Myung-Ho In1, Norbert G Campeau1, John III Huston1, Zijing Dong2,3, Kawin Setsompop4,5, Daehun Kang1, Uten Yarach6, Yunhong Shu1, Joshua D Trzasko1, and Matt A Bernstein1
1Department of Radiology, Mayo Clinic, Rochester, MN, United States, 2A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 3Department of Electrical Engineering and Computer Science, MIT, Cambridge, MA, United States, 4Department of Radiology, Stanford University, Stanford, CA, United States, 5Stanford University, Stanford, CA, United States, 6Department of Radiologic Technology, Faculty of Associated Medical Sciences, Chiang Mai University, Chiang Mai, Thailand
Synopsis
Rapid high-resolution echo-planar-imaging
(EPI) with improved image quality could conceivably be utilized as a high-speed
alternative to conventional T2-weighted fast-spin-echo imaging (T2-FSE). A variant
of multi-shot EPI, DIADEM (distortion-free imaging: a double encoding method), was
recently described (5) and shown to yield high quality, high resolution images free
of spatial distortion. In this work, a T2-weighted DIADEM pulse sequence was
optimized with self-calibrated, tilted-CAIPI reconstruction scheme and then comparatively
evaluated to T2-FSE images of the brain in 24 human subjects by two neuroradiologists.
Introduction
Rapid
(sub-minute) high-resolution echo-planar-imaging (EPI) with improved image
quality is a potential high-speed alternative to conventional MR image such as fast-
or turbo-spin-echo (FSE or TSE) sequence. Previous studies demonstrated that high-resolution
EPI with reduced spatial distortion could be achieved with the benefits of high-performance
gradients (700 T/m/s, 200 mT/m) on a compact 3T MRI1,2 and the image quality was
further improved with multi-shot acquisition.3 However, there was still
noticeable distortion in very high in-plane resolution
(<1mm2) imaging even with a 4-shot acquisition.3 Recently, a novel variant
of multi-shot EPI, termed DIADEM (distortion-free imaging: a double encoding
method)4,5, was described to resolve
the spatial distortion issue. Scan time was accelerated with the tilted-CAIPI
(tCAIPI) reconstruction scheme.6 To adopt this approach as
an alternative to T2-FSE, self-calibrated tCAIPI scheme avoiding additional
calibration scan (~20 s) is proposed in this study. The clinical feasibility was
evaluated in a pilot comparison study with radiological review of 24 subjects.Methods
All experiments were conducted on the compact
3T scanner2,7,8 using a 32-channel head
coil (Nova Medical, USA). Self-calibrated tCAIPI acquisition and reconstruction in DIADEM: Undistorted images without wrap-around
artifact could be calculated from the DIADEM data even with a reduced field of
view (rFOV) factor of 8, leading to a 32-shot acquisition for a 256x256 matrix
size on the compact 3T with high-performance gradients3 that enabled a reduction
in an effective echo-spacing of up to 209.3 ms (Fig. 1A). After considering
this as the “full” acquisition (Fig. 1B), a further undersampling factor of 3
was applied with self-calibration (i.e., 2 additional shots at the k-space center
in the spin-warp phase-encoding (SW-PE) dimension along with 7/8 partial
Fourier, which resulted in an 11-shot accelerated acquisition (i.e., 32/3×7/8+2). Without additional
3D calibration data acquisition as suggested in previous studies, a 2D tilted-CAIPI
reconstruction was applied along the tilted (i.e. ky-ks) axis with self-calibration
included in the DIADEM acquisition (Fig. 1C). Radiological evaluation of T2-DIADEM
relative to T2-FSE: After written consent and under an IRB-approved protocol, a pilot
study of 24 subjects were scanned with standard clinical T2-FSE and proposed T2-DIADEM protocols at equivalent
spatial resolution (0.86×0.86×4 mm3) and
imaging matrix size (256×256×38). The acquisition times
were 55 seconds and 1 minute 58 seconds, respectively. Those exams were
reviewed by two board-certified neuroradiologists.
Further artifact reduction: In radiological review of
the pilot data, Gibbs ringing artifacts were observed along the readout
direction with DIADEM. To address this issue, a 1D Fermi filter was applied in
k-space along the corresponding dimension. In addition, pulsation-induced flow
effects caused phase shifts that vary between shots in DIADEM acquisition and
resulted in ghost artifacts along the PE direction of the undistorted image. To
reduce this artifact, flow compensation gradients (FCG) designed to null the
velocity-dependent phase shifts were implemented into the DIADEM sequence (Fig.
2A). The reduction was confirmed by DIADEM scans with and without the FCG from additional
three-subject exams.Results and Discussion
The acquisition time for the DIADEM is faster
by a factor of 2.14 compared to the standard FSE acquisition. Despite this
considerable acceleration, the spatial resolution and contrast were not
compromised (Fig. 2B). Without any additional 3D calibration scan for 3D tilted-CAIPI
reconstruction, in addition, the 2D tilted-CAIPI reconstruction was able to successfully
unfold the DIADEM data only with four calibration lines (i.e. shots) at the
k-space center in the SW-PE dimension (Fig. 1C).
