Berk Silemek1, Frank Seifert1, Bernd Ittermann1, and Lukas Winter1
1Physikalisch-Technische Bundesanstalt (PTB) Braunschweig and Berlin, Berlin, Germany
Synopsis
The sensor Q Matrix ($$$Q_S$$$) concept is introduced for safe MRI of elongated
implants without compromising imaging quality. For this purpose, small and low-cost
sensors, parallel transmission (pTx) and the orthogonal projection method are
sufficient without the need to perform simulations, safety testing or additional
MR imaging. The $$$Q_S$$$
can also be
implemented along with other pTx mitigation methods such as null modes. The $$$Q_S$$$ based
orthogonal projection method and the null modes method are demonstrated in testbed
and MRI experiments.
Introduction
Parallel transmission (pTx) systems
allow to control the RF fields around and the currents inside implants,
providing RF safety along with the necessary degrees of freedom to maintain high
image quality.1–6 Based on E-field and
temperature information, a sensor Q Matrix $$$Q_{S}$$$ is introduced,
which can be applied with small and cheap RMS sensors embedded in an implant.7,8 This safety concept does not require any
simulation or MR imaging to predict, mitigate and measure RF induced heating at
an implant tip in real-time. Furthermore, we investigate the $$$Q_{S}$$$ based pTx mitigation using the orthogonal
projection (OP)5,9 method and the null modes (NM)2,10 approach in terms of RF-induced heating and
image quality.
Methods
As an implant substitute a CAT-8 cable was used and a diode
(ON Semiconductor:
MMDL101T1G) and a thermistor (Murata:
NCP18XH103F03RB)
were embedded at its uninsulated tip. The $$$Q_{S}$$$
is
constructed as follows based on the sensor measurements ($$$X$$$, either RMS
E-field or
temperature change):
$$
Q_{S,kl}^X=\
\begin{array}{l}(X_{kl}-X_{k}-X_{l})+j(X_{kl}^\dagger-X_{k}-X_{l})\;for\;k≠l\;and\;k<l\\
(X_{kl}-X_{k}-X_{l})-j(X_{kl}^\dagger-X_{k}-X_{l})\;for\;k≠l\;and\;k>l\\
\qquad\qquad\qquad\quad2X_{k}\;\;\;\;for\;k=l
\end{array}\quad,\;X\in\mathbb{R}_{>0}
$$
With $$$N^2$$$ measurements for an $$$N$$$-channel pTx system, $$$X_{kl}$$$ denotes the
measured sensor values when transmitting with channel
k and
l and $$$X_{kl}^\dagger$$$ denotes a $$$\frac{\pi}{2}$$$ phase
difference between these channels. The channels’ amplitudes are the same for
all transmissions. For pTx mitigation, eigenvalues of $$$Q_{S}$$$ are
computed and the corresponding Worst-Case(WC)
5,9, orthogonal projection
(OP)
5,9 and null mode (NM)
2 are determined.
The pTx mitigation concept was investigated in an implant-safety
testbed
7 at 300 MHz (Figure-1A) and inside a 3T scanner (Figure-1B).
Testbed ExperimentsDiode based pTx mitigation ($$$Q_S^E$$$
) An automated 8-channel pTx experiment with
diode based $$$Q_S^E$$$ acquisition and pTx mitigation was performed
in a 2D area (140mm×110mm, 10 mm steps) at 154 implant
locations. At every location six measurements using three pTx modes (CP, OP and
NM) (P
fwd=2.86W) were performed:
- Measurement of the RF-induced steady-state
diode signals during RF transmission ($$$3\times1.26\,ms$$$).
- Measurement of the RF-induced heating
rate using a thermistor during RF transmission ( $$$3\times2\,s$$$).
Thermistor based pTx mitigation ($$$Q_S^T$$$)
At 7 of the aforementioned 154 locations,
thermistor based $$$Q_S^T$$$ acquisition ($$$0.5\,s$$$ RF heating and $$$0.5\,s$$$ cooling per measurement, total acquisition time $$$8\times8\times1\,s=64\,s$$$ per location)
and pTx
measurements were performed. The aforementioned 6 measurements were repeated,
now with OP and NM calculated based on $$$Q_S^T$$$.
MRI Experiments
The implant substitute was embedded in a rectangular PVP
phantom together with 9 plastic tubes placed next to the implant. The tubes are
chosen to mimic contrast and resemble a situation where diagnostic imaging is
required outside the implant region. At the central tube, $$$B_1^+$$$ was maximized
by adjusting the phase (same amplitude per channel) based on previously
acquired $$$B_1$$$-maps; this is referred to as the B
1-shim transmission mode. This setup was positioned inside
an 8‐channel pTx RF coil (RAPID Biomedical) and a 3T scanner.$$$Q_S^E$$$ was acquired
using an FID sequence of 64 pulses of 200µs. The OP mode was obtained by
projecting the aforementioned B
1-shim on the WC-orthogonal subspace.
5,9 Then,
GRE images were acquired for all excitation schemes (WC, B
1-shim, OP, NM) using the same total forward power.
Results
Testbed Experiments
Diode based pTx mitigation
The results for the 154 implant locations are shown in Figure-3.
