David Marlevi1, Jonas Schollenberger2, Maria Aristova3, Edward Ferdian4, Alistair A Young4,5, Elazer R Edelman1, Susanne Schnell3,6, C. Alberto Figueroa2, and David Nordsletten2,5
1Massachusetts Institute of Technology, Cambridge, MA, United States, 2University of Michigan, Ann Arbor, MI, United States, 3Northwestern University, Chicago, IL, United States, 4University of Auckland, Auckland, New Zealand, 5King's College London, London, United Kingdom, 6University of Greifswald, Greifswald, Germany
Synopsis
4D Flow MRI images cerebrovascular blood flow in-vivo, however, estimation of relative
pressure is difficult due to the unique flow and anatomies found in the brain. We
evaluated the performance of three different techniques (reduced Bernoulli (RB);
unsteady Bernoulli (UB); virtual Work-Energy Relative Pressure (vWERP)) for cerebrovascular assessment.
Using patient-specific in-silico models,
we show that accurate estimations are dependent on sufficient spatial
resolution (dx < 0.75 mm3) and used approach (vWERP achieving accurate estimates; RB/UB
showing systematic underestimation bias). With similar dependencies indicated in-vivo, these results underline both
potentials and challenges of mapping cerebrovascular relative pressure by 4D
Flow MRI.
Introduction
Changes in regional blood pressure are indicated
in a number of cerebrovascular disease1,2, and 4D Flow MRI allows
for full-field capture of blood flow throughout the cerebrovasculature3.
However, derivation of relative pressure from 4D Flow MRI is challenging as
unique flow and anatomies preclude accurate translation of estimation techniques
used in other cardiovascular domains. Though attempts have been made to assess
cerebrovascular pressure changes4,5, a systematic evaluation of
image-based techniques has yet to be performed. Therefore, the aim of this
study was to investigate the accuracy of three different methods to assess cerebrovascular
relative pressure from 4D Flow MRI data: the reduced Bernoulli (RB)6,
the extended unsteady Bernoulli (UB)7, and the full-field virtual
Work-Energy Relative Pressure (vWERP)
method8.Methods
The three estimation
approaches all originate from the Navier-Stokes equations, however, different
assumptions on the assessed flow field are imposed when extracting relative
pressure: RB assumes negligible transient and viscous effects and extracts
estimates from peak velocity measurements6; UB includes transient
effects but still reduces estimations to an integration line between estimated
inlet and outlet7; vWERP
does not impose any assumptions on the assessed flow, but rather transfers these
onto an auxiliary virtual field introduced through a virtual work-energy
evaluation8. As such, the three approaches can be expressed as
$$
\Delta p_{RB}=\frac{1}{2}\rho(v_o^2-v_i^2 ); \quad \Delta p_{UB}=\frac{1}{2}\rho(v_o^2-v_i^2 )+\rho\int_i^o \frac{\partial \mathbf{v}}{\partial t} ds; \quad \Delta p_{\nu WERP}=-\frac{1}{Q_e} (\frac{\partial}{\partial t} K_e+A_e+V_e )
$$
where $$$\Delta p$$$ is the relative pressure estimate, $$$v_{i,o}$$$ is the velocity at inlet and outlet, $$$\mathbf{s}$$$ is an integration line connecting the inlet
and outlet, $$$\rho$$$ is the fluid density, and $$$K_e$$$, $$$A_e$$$, $$$V_e$$$, and
$$$Q_e$$$ are different virtual energy and flow
components extracted during computation of vWERP8.
To study estimation accuracy
in a controlled environment, synthetic 4D flow MRI images were simulated from a
calibrated patient-specific computational fluid dynamics (CFD) model of the
arterial cererbrovasculature (see Figure 1(left)). The model was created by
reconstructing the anatomy from TOF-MRI and specifying patient-specific
boundary conditions using a combination of phase-contrast-MRI and arterial spin
labelling (ASL) to recover realistic flow and pressure fields9,10.
Images were generated at increasing spatiotemporal resolution, covering dx =
0.25-1.0 mm3, and dt = 20-80 ms, respectively. In all models,
relative pressures over the internal carotid artery (ICA), as well as over a
transition into the middle cerebral artery (ICA-MCA) were estimated using RB,
UB, and vWERP, respectively, and
compared against reference CFD pressure data.
Furthermore, to evaluate
spatial dependencies in-vivo, 4D Flow
MRI (Siemens Magnotom Skyra, 3T, 20-channel head/neck coil, k-t GRAPPA accelerated dual venc (120/60
cm/s) sequence11) was collected from 8 healthy volunteers (see
Figure 1(right)) at two different spatial resolutions: dx = 1.1 and 0.8 mm3.
Cerebrovascular relative pressures were estimated over an ICA-MCA section using
RB, UB, and vWERP, respectively. Results
Figure 2 shows linear
regression plots for in-silico relative
pressures estimated from RB, UB, and vWERP,
respectively, compared against reference relative pressure extracted from the
CFD data. As shown, estimation accuracy was highly dependent on spatial
resolution and estimation approach: at dx = 1.0 mm3, underestimation
bias was apparent for all methods (regression coefficient k = 0.44, 0.46, and
0.55 for RB, UB, and vWERP,
respectively). As resolution increased, vWERP
accuracy increased (k = 0.98 at dx = 0.5 mm3), however,
underestimation bias remained for RB and UB (k = 0.42 and 0.46 at dx = 0.5 mm3).
