Daniel Christopher Hoinkiss1, Han Nijsink2, Paul Borm3, Sabrina Haase1, Jan Strehlow1, Jurgen Futterer2, and Torben Pätz1
1Fraunhofer Institute for Digital Medicine MEVIS, Bremen, Germany, 2Radboud University Medical Centre (Radboudumc), Nijmegen, Netherlands, 3Nano4imaging GmbH, Aachen, Germany
Synopsis
MR-guided endovascular interventions will enable widely distributed
therapies to be performed without radiation. We present a workflow for interventional
slice steering based on endovascular device tracking to increase the precision
during MR-guided interventions. Merging real-time tracking information based on
passive MRI markers with pre-calculated therapy planning allows the slice
steering to automatically adjust itself to the predicted vessel pathways for smooth
and accurate device monitoring.
Introduction
MR-guided interventions allow widely
distributed therapies to be performed without radiation, one example being
Percutaneous Transluminal Angioplasty (PTA) [1-2]. To support these procedures,
we developed a platform including intervention planning and real-time
intervention support. In previous work we presented a guidewire tracking
algorithm that monitors the position of endovascular devices with high accuracy
[3]. In this abstract we merge the tracking information with the intervention
planning for real-time slice steering to follow the endovascular device during
intervention.Methods
Three parallel MRI slices are used for three-dimensional wire
tracking. A gradient echo sequence was designed for non-flow-dependent
depiction of the guidewire which is equipped with passive MRI markers providing
negative MRI artifacts (Nano4Imaging GmbH, Dusseldorf, Germany). Accelerated by
common techniques like GRAPPA [4] and Partial
Fourier [5], the sequence acquires the three slices in
about 500ms with following sequence parameters: 300mm x 300mm FOV,
144 x 144 acquisition matrix, 5mm slice thickness, TE/TR = 1.7ms/3.4ms, 6°
flip angle, 75% phase resolution, 6/8 partial Fourier and 3-fold
integrated GRAPPA. The wire design, marker detection and wire
reconstruction algorithms were presented in [3] and
are summarized in Figure 1.
The reconstructed guidewire is used to
estimate position and orientation for real-time MRI slice steering. Since the
reconstruction of the guidewire uses a second-degree polynomial, the fitting imaging
plane can be calculated using simple plane equations with three sampling points
(Fig. 2). A bilateral communication protocol is used such that the directions
of the readout and phase-encoding axes are known prior to slice
adjustment. The missing degree of freedom (in-plane orientation) is solved by
automatically aligning the wire tip along the encoding direction that results
in less rotation, which helps to reduce spin history artifacts as well as
sudden direction changes. The plane center is then positioned at the wire tip. Additionally,
an angle α is determined from the curvature of the reconstructed guidewire
(Fig. 2a). Adaption of the plane orientation is only performed when α is larger
than a pre-defined minimum.
A vessel mask with definition of the
envisioned intervention path can be used as a-priori information to align the real-time
slices to the vascular system. Figure 3 shows, exemplarily for a vessel MRI
phantom, the segmented vessel mask with a defined path for intervention (a). A
smoothed representation of the path is computed and synchronized to the wire
tip (b). The image plane calculation uses samples of the path representation
that are (1) in front of the wire tip, (2) near to the wire tip and (3) behind
the wire tip to align the new image plane with the vascular system (c-d).
The full workflow of the real-time
scanner interaction is summarized in Figure 4. After retrieving the MRI images
via ethernet, the guidewire is located and reconstructed before calculating the
new image plane position and orientation. All image processing and calculation
steps are performed on a separate computer and results are fed back to the scanner
console. The real-time slice steering was tested in experiments using 3T
MRI scanners (MAGNETOM Skyra, Siemens Healthineers) at two different sites based on vessel phantoms mimicking vessels with different complexity. Results
In the experiments, the algorithms
included in the wire tracking and reconstruction of the wire geometry together with the calculation of the new imaging plane had an
average computation time below 200ms, ensuring real-time capability of the slice steering. Figure 5a shows the
additional value of using the vascular information for slice steering by
removing the jittering observed when only using the wire geometry (see differences in the blue circles). Figure 5b
and 5c show samples of a time series of real-time images with wire motion and automatic slice
steering enabled.Discussion
Experiments show the importance of
using information of the vascular system for automatic slice steering in MRI
interventions. Otherwise, the real-time slice can become unstable and is prone
to tracking errors. With the additional requirement regarding the angle between
the sample points, the slice steering becomes smoother. In situations where
this angle is not large enough, the adaption of the slice position only is
sufficient for following the device. Low-level ghosting artifacts appear in the
images when moving the imaging plane. This has to be observed in further work.
The images also show distinct wrap-around artifacts that did not disturb wire
tracking. This could be handled by using stronger oversampling in PE direction.
However, in the sequence design, we aimed for high temporal resolution first. The real-time slice steering was implemented using a scanner-specific real-time interface (see Fig. 5c) (Access-i, Siemens Healthineers). However, we also set up the same workflow using the vendor-independent MRI framework gammaSTAR that allows the same, complex real-time interactions in a scanner-agnostic scenario (see Fig. 5b) [6-7].Conclusion
We presented a fast, real-time slice
steering workflow for endovascular interventions that simplifies navigation
through the vascular system by detecting the position of endovascular devices
and adjusting the MRI acquisition in real time to follow the device. Vascular information
from intervention planning are used to predict the course of the vessels for
ideal slice positioning. Ultimately, this workflow can increase precision of
positioning during MR-guided interventions and reduce radiation in widely
applied treatments.Acknowledgements
Funding from the Eurostars-2 joint programme with co-funding from the EU Horizon2020 research and innovation programme (Eurostars E! 11263 - SPECTRE)References
[1] Chambers. Occupational health risks in interventional cardiology. JACC 2015;8:628–30.
[2] Picano et al. The radiation issue in cardiology:
the time for action is now. Cardiovasc Ultrasound. BioMed Central 2011;9:35.
[3] Hoinkiss et al. Guidewire tracking based on passive MRI markers
for MR-guided endovascular interventions. ESMRMB 2020;L01.19.
[4] Griswold et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). MRM 2002;47(6):1202–10.
[5] Feinberg et al. Halving MR imaging time by conjugation: demonstration at 3.5 kG. Radiology 1986;161(2):527–31.
[6] Cordes et al. Portable and platformâindependent MR pulse sequence programs. MRM 2019;86(4):1277-90.
[7] Hoinkiss et al. Event-based traversing of hierarchical sequences allows real-time execution and arbitrary looping in a scanner-independent MRI framework. ISMRM 2020;1043.