Bhumi Bhusal1, Ehsan Kazemivalipour2, Jasmine Vu1, Stella LIn1, Bach Thanh Nguyen1, John Kirsch3, Elizabeth Nowac4, Julie Pilitsis5, Joshua Rosenow1, Ergin Atalar2, and Laleh Golestanirad1
1Northwestern University, Chicago, IL, United States, 2Bilkent University, Ankara, Turkey, 3Massachusetts General Hospital, Charlestown, MA, United States, 4Illinois Bone and Joint Institute, Wilmette, IL, United States, 5Albany Medical College, Albany, NY, United States
Synopsis
Though there are many studies reporting RF heating of implants in
horizontal MRI scanners, there is almost no literature on vertical scanners
that have 90° rotated transmit coils and a fundamentally different distribution
of RF fields. Here we evaluate RF heating of deep brain stimulation (DBS)
implants during MRI in a 1.2T open-bore vertical scanner compared to a 1.5T
horizontal system with both numerical simulations and experimental
measurements. We found a significant reduction in RF heating using vertical vs
horizontal RF coils which are attributable to the orthogonal orientation of RF
electric fields.
Introduction
RF heating of tissue around conductive implants is a major issue,
restricting access to MRI for thousands of patients annually1-5. To date, majority
of studies on RF heating of implants have been performed on horizontal
close-bore (solenoidal) MRI scanners6-10. Vertical MRI
systems, originally introduced as open low-field scanners, are now available at
1.2T, capable of high-resolution structural and functional imaging. As these
scanners have a 90° rotated B1+, they generate a fundamentally
different RF field distribution, and by proxy RF heating, compared to
horizontal scanners. Almost no implant SAR literature exists for this class of
scanners, except a recent simulation study in 3 patients with deep brain
stimulation (DBS) implants which showed a substantial SAR reduction in vertical
vs horizontal MRI systems11. Here we report results of the first
large-scale simulation study of SAR in 90 patient-derived DBS lead models during
MRI in a 1.2 T OASIS Hitachi scanner compared to a 1.5T horizontal system. Simulation
results were validated in experiments with a commercial DBS device implanted in
an anthropomorphic phantom. We found a significant reduction in SAR (30-fold, P˂1×10-5) and RF heating (10-fold, P ˂ 0.025) of implanted leads
during RF exposure with Hitachi 1.2T OASIS coil compared to Siemens 1.5T Aera
coil.Methods
We
constructed numerical models of a
high-pass radial planar birdcage coil made of 12 vertical rungs tuned to 50.4
MHz (1.2T proton imaging) and a 16-rung high-pass birdcage coil tuned to 64 MHz
(1.5T proton imaging). The coil configurations and respective field distributions
are shown in Figure 1 (A-D). The vertical coil geometry and circuitry was based
on information
provided by the vendor. The horizontal coil was modelled using
dimensions similar to those reported in the literature12. To distinguish the effects of resonance
frequency from field orientation, we also simulated the SAR generated by a
horizontal birdcage coil with the same physical dimensions as the 1.5 T
birdcage coil but tuned to 50.4 MHz for comparison.
Ninety
(90) unique lead models were simulated with realistic trajectories extracted
from postoperative computed tomography (CT) images of 53 patients with both
lead-only and fully implanted DBS systems (Figure 2). To account for variation
in lead trajectories, patients were recruited from two DBS centers operated by
different neurosurgeons (JR and JP). Simulations were performed at head imaging
landmark representing MRI for DBS target visualization.
Experiments were
performed in a 1.2T vertical scanner (Hitachi 1.2T OASIS) and a 1.5T horizontal
scanner (Siemens 1.5T Aera) using an anthropomorphic phantom which consisted of
a human body shaped torso and a 3D printed skull (Figure 3 (I)). The skull was
filled with a tissue-mimicking gel with electric and thermal properties similar
to brain gray matter (σ =0.40 S/m and thermal conductivity 0.55 J/k-s) and the body
was filled with saline (σ =0.48 S/m). The phantom was implanted with 40 cm wires
representing lead-only DBS systems and 100 cm wires representing full DBS system,
as well as a commercial DBS device (Medtronic Inc., Minneapolis, MN). Temperature
increase during MRI was measured using flouroptic temperature probes attached
to the lead tip (Figure 3 (II). For each lead/wire model, four different
trajectories were implemented as shown in Figure 3 (III). Trajectory A
consisted of two concentric loops at the surgical burr hole, Trajectory B
consisted of two concentric loops moved farther from burr hole along the
coronal suture and Trajectory C consisted of two loops above the temporal bone.
