Julian Rauch1,2, Frederik B. Laun3, Theresa Palm3, Jan Martin3,4, Maxim Zaitsev5,6, Mark E. Ladd1,2,7, Peter Bachert1,2, and Tristan A. Kuder1
1Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Division of Physical Chemistry, Lund University, Lund, Sweden, 5Medical Physics, Department of Radiology, Faculty of Medicine, Medical Center University of Freiburg, Freiburg, Germany, 6High Field Magnetic Resonance Center, Center for Medical Physics and Biomedical Engineering Medical University of Vienna, Vienna, Austria, 7Faculty of Medicine, Heidelberg University, Heidelberg, Germany
Synopsis
Concomitant or Maxwell fields cause intravoxel
dephasing which can lead to strong image artifacts. In this study, we present a
new method for concomitant field correction in double diffusion encoding
sequences with single pairs of bipolar gradients on each axis. Additionally
implemented oscillating gradients remove the dephasing without changing the
desired image. Phase and magnitude images are analyzed with respect to
concomitant field induced artifacts and the proposed correction method. We show
that the compensation eliminates these artifacts without further consequences
for image quality. The method also may be included in other imaging sequences
to achieve concomitant field compensation.
Introduction
In diffusion-weighted MRI, strong gradients are usually
used to obtain information about diffusion processes within a sample. As a
consequence of Maxwell´s equations, these gradients generate concomitant or
Maxwell fields, which introduce an additional phase on the transverse magnetization which cannot be rephased as a gradient echo
[1, 2].
Diffusion schemes that use single pairs of bipolar
diffusion gradients on each axis, being a common double diffusion encoding
(DDE) approach [3], are prone to severe artifacts induced by concomitant fields
such as signal voids. Unlike the conventional diffusion encoding approach
proposed by Stejskal and Tanner [4], these DDE sequences are not intrinsically
compensated for intravoxel dephasing induced by concomitant fields.
In this study, we present a novel method for
concomitant field compensation in the context of DDE using additional
oscillating gradients.Methods
Experiments were carried out on a SIEMENS Prisma 3T
imaging system. A DDE sequence with EPI readout [5] was adapted for concomitant
field compensation (Figure 1).
Occurring Maxwell fields are given to lowest order by $$\mathrm{B_C(x,y,z,t)} = \frac{1}{2\mathrm{B_0}}(\mathrm{G_x^2z^2+G_y^2z^2+G_z^2\frac{x^2+y^2}{4}-G_xG_zxz-G_yG_zyz}).\qquad(1)$$
$$$\mathrm{B_0}$$$ is the static magnetic field which is modified by
the gradients $$$\mathrm{G_x}$$$, $$$\mathrm{G_y}$$$ and $$$\mathrm{G_z}$$$ along the physical scanner axes $$$\mathrm{x}$$$, $$$\mathrm{y}$$$ and $$$\mathrm{z}$$$, respectively. At the isocenter, the concomitant fields are zero.
The concomitant phase results from
$$\mathrm{\phi_C(x,y,z)=\gamma\int B_C(x,y,z,t)\,dt},\qquad(2)$$ where $$$\gamma$$$ is the gyromagnetic ratio.
Oscillating gradients were implemented such that the compensation condition
$$\mathrm{\int_0^{TE}s_n(t)\,G(t)^2\,dt =0}\qquad(3)$$ is fulfilled, with $$$\mathrm{TE}$$$ being the
echo time and $$$\mathrm{s_n(t) = (-1)^{n_{rf}(t)}}$$$, where $$$\mathrm{n_{rf}(t)}$$$ is the
number of refocusing pulses within time $$$\mathrm{t}$$$ and
$$$\mathrm{TE}$$$.
Since concomitant
phase compensation of one axis is done concurrently with diffusion encoding
along the second axis, theoretically no increase of the echo time is required.
A water phantom with a length of 30 cm was used
in all measurements. Coronal slices were acquired. The phantom center coincided
with the isocenter of the scanner. The field-of-view was 350 mm with a matrix
size of 128×128. Phase encoding direction was along the
$$$\mathrm{x}$$$-axis. The maximum gradient strength was $$$\mathrm{G_x}$$$ = 70 mT/m
for both diffusion and oscillating gradients. Durations of the diffusion-weighting gradients
were constant at $$$\mathrm{\delta}$$$ = 15.3 ms.
