Raphael Tomi-Tricot1,2,3, Jan Sedlacik2,3, Jonathan Endres4, Juergen Herrler5, Patrick Liebig6, Rene Gumbrecht6, Dieter Ritter6, Tom Wilkinson2,3, Pip Bridgen7, Sharon Giles7, Armin M. Nagel4, Joseph V. Hajnal2,3, Radhouene Neji1,2, and Shaihan J. Malik2,3
1MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 2Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 3Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 4Institute of Radiology, University Hospital Erlangen, Erlangen, Germany, 5Institute of Neuroradiology, University Hospital Erlangen, Erlangen, Germany, 6Siemens Healthcare GmbH, Erlangen, Germany, 7School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom
Synopsis
Direct Signal Control (DSC) uses parallel transmission (pTx)
with more flexibility than conventional static RF shimming to tackle RF
inhomogeneity at ultra-high field in fast spin echo (FSE) sequences by varying complex
weights of successive RF pulses independently along the refocusing train. Also,
unlike other dynamic pTx methods, it preserves RF pulse properties and sequence
timing. This work demonstrates the applicability of DSC in routine conditions
for neuroimaging, with minimal workflow disruption. In-vivo T2-weighted FSE
results exhibit higher signal and better homogeneity when using DSC over RF
shimming, while explicitly ensuring safe operation.
Introduction
Transmit radiofrequency (RF) field ($$$B_1^+$$$)
inhomogeneity is a challenge at 7T, especially for specific absorption rate
(SAR) demanding sequences such as 2D fast spin echo1 (FSE). Previously, this has
been tackled using parallel transmission and adjusting complex weights on each
channel2 (static RF shimming, [RFShim]), or by using combinations of RF and gradients
to target a specific flip angle (FA) (SPINS3, kT‑points4,5, spokes6). Alternatively, direct
signal control7,8 (DSC), which has previously
been experimentally applied to 2D9 and 3D7 FSE, leverages the degrees of
freedom offered by the multiple refocusing pulses in FSE to dynamically adjust
RF shimming and target homogeneous signal – with no impact on pulse shapes or
sequence timing. In this work we show first results of an optimised and fully
integrated implementation of DSC to 2D FSE brain imaging, under SAR control. Methods
Environment
All experiments were performed on a MAGNETOM Terra (Siemens Healthcare,
Erlangen, Germany) 7T scanner in research configuration, with an 8-Tx-channel
head coil (Nova Medical, Wilmington MA, USA). Human subject scanning was
approved by the Institutional Research Ethics Committee (HR-18/19-8700).
The scanner allowed for fully integrated
field mapping and pulse design, happening in the background as part of the
pre-acquisition adjustment steps. A FA map for each transmit channel was automatically
acquired at a 4.0x4.0x10.0 mm3 resolution using saturation-prepared
turbo FLASH10 (satprepTFL). Then, the scanner’s built-in pulse design environment was
launched in the background during sequence adjustments.
Imaging was performed under first
level SAR supervision on five healthy volunteers with the following T2w 2D
FSE protocol: TE/TR = 79/5500 ms; echo spacing: 9.88 ms; echo train length: 11;
576x365 matrix, 0.36x0.48 mm2 in-plane resolution; 34 transversal slices,
3 mm thickness; GRAPPA 3.
DSC Description and Comparison
DSC was assessed against RFShim
in a 2D FSE imaging experiment. For RFShim, excitation used a 90°
pulse and refocusing was performed under CPMG conditions11,12 at 135° (157.5° for
the first refocusing). A set of RF complex weights was determined and applied
to all pulses using the vendor’s $$$B_1^+$$$ shimming technique,
which solves a magnitude least-squares problem to minimise the spread of FA
values in the field of view, within 130% of circular polarisation (CP) mode
SAR.