Trajectory D has no loop on the skull. Trajectory A and C are shown to minimize
RF heating of fully implanted DBS device at 3T and 1.5T respectively whereas
Trajectory D is shown to produce higher heating at both 1.5T and 3T horizontal
scanners10. A high SAR turbo spin echo (TSE) sequence was used, with parameters
adjusted to produce rms B1+ of 4 µT at both
scanners.Results
The
mean ± standard deviation of maximum value of 0.1g averaged SAR (0.1g-SARmax) was 17±26.8 W/kg for the
1.2T OASIS coil, 519.3±694.6
W/kg for the 1.5 T horizontal birdcage coil, and 608.9±1077.1 W/kg for 1.2T horizontal birdcage
coil. This indicates ~30-fold reduction in mean value of 0.1g-SARmax
in 1.2T vertical coil compared to 1.5T and 1.2T horizontal coils. A paired
one-tailed t-test showed significant reduction in 0.1g-SAR using 1.2T OASIS
vertical coil compared to 1.5 T horizontal birdcage coil (P˂1×10-5, Cohen’s d =0.73) as
well as 1.2T horizontal birdcage coil (P˂1.7×10-4, Cohen’s d =0.55).
The mean± standard
deviation of ΔT was
5.36±7.40 ºC in the 1.5T horizontal scanner and 0.58±0.27 ºC in the 1.2T
vertical scanner, indicating a 10-fold reduction in mean temperature rise using
vertical scanner. Additionally, the worst-case heating at 1.5T horizontal
scanner was 21-fold higher than the worst-case heating at 1.2T vertical scanner
(26.53ºC vs 1.24ºC). A one tailed paired t-test showed a significant reduction
in heating at 1.2T vertical scanner compared to 1.5T horizontal scanner (p ˂
0.025, Cohen’s d =0.64).Discussions and Conclusions
Our work provides strong evidence of a
significantly reduced RF heating around implanted conductive leads generated by
vertical transmit coils compared to conventional birdcage coils. If verified
with other types of implants, this indicates new possibilities for scanning of
patients with electronically active implants.Acknowledgements
This work was supported by the NIH grants R00EB021320 and R03EB025344.References
1. Rezai AR, Phillips M, Baker KB, et
al. Neurostimulation system used for
deep brain stimulation (DBS): MR safety issues and implications of failing to
follow safety recommendations. Investigative radiology. 2004;39(5):300-3.
2. Bhusal B, Bhattacharyya P, Baig T,
et al. Measurements and simulation
of RF heating of implanted stereo‐electroencephalography electrodes during MR
scans. 2018;80(4):1676-85.
3. Vu J, Bhusal B, Nguyen BT,
Golestanirad L. Evaluating Accuracy
of Numerical Simulations in Predicting Heating of Wire Implants During MRI at
1.5 T. 2020 42nd Annual International Conference of the IEEE Engineering in
Medicine & Biology Society (EMBC), 2020. IEEE: 6107-10.
4. Golestanirad L, Kirsch J, Bonmassar
G, et al. RF-induced heating in
tissue near bilateral DBS implants during MRI at 1.5 T and 3T: The role of
surgical lead management. NeuroImage. 2019;184:566-76.
5. Golestanirad L, Angelone LM, Kirsch
J, et al. Reducing RF-induced
heating near implanted leads through high-dielectric capacitive bleeding of
current (CBLOC). IEEE transactions on microwave theory and techniques.
2019;67(3):1265-73.
6. McElcheran CE, Yang B, Anderson KJ,
et al. Parallel radiofrequency
transmission at 3 tesla to improve safety in bilateral implanted wires in a
heterogeneous model. Magnetic resonance in medicine. 2017;78(6):2406-15.
7. Nguyen BT, Pilitsis J, Golestanirad
L. The effect of
simulation strategies on prediction of power deposition in the tissue around
electronic implants during magnetic resonance imaging. Journal of Physics in Medicine and Biology. 2020;65(18):185007.
8. Kazemivalipour E, Keil B, Vali A, et
al. Reconfigurable MRI technology
for low-SAR imaging of deep brain stimulation at 3T: Application in bilateral
leads, fully-implanted systems, and surgically modified lead trajectories.
NeuroImage. 2019;199:18-29.
9. Oh S, Webb AG, Neuberger T, et al. Experimental and numerical assessment
of MRI‐induced temperature change and SAR distributions in phantoms and in
vivo. 2010;63(1):218-23.
10. Bhusal B, Nguyen BT, Sanpitak PP, et
al. Effect of Device Configuration
and Patient's Body Composition on the RF Heating and Nonsusceptibility Artifact
of Deep Brain Stimulation Implants During MRI at 1.5 T and 3T. Journal of
Magnetic Resonance Imaging 2020;(In Press).
11. Golestanirad L, Kazemivalipour E,
Lampman D, et al. RF heating of deep
brain stimulation implants in open‐bore vertical MRI systems: A simulation
study with realistic device configurations. Magnetic Resonance in Medicine. 2020;83(6):2284-92.
12. Yeo DT, Wang Z, Loew W, et al. Local SAR in high pass birdcage and
TEM body coils for multiple human body models in clinical landmark positions at
3T. Journal of magnetic resonance imaging: JMRI. 2011;33(5):1209.