Kartiga Selvaganesan1, Yonghyun Ha2, Basong Wu2, Kasey Hancock3, Charles Rogers III2, Sajad Hosseinnezhadian2, Gigi Galiana2, and Todd Constable2
1Biomedical Engineering, Yale University, New Haven, CT, United States, 2Radiology and Biomedical Imaging, Yale University, New Haven, CT, United States, 3Electrical Engineering, Yale University, New Haven, CT, United States
Synopsis
The ability to form
an image without using spatial encoding gradients allows for cheaper MRI
manufacturing costs, particularly in small low field devices suitable for point-of-care
imaging. The Bloch-Siegert shift effect provides an RF alternative to spatial
encoding with gradient coils. This work presents a new approach to evaluate and
optimize nonlinear RF encoding schemes generated by transmit phased arrays. From
thousands of possible encoding patterns, the algorithm finds the set that minimizes
correlations between spatial patterns while maximizing the spatial encoding. Simulated
images demonstrate that the nonlinear spatial encodings provided by
Bloch-Siegert RF pulses can provide high image quality.
Introduction
Recently, there has been a renewed interest in low-field
MRI with the aim of building low cost systems to increase accessibility.1,2 The cost and complexity of
MRI systems can be reduced by eliminating B0-gradient coils.3 To this end, we have
developed a novel, low-field, open (Table-top) MRI system (Figure 1A) that allows
for gradient-free imaging in a nonuniform B0-field at 24mT. The low-field
readout yields T2s long enough to obtain up to 800 echoes in 1 second (Figure 1B).
This provides significant flexibility in designing RF encoding patterns that
maximize scan efficiency. In the proposed system, we independently control the amplitude
and phase of the RF transmit pulse to each element in a 3x3 planar array, to define
encoding patterns analogous to that of nonlinear projections. Considering combinations
that only apply RF pulses to 8 channels with amplitudes of 0 or 100%, and phase 0 or π leads to 2304 possible patterns, making it
challenging to choose an appropriate subset for image encoding. Here we propose
a metric for ranking and assembling a set of encodings with minimal correlation.
Images simulated with this encoding set demonstrate the feasibility of
performing Bloch-Siegert shift4 spatial encoding with
typical nonlinear RF coil profiles.Methods
Slice selection: The B0-field of the Table-top system is generated by an electromagnet that can be dynamically
controlled to achieve nonplanar slice selection. A single nonplanar slice
containing uniform magnetic field strength (1MHz), that was on average 5cm from
the bottom of the flat plane of the magnet was used in our 2D
simulations.
Encoding scheme calculation: A 3x3 planar RF array coil was
designed for Bloch-Siegert shift based encoding (Figure 2A). The B1-field
generated by each coil element ($$$B1_c$$$) was computed using the “Biot-Savart magnetic Toolbox” in MATLAB.5 Figure 2B shows the B1-field
of a coil element across the slice-profile used in our simulations.
Using these B1-fields, encoding patterns were calculated according to the
following equation:
$$B_1=\sum_1^nB1_c*A$$ where
$$$n$$$ is the total number of
transmit channels, and $$$A$$$ is a vector defining the amplitude
and phase of the current passed to each channel.
Data analysis and
optimization: Bloch simulations were
used to compute the phasors generated by each encoding pattern. The L2 norm of
the unwrapped phasor in a 20cmx20cm area in the center of the FOV was used to
quantify the total amount of encoding provided by each candidate phasor. After thresholding
by L2 norm, an optimal set was built by recursively choosing the pattern that
minimized β, a metric which
reflects how much a pattern would increase the L2 norm of the overall
correlation matrix (Figure 3). This was intended to choose a distinctive set of
encodings, while minimizing redundancy in the acquired encodings.
Image reconstruction: We performed algebraic
reconstruction of the MR signal generated by the encoding patterns using the conjugate gradient method for
nonlinear encoding fields.6 The
measured MR signal was calculated as:
$$s_n (τ)= ∫p(r)*C_n (r)*e^{i*ϕ_{BS}(r,τ) } dr$$ where $$$C_n (r)$$$ is the complex coil sensitivity profile, $$$p(r)$$$ is the phantom image, and $$$ϕ_{BS}$$$ is the phase distribution of the encoding
pattern. Results
For the Bloch simulation
experiments we used a spin-echo sequence with TE=16ms, T1=3000ms, T2=250ms,
sampling rate of 0.023ms, and matrix size 128x128 voxels. The Bloch-Siegert
transmit pulse was applied for 6ms (Figure 4A). As seen in Figure 4B-D,
changing the phase offset or the amplitude of the applied Bloch-Siegert RF
pulse creates distinct phase maps, with varying amounts of phase windings. We
can also change the spatial shape of these patterns by selecting different subsets
of coils for Bloch-Siegert transmission. These different patterns assign a
signature phase evolution to each voxel in the image, that can then be
recovered using our nonlinear reconstruction technique. 128 of the 2304 encoding
patterns were selected
(analogous to 128 projections), using the method described in the previous
section, for image reconstruction. Figure 5 shows the reconstructed images of a
high resolution brain scan.Discussion
We
have demonstrated the feasibility of performing nonlinear Bloch-Siegert shift
based imaging on our low-field Table top MRI scanner, through simulation
experiments. The overall image quality of the reconstructed images is
excellent, with minimal blurring at the peripheries. The lower resolution at
the edges is due to the fact that the encoding patterns were optimized for a
region in the center of the FOV, and as such the phase winds more slowly
outside this region. We can address this issue by either selecting encoding
patterns that have been optimized for the entire FOV, or by increasing the
amplitude of the applied Bloch-Siegert transmit pulse.
The simulation algorithm developed
in this study allows us to robustly assess various encoding patterns, build
optimal encoding schemes, and reconstruct MR images. This provides us with the
freedom to generate and evaluate encoding patterns from any RF array coil. We
are currently extending this to evaluate encoding patterns created from
transmitting on different numbers of coils (3, 4, or 5 coils etc.),
as well as expanding the range of phase offsets between coil elements.Conclusion
In
this study we have shown that Bloch-Siegert shift based nonlinear spatial
encoding can be used to perform low-field MR imaging in a nonuniform B0-field.Acknowledgements
No acknowledgement found.References
1. Wald, L. L. Portable MR Technology, <https://www.nmr.mgh.harvard.edu/lab/mr-pig/research/portable-mr-technology> (2020).
2. Rothberg, J. Hyperfine. (2020).
3. Sharp, J. C. & King, S. B. MRI
using radiofrequency magnetic field phase gradients. Magnetic Resonance in Medicine 63,
151-161, doi:10.1002/mrm.22188 (2010).
4. Kartäusch, R. et al. Spatial phase encoding exploiting the Bloch–Siegert shift
effect. Magnetic Resonance Materials in
Physics, Biology and Medicine 27,
363-371, doi:10.1007/s10334-013-0417-0 (2014).
5. L.Queval. BSmag Toolbox User Manual (University of Applied Sciences
Dusseldorf, Germany, 2015).
6. Li, S. et al. Algebraic reconstruction technique for parallel imaging
reconstruction of undersampled radial data: application to cardiac cine. Magnetic resonance in medicine 73, 1643-1653, doi:10.1002/mrm.25265
(2015).