Mahesh Bharath Keerthivasan1,2, Lavanya Umapathy2,3, Jean-Philippe Galons2, Diego Martin4, Ali Bilgin2,3,4, and Maria Altbach2
1Siemens Medical Solutions USA Inc, New York, NY, United States, 2Medical Imaging, University of Arizona, Tucson, AZ, United States, 3Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States, 4Biomedical Engineering, University of Arizona, Tucson, AZ, United States
Synopsis
Radial TSE
techniques have been proposed for abdominal T2-weighted (T2w) imaging and T2
mapping. Slice efficiency of breath-held RADTSE is limited by the specific
absorption rate (SAR). We present a reduced SAR variable refocusing flip angle RADTSE (RADTSE-VFA) technique designed for
efficient slice coverage and improved
T2 estimation. The flip angles
are designed to (1) minimize T2 estimation error, (2) improve lesion-liver
relative contrast, and (3) minimize SAR. RADTSE-VFA generated T2w images with
comparable contrast as constant flip angle RADTSE while resulting in a 60%
increase in slice coverage at a 1.5x reduction in SAR.
Introduction
T2-weighted (T2w) imaging is
routinely used in the clinic for the diagnosis of focal liver lesions. The
difference in signal intensity in T2w images allows radiologists to
differentiate malignancies from the most common benign lesions (cysts and
hemangiomas). The qualitative analysis of T2w images for differentiating
abdominal neoplasms is inherently subjective and prone to inconsistency1.
As an alternative, the estimation of T2 relaxation times through an accelerated radial TSE (RADTSE) sequence2,3 that yields TE images with high spatial and temporal
resolution from a single
breath hold has been
proposed. In breath-held TSE pulse sequences, slice coverage is limited by the
specific absorption rate (SAR). Although, the use of refocusing pulses with
flip angles (FA) less than 180o can be exploited to reduce SAR and
increase slice coverage4,5, the late echoes are affected by noise
due to signal decay which compromises accuracy of T2 estimation. In this work
we present a reduced SAR variable refocusing flip
angle (VFA) RADTSE technique designed for efficient slice coverage and improved T2 estimation.Methods
The
VFA scheme was designed based on 4
control angles6 optimized to (1) maintain the signal
above the noise floor (S/N) through the echo
train length (ETL), (2) minimize variance of T2 estimation based
on the Cramer Rao lower bound (CRLB), (3) improve relative contrast
between malignant lesions and liver,
and (4) reduce SAR. The area
under the T2 decay curve (AUDC) was used to compute S/N; $$$b1_{+rms}$$$ was used as a subject independent measure of SAR; relative
contrast lesion-to-liver contrast was defined as $$$relC_{lesion\_to\_liver}= \frac{s_{lesion} (TE)-s_{liver} (TE)}{s_{liver} (TE)}$$$ and computed at TE=90ms. Figure
1A shows 2D plots of the objective function and constraints for
the CRLB, AUDC, $$$relC_{lesion\_to\_liver}$$$, and $$$b1_{+rms}$$$. Figure 1B shows the VFA scheme
for ETL=32 based on the optimal choice of control angles $$$ \vec{\alpha}=[60^∘,120^∘,40^∘,130^∘] $$$ used in
this study.
Phantom data
were acquired with the VFA and constant FA RADTSE (ETL=32, echo spacing=8ms,
TR=2sec, constant FA=150o). Reference T2 estimates were acquired
using a single-echo spin-echo sequence (32 TEs in increments of 8ms). In vivo data were
acquired on 17 subjects at 1.5T with VFA and constant FA RADTSE (ETL=32, echo
spacing=6.7ms, TR=2.5sec, breath-hold=18sec).
