Edwin Versteeg1, Sarah M. Jacobs1, Ícaro A.F. Oliveira2, Dennis W.J. Klomp1, and Jeroen C.W. Siero1,2
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Spinoza centre for neuroimaging Amsterdam, Amsterdam, Netherlands
Synopsis
Sound
levels in MRI can be reduced by switching a silent gradient axis beyond the
hearing threshold. In this work, we implemented a silent readout module that
applies such a silent gradient axis (at 7T) to a 3D MPRAGE sequence. This resulted in a
sequence that featured a much lower peak sound level (26 dB reduction), similar
image contrast and imaging time compared to a conventional MPRAGE-scan. This shows that a silent gradient axis provides
a pathway to fast and quiet brain imaging with the potential to translate
to other field strengths.
Introduction
Loud acoustic noise from gradient intensive sequences can
cause anxiety in patients and lead to a degradation in image quality [1,2]. Therefore, sound level
reduction is important to enhance patient comfort and ensure consistent image
quality. This sound level reduction is usually achieved by reducing the amount
of gradient switching. For example, methods like PETRA, RUFIS, ZTE BURST and
Looping Star utilize a modified zero
echo time acquisition and slowly updating gradients [3–6]. Alternatively, we have
shown that considerable reduction in noise perception can be achieved by increasing
the gradient switching frequency beyond the hearing threshold [7]. Here the readout gradient axis is switched at the inaudible
frequency of 20 kHz. In this work, we have
implemented fast and quiet anatomical brain imaging using this silent MRI
readout setup (at 7T) for a 3D MPRAGE T1-weighted sequence. Methods
The
silent gradient axis operates as an additional 4th gradient axis (z-direction)
to the whole-body gradient system and consists of a lightweight gradient head insert
driven by an audio amplifier (18 kW, Powersoft, Italy). This produces 28.6 mT/m
at 20.2 kHz (3600T/m/s, and no noticeable PNS [8]). The setup features an
integrated birdcage transmit coil, fits a 32-channel receive coil (NOVA
Medical, USA), and was placed in a 7T MR system (Philips, The Netherlands).
The
silent readout module uses an inaudible oscillating readout gradient in the phase-encoding
direction to provide additional spatial encoding during the fast gradient-echo readout
of the MPRAGE sequence (Figure 1). Due to the increased encoding efficiency, the
sound production of the other, whole body, gradients in the sequence were decreased
by reducing the slew rate without effecting the scan time. However, the lower slew rate did result in a
longer minimum echo (TE) and repetition time (TR). To compensate for the longer
TE and TR, the signal in white matter, gray matter and CSF was simulated using
extended phase-graphs simulations, which allowed us to optimize the gray versus
white matter contrast while nulling CSF in the quiet MPRAGE scan. Additionally,
a TR-FOCI pulse was used for inversion to mitigate low B1+ amplitude
and inhomogeneities and more uniform image contrast (Figure
2) [9]. A CAIPI-like sampling pattern was used to limit noise
enhancements from variations in the sampling density.
A
volunteer was imaged using both the quiet and a conventional MPRAGE scan while positioned
in the gradient insert. A whole-brain field-of-view of 256x256x172 mm3
was acquired with a 1 mm isotropic resolution, a shot interval of 3000 ms,
radial (ky-kz) ordering, and an inversion time of 1000 ms (Table 1). Data was
reconstructed offline in MATLAB using an iterative SENSE reconstruction [10,11]. For this reconstruction, the spatiotemporal behavior of the
oscillating gradient field was first characterized using a field camera (Skope,
Switzerland) and was used as an input for the reconstruction.
Relative
sound measurements of both scans were performed using a condenser microphone
(Behringer ECM8000). The audio data was processed in MATLAB. Here, exponential filtering and A-weighting
was applied to mimic the fast response setting and output of a sound level
meter. The same data was also used to determine the effect of the 20 kHz sound,
which was estimated from Z-weighting of the audio signal. Results and Discussion
Figure
3 shows very similar gray and white matter contrast for both the conventional
and quiet MPRAGE scans, which was expected from the EPG-simulations. Some
residual loss of contrast is visible in both scans due to insufficiently low B1+
amplitude. Additionally, the quiet scan
shows some signal loss in the neck region, which originates from the short
linear region of the gradient insert. We foresee that part of the signal loss can
be recovered by incorporating the non-linear field into the reconstruction
using e.g. a PSF-reconstruction as is used in WAVE-CAIPI [12] or an expanded signal model [13].
Importantly,
the sound measurements (Figure 4) show a 26 dB reduction in the peak sound
pressure level during the scan. Here, the main source of residual sound
originated from the slowly switching whole-body gradients resulting in a low-frequency
humming sound during the MPRAGE readout. While not perceived as sound, the silent
gradient axis nevertheless produces sound pressure. When this 20 kHz sound is
included this still result in a 13 dB lower sound pressure level in the quiet scan
compared to a traditional MPRAGE scan.
The
quiet MPRAGE scan was 15% slower than the conventional scan due to duty cycle
limitations of the current audio amplifier, which limited the gradient
amplitude. However, the silent gradient axis still allows for an additional
acceleration by using traditional parallel imaging techniques in the
slice-encoding direction, which would allow us to compensate for the hardware
limitations.Conclusion
We
have shown that a silent gradient axis can considerably reduce
sound (26 dB) for MPRAGE scans while maintaining similar image contrast as a
conventional MPRAGE scan. Such a silent gradient axis provides a way to fast
and quiet imaging that can be readily translated to 3 tesla field strengths with
the potential to greatly improve patient comfort for a wide variety of clinical
scans.Acknowledgements
No acknowledgement found.References
1 McNulty, J. P. et al. Radiography 15,
320–326 (2009)
2 Dewey, M. et al. J. Magn. Reson. Imaging 26, 1322–1327
(2007)
3 Grodzki, D. M. et al. Magn. Reson. Med. 67, 510–518 (2012)
4 Solana, A. B. et al. Magn.
Reson. Med. 75, 1402–1412 (2016)
5 Schulte, R. F. et al. Magn. Reson. Med.
81, 2277–2287 (2019)
6 Wiesinger, F. et al. Magn.
Reson. Med. 81, 57–68 (2019)
7 Versteeg, E. et al. in Proceedings of the 27th Annual Meeting
of ISMRM #4586 (2019)
8 Verbeek Spijkerman, J. et al. in Proceedings of the 28th Annual
Meeting of ISMRM #0114 (2020)
9 Hurley, A. C. et al. Magn. Reson. Med. 63, 51–58 (2010)
10 Pruessmann, K. P. et al. Magn. Reson. Med. 46, 638–651 (2001)
11 Versteeg, E. et al. in Proceedings of the 28th Annual Meeting
of ISMRM #3434 (2020)
12 Polak, D. et al. Magn. Reson. Med.
79, 401–406 (2018)
13 Wilm, B. J. et al. Magn. Reson. Med. 65, 1690–1701 (2011)