The $$$Q_S^E$$$ acquisition and mitigation is real-time capable
(acquisition and mitigation $$$\sim 20\,ms$$$), and OP was able to reduce RF induced heating in
all locations, while for NM two locations were unsuccessful (<1% higher than
CP). OP and NM reduced the voltage signals by factors $$$\geq\,2$$$ in 99% and 93%,
respectively, of the mapped area compared to CP mode (Figure-3C). RF heating
correlated well to the sensor readings (Figure-3B). NM transmission showed
lower induced signals in 62% of the locations compared to OP (Figure-3C).
Thermistor based pTx mitigation
The results for the seven implant locations are shown in Figure-4.
pTx mitigation was successful for OP and NM compared to WC and CP for all
locations. For OP, thermistor-based mitigation gave slightly better results
compared to the diode sensor.
MRI Experiments
For WC and B1-shim, the signal intensity along the implant
increases, indicating stronger coupling compared to OP and NM (Figure-5). This is
confirmed by the measured diode voltages, which could be reduced substantially by
both OP and NM. Despite successfully suppressed RF heating, image quality produced
by the OP mode is comparable to the B1-shim setting in the target area. Discussion and Conclusion
This work demonstrates how cheap (<1€) and small (<1.5mm3)
RMS sensors can be used to substantially reduce the risk for RF-induced heating
in an implant by using $$$Q_{S}$$$-based pTx mitigation. This
concept was successfully demonstrated in 155 different locations and 932 pTx
excitations for a 300-MHz testbed and inside a 3T MRI. Only the RMS tip sensor
signal was needed to obtain these results; no simulations, imaging techniques
or ex ante safety testing was required. The method works in real-time and the sensors
can also be used as an additional fast and sensitive safety watchdog.11 Both OP and NM mitigation
methods show comparable results in terms of RF induced heating reduction, while
the OP mode gives better imaging quality in regions outside the implant. Acknowledgements
This project
(17IND01 MIMAS) has received funding from the EMPIR programme co-financed by
the Participating States and from the European Union's Horizon 2020 research
and innovation programme. References
1. Eryaman Y, Akin B, Atalar E. Reduction
of implant RF heating through modification of transmit coil electric field. Magn
Reson Med. 2011;65(5):1305-1313. doi:10.1002/mrm.22724
2. Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Controlling radiofrequency-induced
currents in guidewires using parallel transmit. Magn Reson Med.
2015;74(6):1790-1802. doi:10.1002/mrm.25543
3. Silemek
B, Acikel V, Oto C, et al. A temperature sensor implant for active implantable
medical devices for in vivo subacute heating tests under MRI. Magn Reson Med.
2018;79(5):2824-2832. doi:10.1002/mrm.26914
4. McElcheran
CE, Golestanirad L, Iacono MI, et al. Numerical Simulations of Realistic Lead
Trajectories and an Experimental Verification Support the Efficacy of Parallel
Radiofrequency Transmission to Reduce Heating of Deep Brain Stimulation
Implants during MRI. Sci Rep. 2019;9(1):1-14.
doi:10.1038/s41598-018-38099-w
5. Winter
L, Silemek B, Petzold J, et al. Parallel transmission medical implant safety
testbed: Real-time mitigation of RF induced tip heating using time-domain
E-field sensors. Magn Reson Med. 2020;84(6):3468-3484.
doi:https://doi.org/10.1002/mrm.28379
6. Guerin
B, Angelone LM, Dougherty D, Wald LL. Parallel transmission to reduce absorbed
power around deep brain stimulation devices in MRI: Impact of number and
arrangement of transmit channels. Magn Reson Med. 2020;83(1):299-311.
doi:10.1002/mrm.27905
7. Silemek
B, Winter L, Seifert F, Pfeiffer H, Ittermann B. Measurement-based safety
assessment, prediction and mitigation of RF induced implant heating with
parallel transmission: temperature matrix. 28th Annual Meeting of ISMRM.
; 2020:1135.
8. Silemek
B, Winter L, Seifert F, et al. Real-time safety assessment and mitigation of RF
induced implant heating with parallel transmission and low-cost RMS sensors.
28th Annual Meeting of ISMRM. ; 2020:4196.
9. Seifert
F, Weidemann G, Ittermann B. Q matrix approach to control implant heating by
transmit array coils. 23rd Annual Meeting of ISMRM. ; 2015:3212.
10. Godinez
F, Scott G, Padormo F, Hajnal JV, Malik SJ. Safe guidewire visualization using
the modes of a PTx transmit array MR system. Magn Reson Med.
2020;83(6):2343-2355. doi:10.1002/mrm.28069
11. Özen
AC, Silemek B, Lottner T, Atalar E, Bock M. MR safety watchdog for active
catheters: Wireless impedance control with real‐time feedback. Magn Reson
Med. Published online January 21, 2020:mrm.28153. doi:10.1002/mrm.28153
12. Han H,
Moritz R, Oberacker E, Waiczies H, Niendorf T, Winter L. Open Source 3D
Multipurpose Measurement System with Submillimetre Fidelity and First
Application in Magnetic Resonance. Sci Rep. 2017;7(1):13452.
doi:10.1038/s41598-017-13824-z