Jihun Kwon1, Masami Yoneyama1, Takashige Yoshida2, Kohei Yuda2, Yuki Furukawa2, Johannes M. Peeters3, and Marc Van Cauteren3
1Philips Japan, Tokyo, Japan, 2Tokyo Metropolitan Police Hospital, Nakano, Japan, 3Philips Healthcare, Best, Netherlands
Synopsis
Shoulder MRI is typically acquired with multiple number of
signals averaged (NSA) in order to average out breathing motion artifacts.
However, higher NSA leads to a longer scan time and patient discomfort. In this
study, we investigated the use of a deep learning-based reconstruction
algorithm to highly accelerate shoulder MRI. Adaptive-CS-Net,
a deep neural network previously introduced at the 2019 fastMRI challenge, was
expanded and presented here as a Compressed-SENSE Artificial Intelligence (CS-AI)
reconstruction. The purpose of this study was to compare the image quality of
shoulder MRI between reference and accelerated methods; SENSE,
Compressed-SENSE, and CS-AI.
Introduction
Shoulder MRI is useful for assessment of shoulder
disability and pain1. Typically, shoulder MRI is acquired with multiple number of signals
averaged (NSA) in order to average out breathing motion artifacts over the
course of the scan. However, higher NSA leads to a longer scan time and patient
discomfort. Recently, integrating artificial intelligence (AI) into the MRI
reconstruction has attracted much attention. This allows to further minimize
scan time2. At the 2019 fastMRI challenge, a novel deep neural network was
introduced as Adaptive-CS-Net and showed superior performance for reconstructing
knee images from highly undersampled k-space data3–6. The Adaptive-CS-Net was expanded to multiple contrasts and anatomical
areas and is presented here as Compressed SENSE AI (CS-AI) reconstruction. It
is hypothesized that the acquisition time for shoulder MRI may be significantly
shortened while maintaining image quality by using the CS-AI reconstruction
algorithm. The purpose of this study was to acquire highly accelerated shoulder
MRI using the CS-AI reconstruction and compare the image quality with the conventional
method, SENSE, and compressed-SENSE (C-SENSE).Methods
The study was approved by the local IRB,
and written informed consent was obtained from all subjects. A total of 2
volunteers were examined on a 3.0T whole-body clinical system (Ingenia Elition
X, Philips Healthcare) using an 8-channel shoulder coil. Four multi-slice 2D sequences;
proton density weighted (PDw), fat-suppressed PDw, T2-weighted (T2w), and fat-suppressed
T2w were all acquired in the coronal and axial plane.
The image quality was compared qualitatively
between the reference method (commonly used sequence in the clinic based on
C-SENSE) and accelerated methods: SENSE, C-SENSE, and CS-AI. The reference scan
was acquired with C-SENSE acceleration factor=2 and NSA=2. The total scan time
in the reference method for PDw, fat-suppressed PDw, T2w, and fat-suppressed T2w
was 2:10min, 2:10min, 2:08min, and 1:51min, respectively. For the accelerated scans,
data was acquired in two breath holds with acceleration factor=4 and NSA=1. The
total scan time for PDw, fat-suppressed PDw, T2w, and fat-suppressed T2w was 40.0sec,
40.0sec, 38.9sec, and 35.2sec, respectively. The TR/TE for PDw, fat-suppressed PDw,
T2w, and fat-suppressed T2w were 2500/30, 2500/30, 2778.8/80, and 3000/65,
respectively. Following parameters were common to all examinations: FOV=160×160mm,
22 slices, voxel size=0.5×0.7×3mm, FA=90.
The CS-AI model used in this study is the
extension of the previously introduced AI-based reconstruction algorithm,
Adaptive-CS-Net5,6. In CS-AI, the iterative optimization procedure in the C-SENSE
reconstruction chain is unrolled for a fixed number of reconstruction blocks.
Each block consists mainly of Unet-like architecture, which performs as a
denoiser. The model was trained on more than 700,000 images, including 2D and
3D data, and multiple contrasts and anatomical areas.Results and Discussions
Overall, the CS-AI images demonstrated high
denoising performance. In all sequences, the CS-AI showed image quality comparable
to that of the reference method, which took more than a factor of 3 longer than
the accelerated scans.
Figure 1 shows the comparison of the
reference, SENSE, C-SENSE, and CS-AI for PDw images. The reference method and
CS-AI showed less noise than other methods. The CS-AI produced visually sharper
images compared to SENSE and C-SENSE.
Figure 2 shows the comparison of the
reference, SENSE, C-SENSE, and CS-AI for fat-suppressed PDw images. Both the reference
and CS-AI showed good image quality, but the CS-AI demonstrated a better depiction
of small structures such as suprascapular nerve (arrow). This suggests that the
scan acceleration with the noise reduction by CS-AI reconstruction may be
useful for getting a high-quality image by minimizing the breathing motion.
Figure 3 shows the comparison of the
reference, SENSE, C-SENSE, and CS-AI for T2w images. The reference method
showed an ambiguous appearance of edges compared to other methods likely owing
to the prolonged scan time. The C-SENSE and CS-AI methods showed less noise
compared to SENSE. Small structures may be better visualized in CS-AI than
C-SENSE.
Figure 4 shows the comparison of the
reference, SENSE, C-SENSE, and CS-AI for fat-suppressed T2w images. The SENSE
and C-SENSE images became very noisy and were not suitable for diagnosis. However,
the CS-AI significantly reduced the noise and had a noise level comparable with
the reference method. Due to the reduced noise, structures with high signal such
as synovial fluid (arrow) were well depicted in the CS-AI as well as the reference.Conclusion
In this study, we compared qualitatively the
image quality of shoulder MRI between reference, SENSE, C-SENSE, and CS-AI
reconstruction methods. Our results suggest that the CS-AI may be able to provide
the image quality equivalent to the reference method in a significantly shorter
acquisition time. The CS-AI allows to do regular shoulder MRI scans in 2 breath
holds and this can be a preferable approach over the conventional averaging
method to avoid motion artifacts. Further data collection is required to
determine the advantage of CS-AI over other scan acceleration techniques.Acknowledgements
No acknowledgement found.References
1.
Tashjian RZ. Epidemiology, Natural
History, and Indications for Treatment of Rotator Cuff Tears. Clin Sports Med. 2012;31(4):589-604.
2. Hammernik K, Klatzer T, Kobler E, et al.
Learning a variational network for reconstruction of accelerated MRI data.
Magn Reson Med. 2018;79(6):3055-3071.
3. Knoll F, Zbontar J, Sriram A, et al.
fastMRI: A Publicly Available Raw k-Space and DICOM Dataset of Knee Images for
Accelerated MR Image Reconstruction Using Machine Learning. Radiol Artif
Intell. 2020;2(1):e190007.
4. Knoll F, Murrell T, Sriram A, et al.
Advancing machine learning for MR image reconstruction with an open
competition: Overview of the 2019 fastMRI challenge. Magn Reson Med.
2020;(January):mrm.28338.
5. Pezzotti N, Yousefi S, Elmahdy MS, et
al. An Adaptive Intelligence Algorithm for Undersampled Knee MRI
Reconstruction. IEEE Access. 2020;8:204825-204838.
6. Pezzotti N, de Weerdt E, Yousefi S, et
al. Adaptive-CS-Net: FastMRI with Adaptive Intelligence. arxiv.
2019;(NeurIPS).