Amir Seginer1, Edna Furman-Haran2,3, Ilan Goldberg4, and Rita Schmidt3,5
1Siemens Healthcare, Rosh Ha'ayin, Israel, 2Life Sciences Core Facilities, Weizmann Institute of Science, Rehovot, Israel, 3The Azrieli National Institute for Human Brain Imaging and Research, Weizmann Institute of Science, Rehovot, Israel, 4Deparment of Neurology, Wolfson medical center, Holon, Israel, 5Neurobiology Department, Weizmann Institute of Science, Rehovot, Israel
Synopsis
We examine the potential to significantly reduce SAR in 7T
fMRI (GRE-EPI) by circumventing the fat-suppression pulse. The resulting
(shifted) lipid artifact is resolved by a parallel-imaging based reconstruction
which separates the lipid and water images.
Simulations, phantom experiments, and fMRI experiments were
performed with the suggested method. SAR was shown to be reduced to less than
half, allowing to shorten the repetition time and/or increase the volume
coverage in fMRI studies.
Introduction
Gradient-echo
echo-planar imaging (GRE-EPI) has become the method of choice for functional
MRI (fMRI) due to its ultrafast acquisition and T2*
sensitivity. However, EPI is also noted for its low effective bandwidth along
the phase encoding (PE) direction1. One consequence of this low
bandwidth is a large apparent spatial shift of the lipids due to the chemical
shift between the lipids and water, resulting in artifacts in the image. The
shift artifact is commonly removed by prepending an RF pulse to suppress the
lipid signal. In this study, we examine the potential to significantly reduce the SAR by circumventing the fat-suppression pulse at 7T.
To compensate for the removal of the fat-suppression pulse we implement a
reconstruction based on parallel imaging to separate the lipid and water
images. By reducing SAR, an EPI scan without fat suppression can offer fMRI
studies greater flexibility to shorten the repetition time and/or increase the
volume coverage.Methods
Separate lipid and
water images can be recovered by considering the acquisition as a simultaneous
multi slice (SMS) problem with a built in CAIPIRINHA shift – due to the
chemical shift – as was demonstrated in Ref. 2. The substantial
lipid-water spatial shift at 7T, for commonly used scan parameters, allows for
a reliable separation. In this study we implemented a reconstruction method for
three parallel imaging aspects: (i) in-plane PE acceleration, (ii) SMS
acceleration, and (iii) lipid-water separation. We used BART3 with
L1 norm to reconstruct the final images. The scans in this study were performed
on a 7T
MAGNETOM Terra (Siemens Healthcare, Erlangen, Germany) using a commercial 1Tx/32Rx head coil (Nova Medical, Wilmington, MA), and a head phantom which
includes both a brain-like compartment and a neck + outer layer (skin, lipids,
and muscle) compartment4. A set of simulations was performed
to examine the effect of the lipid artifact. The simulations were based on
brain GRE water and lipid images, shifting the lipid image to match the EPI
shift. These images and matching sensitivity maps (generated from the images) were
used to reconstruct and simulate an image without Fat Suppression, as well as a
water-only image using the fat separation technique. The lipid image was scaled
to match the different T2* decay (at the EPI TE) of the
lipid and the gray matter (estimated as 12.5±2.5 msec and 24.8±1.4 msec, respectively).
An additional analysis on phantom was performed to examine – as a function of
the Fat Suppression flip angle – both the lipid artifact strength and the
energy contribution of the Fat Suppression pulse. Finally, resting-state fMRI
and an fMRI motor-task experiment were repeated to compare both the tSNR and
the t-test between Fat Suppression and Fat Separation.Results
Figure 1
shows schematically the equivalence of lipid-water separation (and in analogy
also of SMS) to in-plane parallel imaging (e.g. SENSE).
Figure 2
shows the dependence of the lipid artifact strength on the phase of the lipid signal
in simulations. Without fat suppression the artifact reaches a 23% error. Using
“fat separation” a root-mean-square-error (rmse) of less than 1.7% is reached.
The simulation also test the effect of shifting the sensitivity maps by one
pixel to simulate motion during an fMRI experiment. This increased the lipid
error without Fat Suppression to 32%, but the rmse of the water image (after
fat separation) only increased to 3.3%.
Figure 3
shows the (relative) strength of the lipid signal as a function of the Fat Suppression
flip angle. A lipid signal close to zero was found at ~125°, compared to the
sequence’s default flip angle of 110° which contributes ~66% of the total
energy (product GRE-EPI). SAR can be reduced by lowering the flip angle. For
example, 80° will introduce only ~7% lipid artifact, but will still contribute 60%
of the total energy, reducing SAR by only 15%.
Figure 4
gives the g-factor maps for different acceleration factors, demonstrating a
local increase at the lipid artifact region.
Figure 5
demonstrates human imaging at two representative slices out of 60, acquired with ×3 in-plane acceleration and
×2 slice acceleration. SAR levels were 97% with
fat suppression, but only 33% without. SNR and tSNR were estimated in a
resting-state fMRI scenario and demonstrated a respective increase of 30% and 12%-50%
upon removal of the fat-suppression pulse.
Figure 6
summarizes the fMRI motor-task study exhibiting improved statistics using Fat
Separation: the number of voxels with t-test≥2.4 (p=0.99) increased by
×1.45 in one region and
×1.9 in another. Discussion
In this study, we
demonstrated a method which allows to avoid fat suppression in EPI for 7T brain
imaging. This can reduce the GRE-EPI SAR to less than a half. A robust
reconstruction based on SMS-like parallel imaging was demonstrated to separate
the water and lipid images. fMRI scans at 7T can benefit extensively from this
SAR reduction. However, the fat separation method eventually competes for the
limited resources available (multi-channel information) used to accelerate the
acquisition. Further research is required to optimize the FOV shift chosen for
CAIPIRINHA when combined with lipid-water separation.Acknowledgements
No acknowledgement found.References
[1]
Schmitt, F.et al., (1998) Springer Berlin Heidelberg, [2] Uecker M, Lustig M. Proc. Intl. Soc. Mag.
Reson. Med. 20 (2012), 2490. [3] https://mrirecon.github.io/bart,
[4] Jona et.
al, NMR BioMed 2020.