DSC used an RFShim
as starting point for the optimisation, then further tuned complex RF weights
of each pulse independently to minimise signal differences from a target signal
evolution9, solving:
$$
\DeclareMathOperator*{\argmin}{arg\,min}\qquad\qquad\argmin_{W}\sum_t\sum_v\sum_e{c_e\left\Vert\frac{s_{t,v,e}(W)-\hat{s}_{t,v,e}}{\Vert\hat{s}_{t,v,e}\Vert_1}\right\Vert}_2^2\qquad\qquad(1)\\[2em]\qquad\qquad\textrm{s.t. hardware and physiological constraints}
$$with $$$t$$$, $$$v$$$ and $$$e$$$ indexing respectively tissues considered,
spatial locations and echo numbers in the refocusing train; $$$\hat{s}_{t,v,e}$$$ the target signal, $$$s_{t,v,e}$$$ the simulated signal depending on $$$W$$$ the matrix of complex weights applied to
different pulses and channels; $$$c_e$$$ a weighting coefficient adjusting the
contribution of each echo to the optimisation problem – here $$$c_e=0$$$ for early echoes, and $$$c_e$$$ linearly increases towards $$$e=8$$$ (k-space centre). $$$s_{t,v,e}$$$ and $$$\hat{s}_{t,v,e}$$$ were computed using the extended phase graph
representation8. To preserve contrast
two tissues were considered: grey matter (T1/T2 = 2000/100 ms) and cerebrospinal
fluid (T1/T2 =
4400/2000 ms). To accelerate the calculation time 150
voxels were picked for signal optimisation via k‑means clustering as shown in Figure 1(d). Analytical derivatives of the
cost-function and constraints were used. With a 50-iterations cap,
optimisation times were about 1 min.
$$$B_1^+$$$ Map Correction
Since DSC relies on signal calculations,
with errors propagating along the echo train, a prior investigation studied systematic
errors in satprepTFL related to its transient readout. The actual FA imaging
(AFI) sequence13,14 was found more
accurate but lengthier; it helped determine an experimental correction applied
to satprepTFL. Figure 1 shows a scatter plot exhibiting affine
relationship between the AFI and satprepTFL in a healthy volunteer. The
following empirical correction was applied to relative CP mode $$$B_1^+$$$ maps ($$$\dot{\alpha}_{\mathrm{measured}}$$$):
$$
\qquad\qquad\qquad\dot{\alpha}_{\mathrm{corrected}}=1.18\,\dot{\alpha}_{\mathrm{measured}}-0.14\qquad\qquad\qquad(2)
$$
The corrected CP maps ($$$\dot{\alpha}_{\mathrm{corrected}}$$$) were then back-projected into per-channel maps. The
quality and usefulness of the correction was assessed in simulation and in a
DSC measurement.
Quality Assessment
For result comparison, native images were brain-extracted
with FSL15–17 BET, then bias field
correction was applied with FSL FAST to reduce the effect of reception profile
(Figure
2). For a quantitative comparison between RFShim and DSC signals
(respectively $$$S_\mathrm{RFShim}$$$ and $$$S_\mathrm{DSC}$$$), relative signal
difference was calculated from brain-extracted images as:
$$
\qquad\qquad\qquad\qquad\delta{S}=\frac{S_\mathrm{DSC}-S_\mathrm{RFShim}}{S_\mathrm{RFShim}}\qquad\qquad\qquad\qquad(3)
$$Average values were reported for each subject.Results
Figure 3
highlights the importance of $$$B_1^+$$$ correction, especially
for DSC; this is visible in Figure 2
too. Figure 4
provides visual comparison of signal homogeneity and contrast in images
acquired with RFShim and DSC. DSC shows superior homogeneity in lower brain
areas, while preserving image quality in upper regions, under controlled SAR. This
is also confirmed quantitatively in Figure 5,
where DSC exhibits net signal gain over RFShim in all subjects, with average across subjects. Subject-averaged peak local
SAR was 94% (SD: 6%) of limit for RFShim and 98% (SD: 3%) for DSC. Conclusion
We have shown the effectiveness of DSC in providing better
imaging quality than RFShim for 2D T2w TSE in routine conditions. Future
developments will include better $$$B_1^+$$$ correction and measurement,
tackle more weightings and investigate slice-by-slice optimisation. Acknowledgements
The authors wish to thank Alexis Amadon and Franck
Mauconduit (NeuroSpin, CEA Paris-Saclay, Gif-sur-Yvette, France) for providing
the AFI sequence and reconstruction.