In
radiological review, T2-DIADEM demonstrated better apparent spatial resolution
relative to FSE, despite an equivalent acquisition matrix. This may be because
the longer echo train length (i.e. 11) filling different k-space lines in FSE
(Fig. 2A) causes more apparent T2 blurring. Due to stronger T2* weighting over
the long readout acquisition window in DIADEM (Fig. 2A), T2-DIADEM provided superior
depiction of iron-containing structures, e.g. in regions of globus pallidus,
dentate nucleus, red nucleus, substantia nigra, and chronic microhemorrhage
(Fig. 2B). While the strong T2* weighting was helpful to better identify small
developmental venous anomaly and the multiple small T2 hyperintense
perivascular spaces (Fig. 3), it also resulted in strong signal dropouts in the
skull base (Fig. 4a). Due to its multi-shot nature of the DIADEM, in addition, greater
CSF flow/pulsation artifacts were observed around the brainstem and within the
ventricles, which made it difficult to evaluate flow void from basilar artery
and cavernous internal carotid arteries on T2-DIADEM (Fig. 4). Interestingly, reduced
gray-white matter contrast was shown only in periventricular regions on
T2-DIADEM (Fig. 4c). With the radiologist’s feedback, Gibbs ringing and flow
artifacts were additionally minimized with 1D Fermi filter and FCGs,
respectively (Fig. 5). The overall radiologist impression was that the 55 second
DIADEM acquisition would be a suitable alternative to T2-FSE in a rapid brain assessment
protocol (e.g., for an uncooperative patient).Conclusion
Overall,
T2-DIADEM performed well relative to T2-FSE in the 24 subjects evaluated. While
DIADEM offers some advantages over FSE, it is still not a complete replacement
for T2-FSE. In the current form, it would
be most appropriate for use in a rapid brain assessment protocol. Further work is
continuing to improve T2-DIADEM signal dropout and image artifacts.Acknowledgements
This work was supported by NIH U01 EB024450. The
authors would like to thank Jennifer Myers and Erin Gray for their help in
collecting the data.References
1. Tan ET, Lee SK, Weavers PT, Graziani D,
Piel JE, Shu Y, Huston J, Bernstein MA, Foo TK. High slew‐rate head‐only
gradient for improving distortion in echo planar imaging: Preliminary experience.
Journal of Magnetic Resonance Imaging 2016;44(3):653-664.
2. Lee
SK, Mathieu JB, Graziani D, Piel J, Budesheim E, Fiveland E, Hardy CJ, Tan ET,
Amm B, Foo TKF. Peripheral nerve stimulation characteristics of an asymmetric
head‐only gradient coil compatible with a high‐channel‐count receiver array.
Magnetic resonance in medicine 2016;76(6):1939-1950.
3. Tan
ET, Weavers PT, Gray EM, Hardy CJ, Schenck JF, Bernstein MA, Shu Y, Foo TK,
Huston J. ISMRM 2017:p.234.
4. In MH,
Posnansky O, Speck O. High-resolution distortion-free diffusion imaging using
hybrid spin-warp and echo-planar PSF-encoding approach. NeuroImage
2017;148:20-30.
5. In MH,
Tan ET, Trzasko JD, Shu Y, Kang D, Yarach U, Tao S, Gray EM, Huston III J,
Bernstein MA. Distortion‐free imaging: A double encoding method (DIADEM)
combined with multiband imaging for rapid distortion‐free high‐resolution
diffusion imaging on a compact 3T with high‐performance gradients. Journal of
Magnetic Resonance Imaging 2020;51(1):296-310.
6. Dong
Z, Wang F, Reese TG, Manhard MK, Bilgic B, Wald LL, Guo H, Setsompop K.
Tilted‐CAIPI for highly accelerated distortion‐free EPI with point spread
function (PSF) encoding. Magnetic resonance in medicine 2018.
7. Weavers
PT, Shu Y, Tao S, Huston J, Lee SK, Graziani D, Mathieu JB, Trzasko JD, Foo
TKF, Bernstein MA. Compact three‐tesla magnetic resonance imager with
high‐performance gradients passes ACR image quality and acoustic noise tests.
Medical physics 2016;43(3):1259-1264.
8. Foo
TKF, Laskaris E, Vermilyea M, Xu M, Thompson P, Conte G, Van Epps C, Immer C,
Lee SK, Tan ET, Graziani D, Mathieu JB, Hardy CJ, Schenck JF, Fiveland E,
Stautner W, Ricci J, Piel J, Park K, Hua Y, Bai Y, Kagan A, Stanley D, Weavers
PT, Gray E, Shu Y, Frick MA, Campeau NG, Trzasko J, Huston J, 3rd, Bernstein
MA. Lightweight, compact, and high-performance 3T MR system for imaging the
brain and extremities. Magn Reson Med 2018;80(5):2232-2245.