Munish Chauhan1, Saurav ZK Sajib1, Sulagna Sahu1, Enock Boakye1, Willard S Kasoff2, and Rosalind J Sadleir1
1School of Biological and Health System Engineering, Arizona State University, Tempe, AZ, United States, 2Department of Surgery, University of Arizona, Tucson, AZ, United States
Synopsis
Knowledge
of electromagnetic field distribution parameters such as current density,
conductivity, and electric field distribution may play an essential role in
understanding DBS therapeutic effects on living tissues. However, until now
there has been no experimental method that can measure these field parameters. In
this work, we demonstrate that it is possible to quantify electromagnetic field
distribution during DBS therapy using magnetic resonance electrical impedance
tomography (MREIT).
Introduction
Magnetic
resonance current density imaging (MRCDI)1 and magnetic resonance
electrical impedance tomography (MREIT)2 are two emerging techniques
that can produce low-frequency current density and conductivity distributions,
respectively using MR measurements of the z-component
of the magnetic flux density (Bz) data induced due to external
current injection. While the current density can be straightforwardly estimated
using a single set of Bz data, the stable
reconstruction of the conductivity distribution using MREIT requires data from two
independent current administrations2. Administration of two
independent currents are not normally possible in DBS settings.
MRCDI has already been used to demonstrate current
pathways during deep brain stimulation (DBS)3,4 therapy. However, it
is not possible to find the corresponding electric field distributions due to the
lack of a stable conductivity reconstruction method. Recently, Song et al.5 developed an iterative single-current algorithm that can stably reconstruct the ‘apparent’
conductivity distribution from data obtained using a single current administration
by converting the first-order hyperbolic partial differential equation (PDE) used
in the Harmonic Bz MREIT reconstruction algorithm2
into a second-order elliptical-PDE. The aim of this study is to demonstrate the
capability of the new single-current harmonic Bz algorithm to characterize electromagnetic field
distributions produced by DBS electrodes.Methods
Phantom design: A head-shaped phantom formed of agarose-gel was used in this study4.
At 10 Hz, the gel conductivity was measured to be 0.75 S/m. A directional DBS
electrode (Abbott Infinity 6172, Abbott, Abbott Park, IL, USA) was inserted near
the phantom center (Fig.1b). A 50
x 50 mm2 carbon return electrode (HUREV Co Ltd, South Korea) was also
attached at the occipital (Oz) location (Fig.1b). We placed a piece of chicken muscle
between DBS lead and Oz electrode (Fig.1b) to form a conductivity contrast. An MR safe electrical stimulator
(DC-STIMULATOR MR, neuroConn, Ilmenau, Germany) was used to deliver a 1.0 mA
current between the DBS electrode contact combinations and the return Oz
electrode. In the DBS-1-Oz setting, the omni-directional electrode contact-1
(Fig.1a) of the electrode and Oz were used, and the DBS-2B-Oz setting denoted the use of the sectional electrode contact-2B (Fig.1a) facing towards the Oz
direction. This allowed evaluation of the differences between electric fields
produced using conventional omnidirectional DBS electrodes and newer
directional DBS electrode designs.
Magnetic
flux density measurement: All
MR data were measured using a 32-channel RF coil in a 3.0T Phillips scanner
(Phillips, Ingenia, Netherlands) located at the Barrow Neurological Institute
(Phoenix, Arizona, USA). Current induced Bz data was measured at three axial slice positions
using the Phillips mFFE pulse
sequence with an imaging matrix size of 128 x 128
and field-of-view 224x224 mm2.
Other imaging parameters were TR/TE = 50/7 ms, number of echoes = 10,
echo spacing = 3 ms, number of averages = 24. The total scan time was 6 min. A
set of T1-weighted images with matrix size, 224x224x224 resulting in 1 mm3
isotropic resolution, were also collected during the imaging session to build a
numerical model of the phantom and wiring6. Prior to estimating the
electromagnetic field parameters, the echo combined7 optimized Bz
images were corrected to account for wire-created stray magnetic-fields (Fig 1e-f).
