Romain Froidevaux1, Markus Weiger1, and Klaas Paul Pruessmann1
1Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland
Synopsis
Short-T2 imaging needs immediate and rapid encoding, as
provided by zero echo time (ZTE) MRI. However, in ZTE, excitation and early encoding
occur simultaneously and preclude data acquisition in the k-space center, leading
to local undersampling or gap. One way of retrieving the missing data involves
algebraic reconstruction, but it is limited to small gaps and thus requires
short RF pulses that restrain achievable SNR and contrast options.
Here, we demonstrate a method for algebraic reconstruction of large gaps, based
on the knowledge of excitation pulses. It enables the use of longer pulses and overcomes
ZTE flip angle limitations.
Introduction
MRI of
tissues with fast transverse relaxation times T2 or T2*, such as bone or myelin,
requires dedicated pulse sequences capable of acquiring the MR signal quickly
after radiofrequency (RF) excitation (1,2). An efficient method to do so is zero
echo time (ZTE) imaging (3), in which the radial readout
gradient is followed by spin excitation, and data is acquired as soon as
possible after RF transmit-receive (T/R) switching (Fig. 1a). This sequence
timing involves an RF dead time determined in part by the duration of the RF
pulse, which precludes data acquisition in the k-space center (Fig. 1b). If
pulses are short enough, the missed data may be retrieved algebraically, but
only if the gap does not exceed about 3 Nyquist dwells, dk (= 1/FOV with FOV
the field of view) (4). However, due to finite available RF
power, pulse amplitudes and hence flip angles are limited, potentially restricting
SNR and contrast options. If longer pulses are needed, other techniques could be
used, such as PETRA (5), WASPI (6) or HYFI (7). However, they all require additional
data acquisition with changes in readout gradient strength, which may lead to higher
acoustic noise, lower SNR efficiency or propensity to short-T2 related
artefacts (8).
In this work, we demonstrate an alternative method called pulse profile encoded
ZTE (PPE-ZTE) that allows algebraic reconstruction of ZTE datasets acquired with
long RF-pulses. It relies on the knowledge of the spatial profile of the
excitation pulse and its introduction in the ZTE signal model. In this way,
larger flip angles can be realized, thus improving SNR efficiency and
permitting increased T1 contrast in ZTE images, which are commonly proton-density
weighted. Moreover, in PPE-ZTE, the readout gradient strength is kept constant,
hence creating minimal acoustic noise. Theory
To get
sufficiently independent data, PPE-ZTE requires two separate acquisitions of
the k-space support using frequency-swept RF pulses differing only by small
off-resonances $$$+\Delta f$$$ and $$$-\Delta f$$$
respectively (Fig. 2). In this way, the approximately
quadratic phase of the pulse profile is shifted proportionally to $$$\Delta f$$$
while the magnitude and hence the flip angle remain constant over the
FOV.
Reconstruction of a 1D projection in PPE-ZTE is similar to ZTE, with the
exception that the pulse profiles, $$$p_l (r_m )$$$
, are
introduced in the encoding matrix E (Fig. 2):
$$E_{(l,m),n}=e^{(i 2\pi k_m r_n)}\cdot p_l(r_n)$$
where k and r
span k-space and image space respectively. Methods
Imaging
experiments were performed on a 3T Philips Achieva scanner complemented with a
high-performance insert gradient (9). The RF signal was created with a Bruker amplifier (BLA-1000) with
maximum output power of 1000 W. Either hard or sweep pulses were used for
excitation (10). Short-T2 optimized hardware
included: T/R switches capable of switching in about 3 µs (11), an 1H-free birdcage coil (12) and a spectrometer with up to 4 MHz
acquisition bandwidth and a corresponding filter group delay of about 1 µs (13).
A water bottle was used as a phantom, doped with CuSO4 and H2SO4 to match
physiological relaxation times.
The shape of the RF pulses was measured with a pickup coil placed near the
birdcage coil.
The total scan time of ZTE and PPE-ZTE images was matched by adapting the
number of averages.Results
The
robustness with respect to gap size of PPE-ZTE is demonstrated in Figure 3. As
pulse length increases, model violations such as noise or aliased signals
create serious artefacts in conventional ZTE images at large gap (≥4dk). However, PPE-ZTE preserves images quality, independent from the
pulse length.
Figure 4 illustrates improved SNR with PPE-ZTE. In this example, the signal available in ZTE is
limited due to the necessity of using short pulses (max. 2 µs for an imaging bandwidth, BW = 500 kHz). Using longer pulses, the same
image quality is obtained in PPE-ZTE with substantially less peak power. Finally,
with longer pulses and same power, PPE-ZTE is able to create larger flip
angles, increasing signal and SNR by a factor 2 for equal scan time.