This work was supported by a Wellcome Trust collaboration in
science award [WT201526/Z/16/Z], by core funding from the Wellcome/EPSRC Centre
for Medical Engineering [WT203148/Z/16/Z]
and by the National Institute for Health Research (NIHR) Biomedical
Research Centre based at Guy’s and St Thomas’ NHS Foundation Trust and King’s
College London and/or the NIHR Clinical Research Facility. The views expressed
are those of the author(s) and not necessarily those of the NHS, the NIHR or
the Department of Health and Social Care.
References
1. Hennig J, Nauerth
A, Friedburg H. RARE imaging: A fast imaging method for clinical MR. Magn Reson
Med. 1986;3(6):823–33.
2. Padormo
F, Beqiri A, Hajnal JV, Malik SJ. Parallel transmission for ultrahigh-field
imaging. NMR Biomed. 2016 Sep 1;29(9):1145–61.
3. Malik
SJ, Keihaninejad S, Hammers A, Hajnal JV. Tailored excitation in 3D with spiral
nonselective (SPINS) RF pulses. Magn Reson Med. 2012 May 1;67(5):1303–15.
4. Cloos
MA, Boulant N, Luong M, Ferrand G, Giacomini E, Le Bihan D, et al. kT-points:
Short three-dimensional tailored RF pulses for flip-angle homogenization over
an extended volume. Magn Reson Med. 2012 Jan;67(1):72–80.
5. Amadon
A, Cloos MA. Method and apparatus for compensating for B1 inhomogeneity in
magnetic resonance imaging by nonselective tailored RF pulses. WO2011/128847A1,
2011.
6. Saekho
S, Yip C, Noll DC, Boada FE, Stenger VA. Fast-kz three-dimensional tailored
radiofrequency pulse for reduced B1 inhomogeneity. Magn Reson Med. 2006
Apr;55(4):719–24.
7. Malik
SJ, Beqiri A, Padormo F, Hajnal JV. Direct signal control of the steady-state
response of 3D-FSE sequences. Magn Reson Med. 2015 Mar 1;73(3):951–63.
8. Malik
SJ, Padormo F, Price AN, Hajnal JV. Spatially resolved extended phase graphs:
Modeling and design of multipulse sequences with parallel transmission. Magn
Reson Med. 2012 Nov 1;68(5):1481–94.
9. Sbrizzi
A, Hoogduin H, Hajnal JV, van den Berg CAT, Luijten PR, Malik SJ. Optimal
control design of turbo spin-echo sequences with applications to
parallel-transmit systems: Optimal Control of TSE Sequences. Magn Reson Med.
2017 Jan;77(1):361–73.
10. Fautz
HP, Vogel M, Gross P, Kerr A, Zhu Y. B1 mapping of coil arrays for parallel
transmission. In: Proceedings of the 16th Annual Meeting of ISMRM, Toronto,
Canada [Internet]. 2008. p. 1247. Available from:
http://cds.ismrm.org/ismrm-2008/files/01247.pdf
11. Carr
HY, Purcell EM. Effects of diffusion on free precession in nuclear magnetic
resonance experiments. Phys Rev. 1954;94(3):630.
12. Meiboom
S, Gill D. Modified spin-echo method for measuring nuclear relaxation times.
Rev Sci Instrum. 1958;29(8):688–691.
13. Yarnykh
VL. Actual flip-angle imaging in the pulsed steady state: A method for rapid
three-dimensional mapping of the transmitted radiofrequency field. Magn Reson
Med. 2007 Jan;57(1):192–200.
14. Nehrke
K. On the steady-state properties of actual flip angle imaging (AFI). Magn
Reson Med. 2009 Jan;61(1):84–92.
15. Woolrich
MW, Jbabdi S, Patenaude B, Chappell M, Makni S, Behrens T, et al. Bayesian
analysis of neuroimaging data in FSL. NeuroImage. 2009 Mar;45(1):S173–86.
16. Smith
SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TEJ, Johansen-Berg H, et al.
Advances in functional and structural MR image analysis and implementation as
FSL. NeuroImage. 2004 Jan;23:S208–19.
17. Jenkinson
M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. FSL. NeuroImage. 2012 Aug;62(2):782–90.