Reconstruction of electromagnetic parameters: We first used the stray-field-corrected
Bz data to reconstruct the current density distribution, $$$ \mathbf{\mbox{J}}^{P,R}$$$,
within the local region $$$R$$$,
(Fig. 1d) using the regional projected current
density algorithm8. From the $$$ \mathbf{\mbox{J}}^{P,R}$$$,
and the
known conductivity, ($$$\sigma_{b}$$$), at the tissue boundary $$$\partial R$$$, the
first-update of the internal conductivity distribution, $$$\sigma^1$$$
was
reconstructed as5
$$ \begin{cases}\triangledown_{xy}^2 ln\sigma^{1}=\frac{1}{\mu_{0}}\triangledown_{xy} .\Bigg( -\frac{J_y^{P,R}}{(J_x^{P,R})^{2}+{(J_y^{P,R})^{2}}}\triangledown_{xy}^2B_{z},~\frac{J_x^{P,R}}{(J_x^{P,R})^{2}+{(J_y^{P,R})^{2}}}\triangledown_{xy}^2B_{z} \Bigg)~~~~~~in~~~ R~~~~~~~~~~~~~~~~~~~~~~~(1)\\\\~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ln \sigma^{1}=ln \sigma_{b}~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~on ~~\partial R \end{cases}$$
where
$$$\mu_{0}$$$
denotes
the permeability in free space. We set $$$\sigma_{b}$$$ =0.75 S/m.
The
corresponding electric field distribution was then found using Ohm’s law
$$\mathbf{\mbox{E}}=\frac{\mathbf{\mbox{J}}^{P,R}}{\sigma}~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~(2)$$ Results
Figs.
2a (i) and (iv) show the reconstructed $$$ \mid \mathbf{\mbox{J}}^{P,R} \mid $$$distribution due to the 1.0 mA current
injected using the DBS-1-Oz
and DBS-2B-Oz electrode configurations, respectively. Reconstructed
$$$\sigma^{1}$$$ distributions measured using the DBS-1-Oz and
DBS-2B-Oz configurations are displayed in Figs. 2b (ii) and (v), respectively. The
corresponding electric field magnitudes found using Eq. 2 are shown in Fig 2c
(iii) and (vi). All reconstructed electromagnetic field parameters within the
chicken muscle and agar background are summarized in the table in Fig 3. Discussion
As noted in5,
the first update to the iterative reconstruction method $$$\sigma^{1}$$$ is sufficient
to produce a result comparable with the two-current-injection non-iterative
harmonic Bz algorithm9. Therefore, in this study, we only use the
first-updated to the conductivity distribution to determine electromagnetic
field parameters. In the case of the reconstructed $$$ \mid \mathbf{\mbox{J}}^{P,R} \mid$$$
and $$$\bf \mid E \mid$$$ fields (Fig. 3) we also found high standard deviations in the background
agar region, which stems from the fact that most of the field distribution is
concentrated nearby the DBS electrode region (Fig. 2(a), (c)). However, as
expected this high field nearby the DBS electrode does not affect the
reconstructed conductivity in the agar region (Fig. 3b).Conclusion
We
have demonstrated that it is possible to characterize electromagnetic
properties and fields deep within the brain using DBS electrode illumination using
a biological tissue phantom. Acknowledgements
This work was supported by
award RF1MH114290 to RJS.References
1. Scott G C, Joy M L G and Armstrong R L et al.
Measurement of nonuniform current density by magnetic resonance. IEEE Trans. Med.
Imaging. 1991; 10(3): 362-74.
2. Seo J K and Woo E J. Magnetic resonance electrical
impedance tomography. SIAM Rev. 2011;53(1):40-68.
3. Sajib S Z K, Oh T I, and Kim H J et al. In vivo
mapping of current density distribution in brain tissues during deep brain
stimulation (DBS). AIP Adv. 2017; 7(1):015004.
4. Chauhan M, Sajib S Z K and Sahu S et al. Imaging of
current density distribution in deep brain stimulation (DBS). Proc. Intl. Soc.
Mag. Reson. Med. 2020; 28(2020):3180.
5. Song Y, Sajib S Z K and Wang H et al. Low frequency
conductivity reconstruction based on a single current injection via MREIT. Phys.
Med. Biol. 2020;65(22):225016.
6. Sajib S Z K, Chauhan M and Banan G et al. Compensation
of lead-wire field contributions in MREIT experiment using image segmentation:
a phantom study. Proc. Intl. Soc. Mag. Reson. Med. 2019; 27(2019):5049.
7. Kim M N, Ha T Y, Woo E J et al. Improved conductivity
reconstruction from muti-echo MREIT utilizing weighted voxel-specific
signal-to-noise ratios. Phys. Med. Biol. 2012; 57 (11):3643-59.
8. Sajib S Z K, Kim H J and Woo E J et al. Regional
absolute conductivity reconstruction using projected current density in MREIT.
Phys. Med. Biol. 2012;57(18):5841-59.
9. Seo J K, Jeon K and Lee C O et al. Non-iterative
harmonic Bz algorithm in MREIT. Inverse Prob. 2011; 27(8):1-12.