In Figure
5, conventional ZTE images with 500 kHz bandwidth are limited to proton-density
contrast. However, with PPE-ZTE, introducing T1 contrast is enabled due to the
larger available flip angles. Discussion
Utilizing
the encoding properties of the RF pulse in PPE-ZTE compensates for the data
lost during excitation and thus renders image quality of algebraic
reconstructions almost independent of the pulse duration. However, it is still
limited by the T/R switching time (including filter group delays), tT/R.
As a rule of thumb, tT/R should not exceed 2-3 dk. Hence, for the
current setup with tT/R
4µs, the achievable bandwidth is
limited to 500 kHz. However, current developments towards even faster T/R switching (14) hold promise for PPE-ZTE at further increased bandwidths. Conclusion
We demonstrated
algebraic reconstruction of ZTE datasets with large k-space gaps by employing
the PPE-ZTE technique. As opposed to conventional ZTE, image fidelity has only
a minor dependence on pulse duration. Hence, significant improvements in SNR
efficiency as well as T1 contrast become possible, even at high imaging bandwidth.
Moreover, compared to approaches using gap filling, PPE-ZTE does not require
changes in gradient strength and therefore produces negligible acoustic noise. Acknowledgements
No acknowledgement found.References
(1) Bydder GM. Imaging of Short and
Ultrashort T2 and T2* Components of Tissues, Fluids, and Materials in the
Body Using Clinical Magnetic Resonance Systems. EMagRes 2012;1:811–22.
(2) Weiger M, Pruessmann KP. Short-T2 MRI:
Principles and recent advances. Prog Nucl Magn Reson Spectrosc
2019;114–115:237–70.
(3) Weiger M, Pruessmann KP. MRI with zero
echo time. Encycl Magn Reson 2012;1:311–22.
(4) Weiger
M, Brunner DO, Tabbert M, Pavan M, Schmid T, Pruessmann KP. Exploring the bandwidth
limits of ZTE imaging: Spatial response, out-of-band signals, and noise
propagation. Magn Reson Med 2015;74:1236–47.
(5) Grodzki DM, Jakob PM, Heismann B.
Ultrashort echo time imaging using pointwise encoding time reduction with
radial acquisition (PETRA). Magn Reson Med 2012;67:510–8.
(6) Wu Y, Dai G, Ackerman JL, Hrovat MI,
Glimcher MJ, Snyder BD, Nazarian A, Chesler D a. Water- and fat-suppressed
proton projection MRI (WASPI) of rat femur bone. Magn Reson Med 2007;57:554–67.
(7) Froidevaux R, Weiger M, Roesler M,
Brunner D, Pruessmann K. HYFI: Hybrid filling of the dead-time gap for faster
zero echo time imaging. MedRxiv 2020. https://doi.org/10.1101/2020.06.16.20132704
(8) Froidevaux
R, Weiger M, Brunner DO, Dietrich BE, Wilm BJ, Pruessmann KP. Filling the dead-time
gap in zero echo time MRI: Principles compared. Magn Reson Med 2018;79:2036–45.
(9) Weiger M, Overweg J, Rösler MB,
Froidevaux R, Hennel F, Wilm BJ, Penn A, Sturzenegger U, Schuth W, Mathlener M,
et al. A high-performance gradient insert for rapid and short-T 2 imaging at
full duty cycle. Magn Reson Med 2018;79:3256–66.
(10) Weiger M, Pruessmann KP, Hennel F. MRI
with zero echo time: Hard versus sweep pulse excitation. Magn Reson Med 2011;66:379–89.
(11) Brunner
DO, Furrer L, Weiger M, Baumberger W, Schmid T, Reber J, Dietrich BE, Wilm BJ,
Froidevaux R, Pruessmann KP. Symmetrically biased T/R switches for NMR and MRI
with microsecond dead time. J Magn Reson 2016;263:147–55.
(12) Rösler
MB, Weiger M, Brunner DO, Schmid T, Pruessmann KP. An RF birdcage coil
designed for an insert gradient coil dedicated to short-T2 MRI. Proc 26th Annu
Meet ISMRM, Honolulu 2017:2668.
(13) Dietrich BE, Brunner DO, Wilm BJ, BarmeT
C, Gross S, Kasper L, Haeberlin M, Schmid T, Vannesjo SJ, Pruessmann KP. A
field camera for MR sequence monitoring and system analysis. Magn Reson Med
2016;75:1831–40.
(14) Christoph Michael Schildknecht, Klaas Paul Pruessmann, Gallium nitride MOSFETs enable transmit-receive switching in less than 100ns,
Submitted to ISMRM 2021