Seung Su Yoon1,2, Michaela Schmidt2, Manuela Rick2, Teodora Chitiboi3, Puneet Sharma3, Tilman Emrich4,5, Christoph Tilmanns6, Ralph Waßmuth6, Jens Wetzl2, and Andreas Maier1
1Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany, 2Magnetic Resonance, Siemens Healthcare GmbH, Erlangen, Germany, 3Siemens Medical Solutions USA, Inc., Princeton, NJ, United States, 4Department of Radiology, University Medical Center, Johannes Gutenberg-University Mainz, Mainz, Germany, 5Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, United States, 6Diagnostikum Berlin, Berlin, Germany
Synopsis
In cardiac MRI using the Late Gadolinium
Enhancement technique, inversion recovery sequences are acquired for the
correct myocardial nulling for optimal image contrast. In clinical
practice, the selection of the proper inversion time to null healthy myocardium
is manually performed by visual inspection. To standardize the
process, we propose an automated deep-learning-based system which selects the “null
inversion time” where the myocardium signal is darkest, and “contrast inversion
time” where the contrast between the myocardium and blood pool is highest. We
validated the system on a prospective study on different scanners. The system
achieved high accuracy in observers’ annotation range.
Introduction
In cardiac MRI, late gadolinium enhancement
(LGE), a contrast enhanced inversion recovery sequence is generally performed
for the assessment of myocardial viability. For the image contrast optimization
in LGE imaging, inversion recovery techniques are used to null healthy myocardium.
In current clinical practice, the selection of the inversion time (TI) for
correct myocardial nulling is done by visual inspection or a manual
post-processing step. We propose an automated deep-learning-based system1
to select the TI where the myocardium
intensities are the darkest (TInull) and the TI where the contrast
is highest between the myocardium and blood pool signal (TIcontrast)
to increase the efficiency and improve the reproducibility. In this work, we
evaluate the system in a prospective study.Methods
The proposed prototype system
automatically selects the TInull and TIcontrast based on
the myocardium and blood pool signals. It consists of three deep neural networks
and takes a TI scout sequence of a mid-ventricular short-axis (SAX) slice as
input. The last frame is forwarded to a localization network based on Unet2
with ResNet3 blocks, which outputs two right ventricle (RV) insertion
points and left ventricle (LV) midpoint. These landmarks are used for
standardizing the SAX orientations. The orientation of the series is
transformed by aligning the RV insertion points to be parallel to the vertical
image axis, the LV point to be the center of the image and cropping a fixed size
of 128x128 pixels with 1mm resolution. Secondly, the different contrasts of the
standardized series over time points are transformed to a uniform, CINE-like contrast
by the style transfer network built based on Unet2 with DenseNet4
blocks. The processed series is then transferred back to the original
orientation. Third, the segmentation network5 trained on CINE images segments the LV endo-/epicardial and RV epicardial
contours. The mean pixel intensities of the myocardium, LV and RV blood pool at
each time point can then be calculated. The time point of the minimum
myocardium signal is selected as TInull. By
taking the TInull as starting point and examining the next TI window
of 80ms, which is defined based on the analysis of TInull (Fig.3),
the time point where the relative difference between the average LV, RV blood
pool and myocardium signal is highest, is selected as TIcontrast. Experiments
The proposed system was evaluated on
TI scout acquisitions in SAX orientations from 130 patients, acquired on $$$1.5\ T$$$
and $$$3\ T$$$ clinical scanners (MAGNETOM Avanto fit, Aera, Skyra and Prisma, Siemens
Healthcare, Erlangen, Germany) at multiple centers. Detailed information about
the data population and acquisition is shown in Fig. 2a. To validate the
system, mean and standard deviation of the absolute difference between TIcontrast and TInull with two observers (both with $$$>20$$$ years of
cardiac MR experience), and interobserver statistic were analyzed based on
Bland-Altmann analysis. To test its robustness, the system was validated with a
segmented inversion recovery CINE TrueFISP pulse sequence with and without
compressed sensing. The results were qualitatively validated (Fig.4).
Furthermore, the system was
integrated into the scanner software (MAGNETOM Vida; Siemens Healthcare,
Erlangen, Germany) and tested online in a prospective study (Fig.2b) ($$$n=6$$$). The
results were qualitatively validated (Fig.5).Results
The mean difference of $$$1.5\ T$$$ data
($$$n=64$$$) with both observers was $$$-21.2\pm14.6\ ms$$$, $$$4.8\pm14.5\ ms$$$ for TInull, TIcontrast respectively
(Fig.3). The mean difference of $$$3\ T$$$ data ($$$n=66$$$) with both observers was $$$-31.4\pm16.2\ ms$$$, $$$-1.9\pm15.2\ ms$$$ for TInull, TIcontrast,
respectively. TInull shows reasonable deviation between the
system output and both observers and was selected in all cases not later than
observers' annotation. The search window for TIcontrast of $$$[$$$TInull, TInull$$$+80\ ms]$$$ was therefore chosen based on the deviation between the
TInull and both observers’ annotations. TIcontrast was
matched or off by one frame and shows high accuracy in the range of observers’
annotations. In addition, the interobserver variability indicates that the mean
differences between two observers were $$$2.6\pm13.0\ ms$$$ for $$$1.5\ T$$$
and $$$-9.1\pm16.0\ ms$$$ for $$$3\ T$$$, and that the task is user
dependent. In some cases, the TIcontrast was selected the TI with
better contrast than the one chosen by observer as shown in Fig. 5a) and 5b).Conclusion
The results demonstrate the
feasibility of the proposed system to automatically select the correct TI. The
proposed system was integrated into a scanner software and successfully tested online.
It shows great potential for improving automation and standardization of LGE
imaging. Future work will focus on clinical evaluation and validation.Acknowledgements
No acknowledgement found.References
- Yoon S, et al. A
Deep-Learning-based Automated Myocardial Inversion Time Selection for Late Gadolinium Enhancement Imaging. SCMR 2021.
- Ronneberger O, et al. U-net: convolutional
networks for biomedical image segmentation. MICCAI 2015.
- He K, et al. Deep residual learning
for image recognition. CVPR 2016.
- Huang G, et al. Densely connected
convolutional networks. CVPR 2017.
- Chitiboi T, et al. Automated
ventricular volume and strain quantification with deep learning. SCMR 2020.