Constantin von Deuster1,2, Stefan Sommer1,2, Christoph Germann3,4, Natalie Hinterholzer2, Robin M. Heidemann5, Reto Sutter3,4, and Daniel Nanz2,4
1Siemens Healthcare, Zurich, Switzerland, 2Swiss Center for Musculoskeletal Imaging (SCMI), Balgrist Campus, Zurich, Switzerland, 3Radiology, Balgrist University Hospital, Zurich, Switzerland, 4University of Zurich, Zurich, Switzerland, 5Siemens Healthcare, Erlangen, Germany
Synopsis
The large water-fat frequency difference at 7 T renders
imaging of the musculoskeletal (MSK) anatomy very challenging. In particular, through-slice
chemical-shift artifacts may manifest in state-of-the-art 2D turbo-spin-echo (TSE)
images as partial or locally complete fat-signal loss that radiologists are usually not
trained to account for from lower field strengths. In this work, we demonstrate
the range of possible through-slice artifacts in MSK images and show that matched
RF-pulse bandwidths as high as 1500 Hz for the excitation and refocusing RF-pulses
are necessary to consistently perform successful, non-fat suppressed MSK imaging
at 7 T.
Introduction
State-of-the-art Turbo-Spin-Echo (TSE) sequences commonly
employ a selection of unmatched and moderate RF-pulse bandwidths for the excitation
and refocusing pulses to suppress “third arm” artifacts1. However, different bandwidths result in varying chemical-shift
displacements along the slice-selection direction. Particularly at 7 Tesla (T),
this geometrical mismatch between the excitation and refocusing slice profiles
of fat may manifest as a relevant fat-signal loss. Additionally, moderate
RF-bandwidths can cause significant through-slice chemical-shift displacement
resulting in geometric inconsistencies between imaged water and fat slice. In
musculoskeletal applications2, these effects can cause false positive findings,
such as bone fractures or neoplastic bone-marrow disease3. The objective of this work was to explore and reduce
through-slice chemical-shift artifacts in TSE imaging at 7 T.Theory
A slice-selective RF-pulse affects fat and water spins
in different slices that are separated along the slice-selection dimension by $$$d$$$:$$ d = \Delta{z}\frac{\Delta f_{CS}}{\Delta f}$$ $$$\Delta{z}$$$ is the prescribed slice
thickness,$$$\Delta f_{CS} \approx$$$ 1040 Hz @ 7 T is the Larmor-frequency difference between fat and water and $$$\Delta{f}$$$ the RF-pulse bandwidth (Fig. 1). Materials and Methods
All experiments were performed on a 7 T whole-body MR
scanner (MAGNETOM Terra, Siemens Healthcare, Erlangen, Germany) equipped with a
1Tx/28Rx-channel knee coil (QED, Quality Electrodynamics, Mayfield Village, OH,
USA) for knee and foot imaging and a single-channel flex coil (RAPID Biomed,
Rimpar, Germany) for wrist imaging. Foot imaging is not the intended use for
the knee coil.
2D TSE imaging was performed with a standard sequence
utilizing (a) unmatched and low bandwidths (“ULB”), (b) unmatched and moderate bandwidths
(“UMB”) and (c) a prototype sequence with matched and increased bandwidths
(“MIB”). For the ULB, UMB and MIB case, the bandwidths of the excitation/refocusing
pulses were set to 750/469 Hz, 1095/682 Hz and 1500/1500 Hz, respectively. Both
RF-pulse types had a Hanning-filtered Sinc shape. Apart from the RF-pulse bandwidths,
imaging parameters were otherwise identical for each extremity (Fig. 2a).
T1-weighted sagittal knee
MRI with UMB and MIB was performed (i) in 10 healthy volunteers (mean age 31 ±
6 years, 6 male). Additional T1-weighted data were acquired from a single
subject (34 years, male) (ii) in coronal orientation in the knee (ULB, UMB, MIB),
(iii) in coronal orientation in the wrist (UMB, MIB), (iv) in axial orientation
in the foot (UMB, MIB). The
measurements were conducted in adherence with the federal law and written
informed consent was obtained prior to each examination from all subjects.
Chemical-shift
artifacts in the mostly affected sagittal knee images were measured for UMB and
MIB at two standardized locations at
the distal femoral bone-cartilage interface (Fig. 3a). The perceived negative
impact of through-slice chemical-shift artifacts on image quality in the bone
and posterior muscles were graded on a 5-point Likert4
scale (Fig. 2b). Objective and subjective artifact metrics were assessed by two
fellowship-trained musculoskeletal radiologists with 15- and 2-years’
experience in musculoskeletal MR imaging. Artifact size and impact on image
quality were compared between UMB and MIB using a two-way repeated measures ANOVA
and Wilcoxon test, respectively (SPSS, Version 25, IBM Inc, Armonk, NY). The corresponding
inter-reader agreement was assessed with the intra-class correlation
coefficient (ICC) and Cohen’s Kappa coefficient, respectively. Values for the
artifact size and impact on image quality are reported as mean ± standard
deviation and median ± interquartile range, respectively. A p-value < 0.05
was considered statistically significant.Results
Through-slice chemical-shift artifacts at the
bone-cartilage interface were statistically significantly smaller with MIB
compared to UMB: location 1: 0.35 ± 0.20 mm vs.
1.27 ± 0.27 mm, p <
0.001, location 2: 0.25 ± 0.13 mm vs. 1.48 ± 0.46 mm, p < 0.001, ICC: 0.98. The negative impact of chemical-shift artifacts on
image quality was rated statistically significantly smaller with MIB than with UMB:
bone: 2 ± 0 vs. 4 ± 1, p < 0.005 (both
readers); muscle: 3 ± 0 vs. 2 ± 0, p < 0.005 (both readers), Kappa: 0.69 (Figs. 3 d, e). UMB images
displayed multiple artifactual hyperintensities as well as significant
fat-signal loss in all examined anatomical regions. MIB images consistently showed
high image quality with bright T1-weighted fat-signal, excellent visual
resolution of fine tissue structures and improved image sharpness at several
bone-muscle, bone-fat and muscle-fat interfaces (Figs. 3 b, c, 4, 5). The specific
absorption rate for MIB was approximately doubled compared to UMB.Conclusion
Optimization of RF-pulse bandwidths to address
through-slice chemical-shift artifacts allowed consistently successful high-quality
concurrent fat- and water TSE imaging at 7 T. Through-slice chemical-shift artifacts
as well as their perceived negative impact on image quality were significantly
reduced while “third arm” artifacts1 could
not be observed in our measurements, using local transmit-receive coils. The
proposed sequence optimization has the potential to reduce the
likelihood of false readings in musculoskeletal applications at 7 T. Acknowledgements
This research project is partly funded by Balgrist
Campus AG, by the University Hospital Balgrist and by Siemens Healthcare AG:
CvD, SS and RMH are employees of Siemens Healthcare AG. References
1. Haacke EM, Brown RW, Thompson MR, et al. Magnetic Resonance Imaging: Physical
Principles and Sequence Design. 1st ed. New York: John Wiley & Sons;
1999.
2. Juras V, Mlynarik
V, Szomolanyi P, et al. Magnetic Resonance Imaging of the Musculoskeletal
System at 7T: Morphological Imaging and Beyond. Top. Magn. Reson. Imaging. 2019;28(3):125–135.
3. Singhal V, Bredella
MA. Marrow Adipose Tissue Imaging in Humans. Bone. 2019;118:69–76.
4. Jamieson S. Likert
scales: How to (ab)use them. Med. Educ.
2004;38(12):1217–1218.