Johannes Fischer1, Ali Caglar Özen1,2, Matthias Echternach3, Louisa Traser4, Bernhard Richter4, and Michael Bock1
1Dept.of Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Freiburg Site, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Division of Phoniatrics and Pediatric Audiology, Department of Otorhinolaryngology, Head and Neck Surgery, Ludwig-Maximilians-University, Munich, Germany, 4Institute of Musicians' Medicine, Freiburg University Medical Center, Germany Faculty of Medicine, University of Freiburg, Freiburg, Germany
Synopsis
Single point imaging with rapid encoding
(SPIRE) can dynamically image the oscillations of the vocal folds in the
coronal plane with sub-milisecond temporal resolution. To add spatial encoding
in the third dimension, a slower frequency encoding gradient was applied in slice
direction, where motion is minimal. Electroglottography and projection
navigators are used to detect shifts in the larynx position, which are corrected
during reconstruction. The velocity of the mucosal wave is estimated from the
images.
Introduction
Fast oscillatory motion can be imaged with rapidly switched phase
encoding gradients, which has been shown in the vocal folds for one- and
two-dimensional motion1,2. The free choice of the imaging plane makes
MRI the ideal tool to image the motion in the coronal plane, which is not
possible with established clinical imaging methods such as laryngeal
stroboscopy, where only a transverse view from the top is available. During
phonation, the closed vocal folds are pushed along head-feet (IS) direction by
air pressure created in the lungs, and they move along the left-right (LR) direction,
once they are separated by the pressure. In healthy subjects, the motion along anterior-posterior
(AP) direction is small – thus, conventional frequency encoding can be used
to encode this direction, whereas short SPIRE phase encoding gradients are
needed for the rapid IS and LR motion. In this work we present an extended 2.5D
SPIRE technique with which we measure the velocity of the mucosal wave3 in two locations of the vocal folds
simultaneously.Methods and Materials
Assuming that vocal fold motion occurs only in the coronal plane, we use
rapidly switched phase encoding gradients to encode the two in-plane directions,
and a much slower frequency encoding gradient in slab selection direction. A
diagram of the 2.5D SPIRE sequence is shown in Figure 1. To reduce gradient
switching and to optimize sequence timing, parameters were chosen such that slice
rephasing and readout dephasing compensate. This is the case for $$\mathrm{SPS}=\mathrm{TBWP}\cdot\frac{1+\frac{\mathrm{Rdt}}{T_\mathrm{RF}}}{1+\frac{\mathrm{Rut}}{T_\mathrm{ADC}}},$$where SPS is the number of slices per slab, TBWP and $$$T_\mathrm{SS}$$$ are the time
bandwidth product and duration of the
slab selective RF‑pulse, Rdt is the ramp down time of $$$G_\mathrm{SS}$$$, Rut is the ramp
up time of the readout gradient and $$$T_\mathrm{ADC}$$$ is the duration of the ADC. To
obtain SPS=4, we use TBWP=3.6, Rdt=110μs, Rut=80μs,
$$$T_\mathrm{ADC}$$$=1000μs and $$$T_\mathrm{SS}$$$=550μs. Slice
thickness was set to 2.6mm and readout was oversampled by a factor of 2 to
accommodate the continuous excitation profile. PE gradients are switched after
full ramp down of $$$G_{SS}$$$ to avoid simultaneous switching on multiple gradient
axes and are applied as short as
possible ($$$\Delta t_\mathrm{max}$$$ = 640μs, a compromise between peripheral nerve
stimulation and speed), which yields shorter gradient durations towards k‑space
center. Oversampling in SPI is time inefficient, thus saturation bands were
placed superior and inferior of the field of view and a rectangular field of
view was applied in LR direction to further reduce the matrix size. The
following imaging parameters were used: Slice thickness=3mm, FoV=60mm, matrix
size 60x44x8, TE=1.61ms , TR=3.27ms, FA=10°. In Figure 2 the position of the
slab and the saturation-bands can be seen. Navigator data were acquired every
300ms, and saturation pulses were applied every 100ms.
