Daniel Löwen1, Eberhard Daniel Pracht1, Rüdiger Stirnberg1, and Tony Stöcker1,2
1German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany, 2Department of Physics and Astronomy, University of Bonn, Bonn, Germany
Synopsis
Achieving homogeneous fat suppression is important for a wide
range of MRI applications. In this work, we present time-efficient
water-selective, parallel transmit RF excitation pulses for
ultra-high field applications. This method combines the properties of
kT-points with water-selective binomial pulses to achieve short, B1
insensitive water excitation pulses.
Introduction
Homogenization of
the image contrast plays a key role at ultra-high field (7T).
Different approaches have been proposed and parallel transmission
(pTx) has turned out to be particularly effective. However, the
combination of pTx homogenization and spectral selective excitation
increases the pulse duration and therefore limits the echo and/or the
repetition time of fast steady-state imaging sequences.
To achieve high
image homogeneity and spectral selection while maintaining short
RF-pulse durations, we propose interleaved binomial kT-points pulses.
The basic approach is similar to the work described in [1] but
extended to dynamic pTx pulses, leveraging the full potential of
multichannel transmission.Methods
Image homogenization
was achieved by employing non-selective kT-points pulses
[2]. To introduce water-selection into kT-points, we applied
the concept of binomial (1-1) pulses [3].
The most obvious
approach is to repeat the kT-points pulse when fat and water protons
are out of phase to tip the fat magnetization back to the
longitudinal axis, see Figure 1 (middle: binomial classic). If the
total pulse length is longer than half the precession period of fat
(Tfat/2) the second pulse has to be applied at an odd multiple of Tfat/2 later, leading to overly long pulse durations and
significantly increased B0 sensitivity. However, due to the composite
nature of kT-points, the pulse can be split into sub-pulse blocks
shorter than Tfat/2: The first sub-pulse block is repeated directly
after Tfat/2, immediately followed by the second sub-pulse block
played out in the same fashion, and so forth (Figure 1 bottom).
For comparison, a
previously published kT-points based water excitation technique was
implemented [4,5]. Here, each sub-pulse duration is adjusted to suppress fat (~1ms at 7T), see Figure 1 (top: RECT).
Therefore , the only degree of freedom is the number of kT-points,
leading to a compromise between homogenization and total RF-pulse
duration.
Bloch simulations of
the interleaved binomial kT-points pulse were carried out and
compared to the RECT pulse by means of duration, homogeneity, fat
suppression, peak voltage and pulse energy.
The
pulse design was implemented on a MAGNETOM 7T scanner (Siemens AG,
Healthcare Sector, Erlangen, Germany). For validation, in vivo MPRAGE
data were acquired, using a head array coil with 32 receive and 8
transmit channels (Nova Medical, Wilmington, USA) using simple CP
mode pulses and kT-points pulses (7 sub-pulses) without and with
water-selectivity. Imaging parameters:
TI
= 1100 ms (adiabatic CP-based inversion), TE/TR: 4.4 / 2500 ms,
turbofactor 176, 1mm3
isotropic resolution. The sequence
utilizes
an optimized linear reordering scheme
with elliptical scanning and 2D
parallel imaging (CAIPI 2x2z1)
[6].The
total scan
time was 2:40 min.Results
Simulation results
of a few selected example pulses are listed in Table 1. In order to
get a good compromise between water-selection and homogenization for
the RECT water excitation, the number of kT-points was chosen to be
3, as in [5], resulting in 3.22ms total duration.
The interleaved
binomial kT-points pulses were optimized to obtain shorter total
durations. This was achieved by varying the number of kT-points and
the sub-pulse duration. If shorter binomial sub-pulse durations were
used, the number of kT-points could be increased and hence better
homogeneity was achieved, while still maintaining shorter total pulse
durations. However, shorter sub-pulses increase peak voltages and
total energies.
Fig. 2 shows flip
angle maps of a phantom slice calculated in Bloch simulations for
water (top) and fat (bottom) using representative pulses from Table
1. For the fat simulation, the B0 map was shifted to the fat
frequency to simulate a “fat-only” phantom. Using the proposed
interleaved binomial kT-points pulses, the mean fat flip angle
increases only slightly compared to the RECT kT-points pulse, while
water is excited more homogeneously.
The effectiveness of
interleaved binomial kT-points pulses could be demonstrated in vivo
and is shown in Fig. 3. The images acquired in CP mode (top) clearly
lack intensity in the cerebellum whereas the interleaved kT-points
pulse achieves good homogeneity. Additionally, the fat signal is
effectively suppressed for the interleaved binomial kT-points pulses
(bottom right), as predicted by the Bloch simulations.Discussion/Conclusion
In this work, we
demonstrate a time efficient combination of B1-homogenization and
water excitation for ultra-high field applications. In contrast to
the RECT kT-points method, interleaved binomial kT-points add another
degree of freedom to the pulse design to obtain good image
homogeneity even with short water-selective pulses. SAR can be
controlled by choosing appropriate sub-pulse durations and number of
kT-points.
Future prospects
include the extension to binomial 1-2-1 or higher order binomial
pulses, which are less sensitive to B0 inhomogeneities, and thus lead
to more robust fat suppression.Acknowledgements
No acknowledgement found.References
[1] Malik S. J. et
al.: MRM 2010;63:988-997
[2] Cloos M. A. et
al.: MRM 2012;67:72–80
[3]
Hore P. J.: JMR 1983;54:539-542
[4]
Stirnberg R., et al: MRM 2016;76:1517-1523
[5]
Le Ster C., et al. PLoS ONE 2019;14:e0225286
[6]
Brenner, D, et al. Magma (New York, N.Y.), 27(5),
455–462.