The compensation gradients had an oscillation period of 4.2 ms and 14
oscillations giving rise to a maximum $$$\mathrm{b}$$$-value of 14 s/mm2.
Acquired values for $$$\mathrm{q=(2\pi)^{-1}\gamma\delta G}$$$ ranged from 0 mm-1 to 46 mm-1 yielding
$$$\mathrm{b}$$$-values from 0 s/mm2 to 900 s/mm2 with the bipolar diffusion-weighting gradients. For this proof of concept
demonstration, diffusion-weighting and compensation gradients were only applied
along the
$$$\mathrm{x}$$$-axis.
To quantify the
effect on the image phase, measurements using one channel of the integrated Tx/Rx
body coil were carried out. Echo time was set to $$$\mathrm{TE}$$$ = 225 ms. Both uncompensated and compensated
images were acquired.
Magnitude images
were also acquired using a flexible 18-channel body array coil combined with
the spine array and the following parameters: Parallel imaging (GRAPPA, acceleration
factor 2), phase partial Fourier factor 5/8, receiver bandwidth 1954 Hz/pixel, $$$\mathrm{TE}$$$ = 150 ms.Results
The comparison of the phase behavior for both the
compensated and uncompensated acquisitions is shown in Figure 2. Phase values
were evaluated within a small region-of-interest (RoI). Without the oscillating
gradients, the phase exhibits a parabolic dependency on the
$$$\mathrm{q}$$$-value as predicted by Eqs. (1) and (2). A reference
curve based on the distance of the RoI from the isocenter is also depicted in
Figure 2. Experiment and theory show a high degree of accordance. With the
compensation, the phase remains mainly constant and no influence of concomitant
fields is observed.
Figure 3 shows the occurrence of the signal voids induced by the concomitant fields
in the elongated phantom by displaying representative signal profiles along the
$$$\mathrm{z}$$$-axis. The signal was corrected for diffusion attenuation
beforehand. As depicted in (a), beginning with $$$\mathrm{q}$$$ = 37 mm-1,
the displayed phantom profiles show severe signal attenuations. Due to the $$$\mathrm{z^2}$$$-dependency of the concomitant field (see Eq. (1)),
artifacts worsen with distance from the isocenter. As seen in (b), the
oscillating gradients remove the concomitant phase and no related artifacts are
visible.
In Figure 4, uncompensated and compensated magnitude images acquired with $$$\mathrm{q}$$$ = 46 mm-1 are shown together with a reference image ($$$\mathrm{q}$$$ = 0 mm-1).
The full image in (a) again highlights the artifacts which are removed in (b).Discussion and Conclusion
It could be shown that the usage of oscillating
gradients is an efficient tool for the removal of concomitant field induced
artifacts in the frame of DDE. Phase and magnitude were no longer affected by
the concomitant phase induced by the bipolar diffusion gradient pairs. The
oscillating gradients generated a negligible diffusion weighting of 1.6 % of
the acquired
$$$\mathrm{b}$$$-value. Thus, the desired diffusion image was well
conserved.
In order to null the resulting concomitant phase, a
sufficient number of gradient oscillations with adequate strength and duration
must be used which might exceed the stimulation limit.
Ongoing research includes the usage of concomitant
field compensation with simultaneous diffusion weighting on the same axis.
In conclusion, the proposed method can enhance image quality whenever large
gradient strengths and field-of-views cause concomitant field-related artifacts
in DDE.Acknowledgements
No acknowledgement found.References
[1] Bernstein MA
et al. Concomitant gradient terms in phase contrast MR: analysis and correction.
Magn Reson Med. 1998;39(2)
[2] Szczepankiewicz F
et al. Maxwell-compensated design of asymmetric gradient waveforms for tensor-valued
diffusion encoding. Magn Reson Med. 2019;82(4)
[3]
Shemesh N et al. Conventions and nomenclature for double diffusion encoding NMR
and MRI. Magn Reson Med. 2016;75(1)
[4] Stejskal EO, Tanner JE. Spin diffusion measurements: spin echoes in the
presence of a time-dependent field gradient. J Chem Phys. 1965;42(1)
[5]
Martin J et al. Contrast-to-noise ratio analysis of microscopic diffusion
anisotropy indices in q-space trajectory imaging. Z Med Phys. 2020;30(1)