Bhumi Bhusal1, Jason Stockmann2, Azma Mareyam2, John Kirsch2, Lawrence L Wald2, Mark J Nolt1, Joshua Rosenow1, Roberto Lopez-Rosado1, Behzad Elahi1, and Laleh Golestanirad1
1Northwestern University, Chicago, IL, United States, 2Massachusetts General Hospital, Charlestown, MA, United States
Synopsis
We report the results of MRI safety and image artifact assessment of a
commercial deep brain stimulation (DBS) system implanted in an anthropomorphic
phantom, undergoing MRI at 7T. RF
heating was observed to be less than 2°C for all clinically relevant as well as
worst-case configurations evaluated in the study. The magnetic force on the pulse generator was
found to be within the safe limit. Metal-induced image artifact was comparable
to what is observed at lower fields. Our results indicate that 7T MRI could be
performed safely in patients with DBS implants under carefully evaluated device
model and MRI hardware.
Introduction
Deep brain stimulation (DBS) is a neurosurgical procedure increasingly
used to treat neurologic and psychiatric disorders. MRI is useful in DBS
therapy, both for electrode localization and target verification1, and for monitoring functional effects of
stimulation on local and sparse brain networks2-5. To date, studies
that have assessed the safety and feasibility of postoperative MRI for DBS
imaging have focused on 1.5T and 3T scanners due to their clinical prevalence6-11. Ultra-high field
MRI at 7T provides unrivaled opportunity to push the boundaries of DBS target
visualization 12, 13, yet no study has comprehensively assessed safety and image quality of 7T MRI in patients with DBS devices. Here
we report results of RF heating assessment, magnetic force measurement, and
image artifact of a commercial DBS device implanted in an anthropomorphic
phantom undergoing MRI at 7T.Methods
RF exposure: Experiments were
performed on a 7T scanner (Magnetom Terra; Siemens Healthineers, Erlangen,
Germany) using a home-made local transmit/receive head coil consisting of an
8-channel wrap-around receive array nested inside a shielded detunable
quadrature birdcage volume coil (Figure 1 A & B). To allow
realistic configuration of DBS devices, we used an anthropomorphic phantom
consisting of a 3D-printed body-shaped container and a refillable skull
structure designed based on CT images of a DBS patient. The skull was filled
with tissue-mimicking gel having electric and thermal properties similar to
that of brain tissue (σ = 0.43S/m, thermal conductivity =0.55 J/k-s). Flouroptic temperature sensors were attached to the most distal
contact (contact 0) of a Medtronic DBS lead (model 3387 or model 3389) which
was inserted into the skull following the position and penetration angle
similar to surgical approach for targeting subthalamic nucleus DBS (Figure 1 C
& D). We replicated 28 lead-only configurations, where lead trajectories mimicked
those observed in postoperative CT images of patients operated at our
institutions (Figure 2 A & B). Additionally, we replicated 15 full DBS
system configurations, connecting the lead (3389) to an extension (model
3708660) and an implanted pulse generator (IPG) (Activa SC-37603). Finally, we included
trajectories that showed maximum heating at lower fields17. RF heating
measurements were performed using a high SAR HASTE sequence (TE = 99 msec, TR =
2000 msec, FA = 180°, TA=7:04 minutes).
Force measurement: Magnetic force measurement was
performed according to the ASTM F2052 test method, using a custom-made
device which allowed the IPGs to hang freely using a cotton thread (Figure 2C).
The IPG position was varied to find the maximum displacement from the vertical
position.
Image artifact: To assess image artifacts, the gel-filled skull was
replaced with a skull structure that contained a formalin-fixed human cadaveric
brain. A DBS lead (3387) was implanted in the brain following methods similar
to DBS implantation targeting right subthalamic nucleus (STN) (Figure 3). Artifacts
were assessed on MPRAGE, T2*-GRE and T2-TSE sequences optimized for DBS target
visualization14, 15.Results
RF
heating: The mean ± std of temperature rise ΔT was 0.58 ± 0.23°C for the lead-only system with lead 3387, 0.57 ± 0.30°C for the lead-only system with lead 3389 and 0.52 ± 0.32°C for the full system. A single factor ANOVA showed no
significant difference in ΔT between groups (P-value=0.84). The values of
temperature increase and related plots are shown in Figure 4 & 5.
Magnetic force: The maximum displacement of the IPG
from vertical position due to static magnetic force was 25° for IPG Activa PC
(37601) and 36° for IPG Activa SC (37603).