T2w images (32 TEs) were reconstructed from undersampled
data (6 views per TE) using an iterative algorithm2
which incorporates RF pulse information and B1 into the signal model for
accurate T2 estimation. A dictionary based fitting
approach was used to estimate the T2 maps.Results and Discussion
As shown in Figure 2A,
phantom T2 values from VFA and constant FA RADTSE match the reference values. Figure 2B show the decay
curves for phantoms with T240ms and T280ms (representing
liver and malignancies). Note that both
VFA and the constant FA schemes provide a similar relative contrast at TEeff=90ms,
the contrast typically used by radiologists to detect focal liver lesions. The
VFA scheme provides a higher AUDC over the echo train than the constant FA. To
analyze the effect of noise on T2 estimation, noise was added to the phantom
data. As can be seen in Figure 2C, T2 estimation changes significantly with
noise levels for the constant FA with a stronger effect for the phantom with
T2=40ms. This effect is not seen in the VFA scheme, due to the higher signal
level in the late echoes.
Figure 3
shows T2w images (3 out of 32 TEs) and T2 maps for subjects with a hemangioma,
liver metastases, and hepatocellular carcinoma. A major advantage of VFA RADTSE
is that slice
efficiency is 60% higher than the constant FA due to a reduced SAR, (mean SAR VFA=1.08, mean
SAR constant FA=1.62). Moreover, the constant FA T2 maps have “artificially” higher
T2 values in the central part of the liver (arrows), a region of low SNR due to
reduced soil sensitivities. The effect is not seen in the VFA T2 maps. This is consistent
with the results of Figure 2C (for lower T2 species) as well as the AUDC maps
shown in Figure 4 (note the lower AUDC in the central part of liver in the
constant FA compared to the upper liver). The AUDC values for the VFA are
higher through the anatomy and also more uniform within the same organ.
Figure 5A shows the T2
distributions for a total of 24 focal liver lesions. Both constant and VFA
methods yield excellent separation between benign and malignant lesions and the
mean and standard deviation for malignancies and benign are comparable between
the two methods (p=0.435). The relative
contrast between malignant lesions and adjacent liver, which is related to
lesion conspicuity, is higher for the VFA method (p=0.013) as shown in Figure
5B. Conclusion
A variable flip angle radial
TSE sequence optimized for abdominal
imaging was presented. The method yields anatomical images for 32 TEs and a T2
map for 11 slices in a single breath hold. T2 estimation, slice coverage and relative
contrast between malignancies and liver is superior than the constant flip
angle method. An extension of the method
for higher ETLs should provide full liver anatomical coverage with T2 mapping
in only 1-2 breath-holds. Acknowledgements
We would like to acknowledge grant support
from NIH (CA245920), the Arizona Biomedical Research Commission (ADHS14-082996),
and the Technology and Research Initiative Fund Technology and Research
Initiative Fund (TRIF). References
1. Fenlon, H. M., Tello, R.,
Victor, L. S., & Yucel, E. K. (2000). Signal characteristics of focal liver
lesions on double echo T2-weighted conventional spin echo MRI: observer
performance versus quantitative measurements of T2 relaxation times. Journal of
computer assisted tomography, 24(2), 204-211.
2.
Huang C., Bilgin
A., Barr T., Altbach M, T2 relaxometry with indirect echo compensation from highly undersampled data, Magnetic Resonance
in Medicine, 70(4) 2013
3.
Altbach M., Bilgin A., Li Z.,
Clarkson E., Trouard T., Gmitro A.,
Processing
of radial fast spin-echo data for obtaining T2 estimates from a single
k-space data set, Magnetic Resonance in
Medicine, 54 2005
4.
Hennig,
Jürgen, and Klaus Scheffler. "Easy improvement of signal‐to‐noise in RARE‐sequences with low refocusing flip angles." Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine 44.6 (2000): 983-985.
5.
Lebel,
R. Marc, and Alan H. Wilman. "Transverse relaxometry with stimulated echo
compensation." Magnetic resonance in medicine 64.4 (2010): 1005-1014.
6. Busse RF, Brau ACS, Vu A, Michelich
CR, Bayram E, Kijowski R, Reeder SB, Rowley HA. Effects of refocusing flip
angle modulation and view ordering in 3D fast spin echo. Magn Reson Med
2008;60: 640-649