Ali Caglar Özen1,2, Simon Reiss1, Thomas Lottner1, Dursun Korel Yildirim3,4, Ozgur Kocaturk3,5, and Michael Bock1
1Deptartment of Radiology, Medical Physics, University Medical Center Freiburg, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey, 4Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States, 5Transmural Systems, Andover, MA, United States
Synopsis
Interventional devices such as actively visualized catheters or guidewires have a long conductive line between the source and the probe at the tip. Long conductive wires can interract with the electromagnetic fields generated by the transmit coils resulting in RF-induced heating of surrounding tissue. Existing standards do not cover partial immersion and dynamic motion of such devices. In this study, we propose an automated electric field detection and mapping system that is capable of performing hotspot detection, transfer function(TF) measurement and TF validation using different dipole antennae. RF-induced heating can be evaluated without moving the device, thus eliminating positioning errors.
Introduction
RF-induced
heating of implants and devices in the MRI environment is a result of the
coupling of external electromagnetic (EM) transmit fields with elongated
metallic structures [1], [2]. To assess this coupling, Park et al. [3] describe a domain decomposition approach (Tier 3 Clause #8, ISO/TS
10974 Ed2) which analyses the RF exposure by the MRI RF transmit coil using a
set of clinically relevant incident tangential RF electric fields (Etan(z)) along the
lead pathways in combination with a position-independent lead RF response, the
so-called transfer function (TF). Although the ISO
standard is intended for AIMDs, in principle it can also be applied to actively-visualized
interventional catheters and guidewires (AICs and AIGWs). The additional
challenges of AICs and AIGWs that are not covered by the ISO/TS 10974 Ed2 are
that AICs and AIGWs are partially immersed in the body, that their insertion
length is not constant over time, and that the temperature hot spot is not
necessarily an exposed tip.
Previously we showed that insertion length is a
significant parameter in RF-induced heating of AICs [4], [5]. Recently, we also introduced a novel TF
measurement that can perform measurements of partially immersed leads [6]. Using these concepts, in this study TF
measurements are performed in an AIGW for 23.66MHz 64MHz, and 123MHz using a
single setup based on a metal-free broadband electric field sensor. We also
propose a new method for TF validation, where dipole antennae with different
patterns are used to generate uncorrelated Etan(z).Methods
A schematic of the
setup for hotspot detection, TF measurement, and TF-validation modes of the
setup is shown in Fig. 1. A main element of the setup is an electro-optic E-field
sensor (EOS) with a detection bandwidth that is only limited by the sampling
rate of the receiver (UHFLI, Zurich Instruments, Switzerland), 1.8 GSamples/s [7], [8]. E- and B- fields are generated by a dipole antenna and/or a local excitor
which are tuned to the resonance frequency and matched to 50W. In the setup,
the active device (here: an AIGW) can be inserted at various lengths. Here,
insertion lengths of 60, 80, and 100cm were used for the TF measurements and the
validation.
The first step in safety evaluation is hotspot detection. After hotspot
detection, the EOS is positioned close to the hotspot and the excitors specific
to each medium (here, air and saline) are mounted on the translational stage
for TF measurements. Background measurements are also completed and validation
measurements using different dipole antennae are performed by mounting the EOS
back on the translational stage. Finally, to validate the TF, at least two
different dipole antennae are used to generate uncorrelated Etan(z), which is
mapped by the EOS without the AIGW in place. And then absolute E field
scattered around the hotspot is detected to calibrate and validate the TF.
The AIGW used in this study is 150cm-long,
nitinol guidewire with a loopless antenna at the tip (Transmural Systems, Andover, MA). The whole shaft and the
tip of the AIGW was coated by an insulating layer. The measurements were
performed at 23.66MHz,64MHz, and 123MHz corresponding to 0.55T,1.5T, and 3T
static field strengths. The AIGW
was kept in detuned state during the measurements by short circuiting the PIN
diode in the active detuning circuit.Results
Hot spot detection: E field maps for 60,
80, and 100 cm insertion lengths are given in Fig. 1 for 23.66MHz. The hotspot
is located 4mm before the tip of the AIGW at all frequencies for insertion
lengths.
