Lieke van den Wildenberg1, Arjan Hendriks1, Wybe van der Kemp1, Dennis Klomp1, and Jeanine Prompers1
1Radiology Department, UMC Utrecht, Utrecht, Netherlands
Synopsis
MR
measurement of increased concentrations of choline containing compounds (tCho)
in the liver could be used as a non-invasive tumor tissue biomarker. However,
3D 1H-MRSI with conventional water suppression and volume selection does
not yield the sensitivity gain expected at higher fields like 7T. As a result,
it cannot be performed in clinically feasible scan times with full coverage of
the liver and at sufficient spatial resolution to detect tumor metastases. With
spectral-spatial pulses, it is possible to spatially map tCho much faster and
at higher SNR than with conventional MRSI, with excellent suppression of both
water and lipid signals.
Introduction
To assess the effectiveness
of tumor treatment, a biomarker is required which is indicative for the
presence of viable tumor tissue and/or tumor growth. In the liver, increased concentrations
of choline containing compounds (tCho) could potentially be used as a tumor tissue
biomarker [1]–[3].
The use of high-field MR realizes a gain in sensitivity, allowing tCho
concentrations to be spatially mapped with 3D 1H-MR spectroscopic
imaging (MRSI) [4],
potentially in very short scan times. However,
using conventional water suppression the large lipid signals in the liver are
not suppressed. Moreover, volume preselection techniques generally used in MRSI
(STEAM/PRESS/etc.) come with high RF power deposition, and constrain the TE and
TR, which is suboptimal for SNR efficiency. Here we propose a two-dimensional
spectral-spatial (SPSP) excitation pulse to selectively excite spins at the Ernst
angle within a narrow spectral frequency range around the tCho resonance and
within a transversal slice in the liver, to suppress unwanted signals from
water and lipids, and to avoid artefacts resulting from B0
inhomogeneities outside of the liver (e.g. in lungs and bowels). The aim was to
reliably map tCho in the liver with 3D 1H-MRSI with maximum
sensitivity.Methods
Simulations were performed
for 2D SPSP pulses using Bloch equations (PulseWizard; Matlab R2019a, MathWorks,
Natrick, Massachusetts). The designed SPSP pulse (16 subpulses of 0.2ms, with 0.7ms
spacing) has a Gaussian shaped amplitude profile in the spectral as well as in
the spatial dimension (Figure 1A/B). A
whole-body 7T MR scanner equipped with a multi-transmit RF system (Achieva,
Philips Healthcare, Best, Netherlands) was used to acquire 3D 1H-MRSI
in a phantom and two healthy volunteers. For the phantom measurement, we used an
8-channel transmit,
quadrature head coil with 32-channel receive (NOVA Medical, Houston, Texas,
USA) with scanning parameters shown in Figure 2.
The spherical phantom contains ethanol (1.6%), acetate, and methanol in arquad
solution with CuSO4. The frequency offset of the SPSP pulse in the 1H-MRSI
scans was set at the water frequency (0Hz) and at the ethanol-CH2
frequency (-373Hz), and a transversal slice was selected. 1H-MRSI
scans with SPSP pulses were compared with conventional 1H-MRSI with
and without VAPOR water suppression (Figure 2). For the in-vivo measurements, eight transceiver
fractionated dipole antennas with 16 additional receive loops were positioned
symmetrically around the body at the position of the liver [5]. SPSP
1H-MRSI scans were performed at the water (0Hz) and tCho (-441Hz)
frequency, with transversal slice selection (Figure 2). In addition, a single-voxel
measurement with VAPOR water suppression was performed. For all measurements, RF
phase and B0 shimming were performed in the region of interest and in-vivo measurements were performed with
free breathing. The channels were combined using the WSVD method [6].Results