Image artifact: The average
width of the artifact around the DBS contacts was observed to be 4.7mm, 5.4 mm
and 7.3 mm for the sequences T2-TSE, MPRAGE and T2*-GRE respectively.Discussions
There are strong incentives for the application of ultra-high field MRI in
DBS therapy as 1.5T MRI systematically underestimates the boundaries of STN and
globus pallidus, the common DBS targets. Furthermore, the contrast to noise
ratio is significantly improved at ultra-high field MRI compared to low fields,
allowing visualization of DBS small targets. This work represents the first
assessment of MRI safety and image artifact during 7T imaging of DBS. The RF induced heating was observed to be less
than 2°C for all clinically relevant lead configurations, as well as lead configurations
that are shown to generate maximum heating in simulations16. This is
less than the temperature increase that occurs during a fever, and indicates RF
heating of DBS implants may not be worse than what is observed at lower fields17. Magnetic force
measurement showed the force was less than the force due to gravity (angular
displacement <45°), implying no hazard would arise due to magnetic force. Artifact
measurements indicated that the metal-induced artifacts around leads were
comparable to what is observed at lower fields18. The size of artifact, however, was comparable
to the size of subthalamic nucleus, therefore making direct target
visualization challenging.Conclusions
Our results suggest that 7T MRI may be performed safely in patients with
DBS implants for the specific implant models and MRI hardware. However, further
studies with different implant models, scanners and transmit coil will be
required to generalize this result.Acknowledgements
This
work was supported by NIH grant R00EB021320.References
1. Horn
A. The impact of modern-day
neuroimaging on the field of deep brain stimulation. Current opinion in
neurology. 2019;32(4):511-20.
2. Hancu I, Boutet A, Fiveland E, et
al. On the (Non‐) equivalency of
monopolar and bipolar settings for deep brain stimulation fMRI studies of
Parkinson's disease patients. Journal of Magnetic Resonance Imaging.
2019;49(6):1736-49.
3. DiMarzio M, Rashid T, Hancu I, et
al. Functional MRI Signature of
Chronic Pain Relief From Deep Brain Stimulation in Parkinson Disease Patients.
Neurosurgery. 2019;85(6):E1043-E9.
4. Horn A, Wenzel G, Irmen F, et al. Deep brain stimulation induced
normalization of the human functional connectome in Parkinson’s disease. Brain.
2019;142(10):3129-43.
5. Baldermann JC, Melzer C, Zapf A, et
al. Connectivity profile predictive
of effective deep brain stimulation in obsessive-compulsive disorder.
Biological psychiatry. 2019;85(9):735-43.
6. Golestanirad L, Keil B, Angelone LM,
et al. Feasibility of using linearly
polarized rotating birdcage transmitters and close‐fitting receive arrays in
MRI to reduce SAR in the vicinity of deep brain simulation implants. Magnetic
resonance in medicine. 2017;77(4):1701-12.
7. Golestanirad L, Kazemivalipour E,
Keil B, et al. Reconfigurable MRI coil
technology can substantially reduce RF heating of deep brain stimulation
implants: First in-vitro study of RF heating reduction in bilateral DBS leads
at 1.5 T. PloS one. 2019;14(8).
8. Golestanirad L, Iacono MI, Keil B,
et al. Construction and modeling of
a reconfigurable MRI coil for lowering SAR in patients with deep brain
stimulation implants. Neuroimage. 2017;147:577-88.
9. Rezai AR, Phillips M, Baker KB, et
al. Neurostimulation system used for
deep brain stimulation (DBS): MR safety issues and implications of failing to
follow safety recommendations. Investigative radiology. 2004;39(5):300-3.
10. Baker KB, Tkach J, Hall JD, et al. Reduction of magnetic resonance
imaging-related heating in deep brain stimulation leads using a lead management
device. Neurosurgery. 2005;57(4):392-7.
11. Boutet A, Hancu I, Saha U, et al. 3-Tesla MRI of deep brain stimulation
patients: safety assessment of coils and pulse sequences. Journal of
Neurosurgery. 2019;132(2):586-94.
12. Plantinga BR, Temel Y, Duchin Y, et al. Individualized parcellation of the
subthalamic nucleus in patients with Parkinson's disease with 7T MRI.
Neuroimage. 2018;168:403-11.
13. Maruyama S, Fukunaga M, Fautz H-P, et
al. Comparison of 3T and 7T MRI for
the visualization of globus pallidus sub-segments. Scientific Reports.
2019;9(1):1-8.
14. Alkemade A, de Hollander G, Keuken MC,
et al. Comparison of T2*-weighted
and QSM contrasts in Parkinson's disease to visualize the STN with MRI. PloS
one. 2017;12(4):e0176130.
15. Abosch A, Yacoub E, Ugurbil K, Harel
NJN. An assessment of current brain
targets for deep brain stimulation surgery with susceptibility-weighted imaging
at 7 tesla. 2010;67(6):1745-56.
16. Golestanirad L, Kirsch J, Bonmassar G,
et al. RF-induced heating in tissue
near bilateral DBS implants during MRI at 1.5 T and 3T: The role of surgical
lead management. NeuroImage. 2019;184:566-76.
17. Bhusal B, Nguyen BT, Sanpitak PP, et
al. Effect of Device Configuration
and Patient's Body Composition on the RF Heating and Nonsusceptibility Artifact
of Deep Brain Stimulation Implants During MRI at 1.5 T and 3T. J Magn Reson Imaging. 2020.
18. Boutet A, Rashid T, Hancu I, et al. Functional MRI safety and artifacts
during deep brain stimulation: experience in 102 patients. Radiology. 2019;293(1):174-83.