Notably, the same dependency on resolution and approach was not observed with
varying temporal sampling.
No apparent correlation was observed
between the in-vivo estimates at dx =
1.1 and 0.8 mm3 for RB or UB (k = 0.05 and 0.07, Fig. 3), whereas vWERP indicates stronger correlation as
well as lower estimates extracted at lower resolutions (k = 0.52, Fig. 3). Discussion
Based on calibrated in-silico data, vWERP provides
the most accurate estimates of cerebrovascular relative pressure, overcoming
the underestimation bias observed in RB and UB. Evidently, assumptions used in
RB and UB do not translate well into the evaluated cerebrovascular flows.
Further, correlative trends between estimates and true relative pressures
varies over different vascular sections (Fig. 2), making their use for
diagnostic purposes non-trivial. While vWERP
avoids any flow field assumptions, it is still impacted by spatial resolution: at
dx = 1 mm3, vWERP exhibited
underestimation bias, indicating that spatial flow and velocity gradients are not
sufficiently captured through the tortuous vasculature. The correlation between
1.1 and 0.8 mm3 data in-vivo
also corroborates the clinical plausibility of these findings. High-Tesla
acquisition12 or super-resolution imaging13 could
represent practical solutions for quantitative cerebrovascular 4D Flow MRI in the
near-future. Emphasis should still be made on spatial resolution when probing
cerebrovascular hemodynamics. Work remains on in-vivo validation and clinical evaluation of our results, however,
our study underlines both opportunities and challenges associated with
quantitative cerebrovascular 4D Flow MRI.Conclusion
Accurate estimates of
cerebrovascular relative pressure can be achieved by 4D Flow MRI. However,
output is strongly dependent on image resolution (dx < 0.75 mm3
indicated through the circle of Willis) and estimation approach (vWERP outperforming alternative
Bernoulli-methods). With similar indications observed in-vivo, clinical implications are highlighted, with our data
clarifying both possibilities and challenges associated with image-based
cerebrovascular relative pressure mapping.Acknowledgements
No acknowledgement found.References
1.
Leng X, Wong KS, Liebeskind DS. Evaluating intracranial atherosclerosis rather
than intracranial stenosis. Stroke. 2014:45:645-651
2.
Penn DL, Komotar RJ, Connolly ES. Hemodynamic mechanisms underlying cerebral
aneurysm pathogenesis. J Clinical Neuroscience. 2011:18:1435-1438
3.
Stankovic Z, Allen BD, Garcia J, Jarvis KB, Markl M. 4D flow imaging with MRI.
Cardiovascular diagnosis and therapy. 2014:173
4.
Rivera-Rivera LA, Johnson KM, Wieben O. Pressure mapping and hemodynamic
assessment of intracranial dural sinuses and dural arteriovenous fistulas with
4D flow MRI. Am J Neuroradiology. 2018:39:485-487
5.
Vali A, Aristova M, Vakil P, et al. Semi-automated analysis of 4D flow MRI to
assess the hemodynamic impact of intracranial atherosclerotic disease. Magn Res
Med. 2019:82:749-762
6.
Stamm RB, Martin RP. Quantification of pressure gradients across stenotic
valves by Doppler ultrasound. JACC. 1983:707-718
7.
Firstenberg MS, Vandervoort PM, Greenberg NL, et. al. Noninvasive estimation of
transmitral pressure drop across the normal mitral valve in humans: importance
of convective and inertial forces during left ventricular filling. JACC. 2000:36:1942-1949
8. Marlevi D, Ruijsink B, Balmus M, Dillon-Murphy D, Fovargue D et al.
Non-invasive pressure estimations by virtual fields – cardiovascular pressure
drops from 4D flow MRI. ISMRM 2018, Paris, France.
9.
Schollenberger J, Hernandez-Garcia L, Figueroa CA. Calibration of
patient-specific computational models of cerebral blood flow in cerebrovascular
disease using arterial spin labeling. ISMRM, 2020
10. Schollenberger J, Osborne NH, Hernandez-Garcia L, Figueroa CA. A combined computational
fluid dynamics and MRI arterial spin labeling modeling strategy to quantify
patient-specific cererbal hemodynamics in cerebrovascular occlusive disease. in-preparation
11. Schnell S, Ansari A, Wu C, et al. Accelarated dual-venc 4D flow MRI
for neurovascular applications. JMRI. 2017:46:102-114
12.
Gottwald L, Töger J, Block KM, et al. High spatiotemporal resolution 4D flow
MRI of intracranial aneurysms at 7T in 10 minutes. Am J Neuroradiology.
2020:41:1201-1208
13.
Ferdian E, Suinesiaputra A, Dubowitz DJ et al. 4DFlowNet: Super-resolution 4D
Flow MRI using deep learning and computational fluid dynamics. Frontiers in
Physics. 2020:8:138