The Bloch simulations (Figure 1) predicted a spectral bandwidth of ~0.37kHz, and a slice
thickness of ~4cm/~8cm for the SPSP pulse used in the phantom/liver. The
measured slice profiles and positions were in good agreement with the
simulation, both for the phantom and in-vivo
(Figure 3). For the 1H-MRSI
SPSP measurement in the phantom with the offset set to the ethanol-CH2
frequency, all metabolite signals other
than the ethanol-CH2, methanol and some residual water signal were effectively
suppressed (Figure 4). Using the SPSP pulse,
water suppression was in fact better as compared with VAPOR water suppression (Figure 4). The ethanol/water peak intensity ratios were
similar (~1.0×10-3) for the conventional 1H-MRSI
and the SPSP measurements. In-vivo 1H-MRSI
SPSP measurements with the offset set to the tCho frequency showed excellent
water suppression (50- to 75-fold) and lipid suppression, and a well-resolved
tCho signal (Figure 5). The tCho/water peak
intensity ratio was ~4.6×10-3 and ~1.2×10-3 for the two
volunteers, respectively. In comparison, the tCho/water ratio in the single-voxel
STEAM spectra was ~0.6×10-3.Discussion and Conclusion
We showed that both for
phantom and in-vivo liver
measurements, the performance of the SPSP pulses in the spatial as well as in
the spectral domain was in excellent agreement with the Bloch simulations. Care
was taken that the spectral sidebands of the SPSP pulse (Figure 1C/D) fall outside of the spectral region of
interest and do not excite unwanted resonances. When the offset of the SPSP
pulse was set to the ethanol/tCho frequency, this resulted in excellent water
and lipid suppression and the signal of interest was well resolved. Differences
in tCho/water ratio between the 1H-MRSI and single-voxel
measurements may indicate a higher contribution of short-T2
components, such as taurine, in 1H-MRSI with shorter TE, but
interpretation needs a more careful analysis with fitting and T1 and
T2 corrections. With the proposed technique, it is possible to spatially
map tCho (or another tumor marker) much faster than with a conventional 3D 1H-MRSI
sequence with water suppression, because the repetition time can be kept short
and sensitivity is maximized. Moreover, as SNR is very high, acceleration
techniques like EPI and SENSE can be applied to either facilitate very short
scans or increase spatial resolution. In conclusion, 1H-MRSI with the
designed SPSP pulse allows capturing detailed metabolic information with full 3D
liver coverage at 7T.Acknowledgements
H2020-FETopen:NICIReferences
[1] K. Glunde, C. Jie, and Z. M. Bhujwalla,
“Molecular Causes of the Aberrant Choline Phospholipid Metabolism in Breast
Cancer,” Cancer Res., vol. 64, pp. 4270–4276, 2004.
[2] K. Glunde, E. Ackerstaff, N. Mori, M.
A. Jacobs, and Bhuj, “Choline Phospholipd Metabolism in Cancer: Consequences
for Molecular Pharmaceutical Interventions,” Mol. Pharm., vol. 3, no. 5,
pp. 496–506, 2006.
[3] H. J. Shin, H.-M. Baek, J. H. Cha, and
H. H. Kim, “Evaluation of Breast Cancer Using Proton MR Spectroscopy: Total
Choline Peak Integral and Signal-to-Noise Ratio as Prognostic Indicator,” Am.
J. Roentgenol., vol. 198, no. 5, pp. W488–W497, 2012.
[4] R. Otazo, B. Mueller, K. Ugurbil, L.
Wald, and S. Posse, “Signal-to-Noise Ratio and Spectral Linewidth Improvements
Between 1.5 and 7 Tesla in Proton Echo-Planar Spectroscopic Imaging,” Magn.
Reson. Med., vol. 56, pp. 1200–1210, 2006.
[5] B. R. Steensma et al., “An
8-channel Tx Rx dipole array combined with 16 Rx loops for high-resolution
functional cardiac imaging at 7 T,” Magn. Reson. Mater. Physics, Biol. Med.,
vol. 31, pp. 7–18, 2018.
[6] C. T. Rodgers and M. D. Robson,
“Receive array magnetic resonance spectroscopy: Whitened singular value
decomposition (WSVD) gives optimal Bayesian solution,” Magn. Reson. Med.,
vol. 63, no. 4, pp. 881–891, 2010.