_{total}$$$ or $$$V_{new}/(V_{new}+V_{old})$$$. To generate a pseudo r-value the traditional approach17 was utilized. The SNR of the original/averaged/reconstructed images were
then calculated.Results
Figure_2 shows 8 images generated by
averaging (1-8) and one original image. The SNR of the 9 k-space averaged image
was 22 and the SNR of the original image was 6.5. Figure_4 shows the reconstructed images for
three sampling-schemes and three AF. SNRFGRE/SNRx-Centric/SNRSectoral=20/28/25.
Figure_5 shows pseudo-fractional-ventilation
maps obtained for three sampling-schemes and three AF. The figure
legend shows the MRI-based mean r; the r-value obtained for the fully-sampled case
was 0.30 ± 0.01.Discussion and Conclusion
In this
proof-of-concept study, we showed that the SEM
equation can be adapted for fitting the SNR attenuation dependence of the MR-signal,
similar to fitting the time/b-value dependences. The
reconstructed images yielded improved SNR over the original fully-sampled
images (>20 vs. 6.5) for all three imaging schemes, with comparable or
identical imaging times. Due to this
technique not requiring extra hardware, the proposed method could be
implemented in current MRI-systems and yield improved images. Further comparison of identifiable biomarkers between
the reconstructed and original images would determine accuracy of the reconstruction,
with the goal of possible integration into clinical-MRI-systems. One potential limitation of the method is the
CS-based reconstruction leading to the artefacts clearly visible for higher AF
(FGRE/x-Centric); however, this can be minimized by using a Deep Learning-based
correction after the fact.18-20Acknowledgements
No acknowledgement found.References
- Song, X.-x., Liu, Z.-j., Xu, X.-z. & Tang, Q. Highly sensitive
MRI contrast agent for enhanced visualization of tumors. New J. Chem. 38,
3813-3818, doi:10.1039/c4nj00183d (2014).
- Thind, K. et al. Detection
of radiation-induced lung injury using hyperpolarized (13)C magnetic resonance
spectroscopy and imaging. Magn Reson Med
70, 601-609, doi:10.1002/mrm.24525
(2013).
- Fox, M. S. et al.
Detection of radiation induced lung injury in rats using dynamic hyperpolarized
(129)Xe magnetic resonance spectroscopy. Med
Phys 41, 072302,
doi:10.1118/1.4881523 (2014).
- Farag, A. et al.
Unshielded asymmetric transmit-only and endorectal receive-only radiofrequency
coil for (23) Na MRI of the prostate at 3 tesla. J Magn Reson Imaging 42,
436-445, doi:10.1002/jmri.24798 (2015).
- LaPlante, G., Ouriadov, A. V., Lee-Sullivan, P. & Balcom, B. J.
Anomalous moisture diffusion in an epoxy adhesive detected by magnetic
resonance imaging. Journal of Applied
Polymer Science 109, 1350-1359,
doi:10.1002/app.28106 (2008).
- Zhang, Z., Ouriadov, A. V., Willson, C. & Balcom, B. J. Membrane
gas diffusion measurements with MRI. J
Magn Reson 176, 215-222,
doi:10.1016/j.jmr.2005.06.009 (2005).
- Ouriadov, A. V. et al. In
vivo regional ventilation mapping using fluorinated gas MRI with an x-centric
FGRE method. Magn Reson Med 74, 550-557, doi:10.1002/mrm.25406
(2015).
- Richards, M. G. in Advances in
Magnetic and Optical Resonance Vol. 5
(ed John S. Waugh) 305-352
(Academic Press, 1971).
- Wawrzyn, K., Ouriadov, A., Hegarty, E., Hickling, S. & Santyr,
G. Mapping 129 Xenon ADC of Radiation-Induced Lung Injury at Low Magnetic Field
Strength Using a Sectoral Approach. Proceedings
of the 23th Annual Meeting of ISMRM, Toronto, Canada, 1492 (2015).
- Couch, M. J. et al.
Pulmonary ultrashort echo time 19F MR imaging with inhaled fluorinated gas
mixtures in healthy volunteers: feasibility. Radiology 269, 903-909,
doi:10.1148/radiol.13130609 (2013).
- Westcott, A., Guo, F., Parraga, G. & Ouriadov, A. Rapid
single-breath hyperpolarized noble gas MRI-based biomarkers of airspace
enlargement. J Magn Reson Imaging 49, 1713-1722, doi:10.1002/jmri.26574
(2019).
- Abascal, J., Desco, M. & Parra-Robles, J. Incorporation of Prior
Knowledge of Signal Behavior Into the Reconstruction to Accelerate the
Acquisition of Diffusion MRI Data. IEEE
Trans Med Imaging 37, 547-556,
doi:10.1109/TMI.2017.2765281 (2018).
- Berberan-Santos, M. N., Bodunov, E. N. & Valeur, B. Mathematical
functions for the analysis of luminescence decays with underlying distributions
1. Kohlrausch decay function (stretched exponential). Chemical Physics 315,
171-182, doi:10.1016/j.chemphys.2005.04.006 (2005).
- Parra-Robles, J., Marshall, H. & Wild, J. M. Characterization of
3He Diffusion in Lungs using a Stretched Exponential Model [abstract]. ISMRM 21st Annual Meeting, 0820 (2013).
- Ouriadov, A. V. & Santyr, G. E. High spatial resolution
hyperpolarized (3) He MRI of the rodent lung using a single breath X-centric
gradient-recalled echo approach. Magn
Reson Med 78, 2334-2341,
doi:10.1002/mrm.26602 (2017).
- Khrapitchev, A. A., Newling, B. & Balcom, B. J. Sectoral
sampling in centric-scan SPRITE magnetic resonance imaging. J Magn Reson 178, 288-296, doi:10.1016/j.jmr.2005.10.006 (2006).
- Santyr, G. E., Lam, W. W. & Ouriadov, A. Rapid and efficient
mapping of regional ventilation in the rat lung using hyperpolarized 3He with
Flip Angle Variation for Offset of RF and Relaxation (FAVOR). Magn Reson Med 59, 1304-1310, doi:10.1002/mrm.21582 (2008).
- Yang, Y., Sun, J., Li, H. & Xu, Z. ADMM-CSNet: A Deep Learning
Approach for Image Compressive Sensing. IEEE
Trans Pattern Anal Mach Intell 42,
521-538, doi:10.1109/TPAMI.2018.2883941 (2020).
- Hammernik, K. et al.
Learning a variational network for reconstruction of accelerated MRI data. Magn Reson Med 79, 3055-3071, doi:10.1002/mrm.26977 (2018).
- Yang, G. et al. DAGAN: Deep De-Aliasing Generative Adversarial Networks for
Fast Compressed Sensing MRI Reconstruction. IEEE
Trans Med Imaging 37, 1310-1321,
doi:10.1109/TMI.2017.2785879 (2018).