The volunteer (male, 31y) was asked to sing a constant
frequency of 100Hz and was provided visual feedback via a spectrogram. An
MR-safe electroglottogram (EGG)4 was acquired during the measurement and
synchronized with the MR-acquisition using an optical trigger. MR-signal was
acquired with a loop-coil (R=3.5cm) on top of the EGG-electrodes which are
located on either side of the larynx. From the acquired EGG data, the phase of
the periodic vocal fold motion was identified at the time of each PE event and
used for gating of the MR-signal. MR-data were then sorted into 12 continuous
frames, and the resulting 60x44x8x12 k-space was reconstructed with a total
variation constraint along the temporal (4th-) dimension using BART5. Translation of the larynx during measurements
was estimated from projection navigators using phase only cross correlation
(POCC)6 and Fourier shift theorem was used to correct
for in-plane motion at each readout event prior to reconstruction.Results
In Figure 3, vocal fold motion is shown in 12 dynamic frames with a
temporal resolution of 890μs and compared to the mean
EGG signal. A video of the reconstructed motion in four adjacent slices can be
seen in Figure 4, here the motion is replayed at 1/50th its original speed.
From the data of slices 1 and 2 in Figure 4, the velocity of the upwards-traveling
mucosal wave is estimated to be 73±22 cm/s and
70±27 cm/s, respectively (Figure 5).Discussion
The addition of a frequency encoding gradient in slab selection direction to
a 2D SPIRE acquisition allows for the simultaneous measurement of the vocal fold
oscillations at multiple locations, and it reduces the individual slice
thickness from 5mm to 2.6mm. In the reconstructed images, the closure of the
vocal folds can be seen starting at the bottom and moving upwards as the
mucosal wave until finally opening again at the top. Estimated values for the
mucosal wave velocity were about twofold lower than values measured with
Doppler ultrasound in male volunteers3. This interesting finding could be explained
as with MRI the muscle tissue in the vocal folds is measured which moves at
slower velocities than the mucosal wave which is formed by the superficial
layers of the vocal folds. This might also explain the small gap that remains
between both vocal folds in the reconstructed frames.Acknowledgements
No acknowledgement found.References
[1] J. Fischer, T. Abels,
A. C. Özen, M. Echternach, B. Richter, und M. Bock, „Magnetic resonance imaging
of the vocal fold oscillations with sub-millisecond temporal resolution“, Magnetic
Resonance in Medicine, Bd. 83, Nr. 2, S. 403–411, 2020, doi:
10.1002/mrm.27982.
[2] Fischer,
Johannes, Özen, Ali Caglar, Echternach, Matthias, Richter, Bernhard, und Bock,
Michael, „Sub-millisecond 2D MRI of the Vocal Fold Oscillation using Single
Point Imaging with Rapid Encoding (SPIRE)“, gehalten auf der ISMRM 29th annual
Meeting & Exhibition, Sydney, 2020.
[3] C.
R. Krausert, A. E. Olszewski, L. N. Taylor, J. S. McMurray, S. H. Dailey, und
J. J. Jiang, „Mucosal Wave Measurement and Visualization Techniques“, Journal
of Voice, Bd. 25, Nr. 4, S. 395–405, Juli 2011, doi:
10.1016/j.jvoice.2010.02.001.
[4] A.
C. Özen u. a., „Ensuring safety and functionality of
electroglottography measurements during dynamic pulmonary MRI“, Magnetic
Resonance in Medicine, Bd. 76, Nr. 5, S. 1629–1635, Nov. 2016, doi:
10.1002/mrm.26037.
[5] M.
Uecker u. a., „Berkeley Advanced Reconstruction Toolbox“, in In
Proc. Intl. Soc. Mag. Reson. Med., Toronto, 2015, Bd. 23, S. 2486.
[6] A.
de Oliveira, J. Rauschenberg, D. Beyersdorff, W. Semmler, und M. Bock,
„Automatic passive tracking of an endorectal prostate biopsy device using
phase-only cross-correlation“, Magnetic Resonance in Medicine, Bd. 59,
Nr. 5, S. 1043–1050, Mai 2008, doi: 10.1002/mrm.21430.