Normalized TFs for different insertion lengths
are given in Fig.2. Calibration and validation measurements were performed
using two different dipole antennae, resembling the Dipole 1 and 2 in Fig.1E.
Estimated SAR values for 1kW homogeneous
incident field are compared in Fig.4. 80cm insertion length results in the lowest SAR
at 23.66MHz, whereas 100cm insertion length yields the lower tip SAR
at 123MHz than 64MHz. For all insertion lengths, SAR is lower at 23.66 MHz.
FDTD simulation results of the dipoles used for
TF validation are given in Fig.5, together with the line profiles along the
AIGW trajectory, which was kept as a straight line to eliminate positioning
errors.Discussion
The main advantage of the EOS measurement
setup is that a single setup is used for hot spot detection, TF evaluation in
cascade of different media, and validation. Performance of our setup was
evaluated on an AIGW. The setup can be improved by addition of dipole sources,
for example, to create circularly polarized incident fields. Recently, a test
setup was proposed, which optimizes trajectories with minimal correlation of
the incident E-field [9]. Etan(z) generated by different dipole antennae, as proposed in this
study, are intrinsically uncorrelated, therefore ideal for validation
measurements. Especially for short leads various dipole excitation patterns can
be used to obtain complete and accurate validation results.
Advantages of using dipole antennae for
calibration and validation of TF can be listed as:
- Compact source, accessible,
easy to reposition and not complicated drive system
- CP, EP, LP fields can be
generated using multiple dipoles driven in parallel
- Device trajectory is fixed, thus position
dependent uncertainties are eliminated.
Acknowledgements
No acknowledgement found.References
[1] F. G. Shellock, “Radiofrequency
Energy-Induced Heating During MR Procedures: A Review,” J. Magn. Reson.
Imaging, vol. 12, no. 1, pp. 30–36, Jul. 2000, doi:
10.1002/1522-2586(200007)12:1<30::AID-JMRI4>3.0.CO;2-S.
[2] C. J. Yeung et al., “RF heating
due to conductive wires during MRI depends on the phase distribution of the
transmit field,” Magn. Reson. Med., vol. 48, no. 6, pp. 1096–1098, 2002,
doi: 10.1002/mrm.10310.
[3] S. M. Park et al., “Calculation
of MRI-induced heating of an implanted medical lead wire with an electric field
transfer function,” J. Magn. Reson. Imaging, vol. 26, no. 5, pp.
1278–1285, 2007, doi: 10.1002/jmri.21159.
[4] A. C. Özen et al., “Safety of
active catheters in MRI: Termination impedance versus RF‐induced heating,” Magn.
Reson. Med., vol. 81, no. 2, pp. 1412–1423, Feb. 2019, doi:
10.1002/mrm.27481.
[5] A. C. Özen et al., “MR safety
watchdog for active catheters: Wireless impedance control with real‐time
feedback,” Magn. Reson. Med., vol. 84, no. 2, pp. 1048–1060, Aug. 2020,
doi: 10.1002/mrm.28153.
[6] T. Lottner et al., “A Transfer
Function Measurement Setup with an Electro-optic Sensor for MR Safety
Assessment in Cascaded Media,” IEEE Trans. Electromagn. Compat., pp.
1–12, 2020, doi: 10.1109/TEMC.2020.3040756.
[7] S. Reiss et al., “An optical
setup for electric field measurements in MRI with high spatial resolution,” Phys.
Med. Biol., vol. 60, no. 11, pp. 4355–4370, Jun. 2015, doi:
10.1088/0031-9155/60/11/4355.
[8] S. Reiss et al., “Analysis of
the RF Excitation of Stents in Small Gap and Overlap Scenarios using an
Electro-optical E-field Sensor,” pp. 1–20.
[9] Y. Wang et al., “A Fast and Accurate Transfer
Function Validation Strategy Using Optimized Rotation-Invariant Lead
Trajectories,” IEEE Trans. Electromagn. Compat., pp. 1–8, 2020, doi:
10.1109/TEMC.2020.3039488.