Mostafa Berangi1,2, Andre Kuehne1, and Thoralf Niendorf1,2
1MRI.TOOLS GmbH, Berlin, Germany, 2Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany
Synopsis
Bio-degradable
implants have the benefit of eliminating implant removal surgery. Yet, it’s essential
to monitor the implant and tissue\bone after implementation. MRI monitoring in
the vicinity of the implant might be compromised by tissue heating and transmission
field inhomogeneities. To address these challenges here an RF transceiver array
is designed, manufactured and evaluated aiming to facilitate 7T MRI in the
vicinity of implants. For this purpose, EMF simulations are performed to derive
an optimum excitation vector that reduces local RF power deposition while
ensuring a proper B1+-field. The RF array is evaluated in
phantom experiments and benchmarked with the simulations.
Introduction
Bio-degradable Mg-based
implants1 promote patient comfort by eliminating implant removal surgery. Post
implantation, the degradation kinetics of the Mg alloys might vary for
different environments/patients. Therefore the healing process of the surrounding
bone/tissues and the corrosion state of the implants need to be monitored2. MRI may be exploited to examine the degradation status of biodegradable implants and to study the
implant/tissue interface3. The metallic nature of conductive Mg implants constitutes
challenges for MRI. This includes 1) MR safety risks due to the elevation of the
specific absorption rate (SAR) in the vicinity of implant4 and 2) B1 field inhomogeneities caused by interferences
of the radio frequency (RF) field with the implant. These constraints can be
offset by leveraging the degrees of freedom facilitated by parallel
transmission using multichannel RF coil arrays. Recognizing this opportunity,
this work examines the applicability of RF array configurations for safe MRI of
biodegradable implants at 7.0T. For this purpose RF array configurations comprising
loop elements and/or fractionated-dipoles5 are characterized in electromagnetic field simulations using the
metrics peak-SAR10g and transmission field uniformity to derive optimum
B1 shim patterns using genetic algorithm (GA). The results are
benchmarked against the birdcage mode6 (CP) and the orthogonal projection method7 (OP) in simulations and in a phantom study with a focus on B1+, specific
absorption rate (SAR) and temperature maps.Methods
EMF simulations
were performed (CST Studio Suite 2020, CST MWS, Darmstadt, Germany) for three cylindrical
array configurations of: (i) 8 loop elements, (ii) 8 dipoles, (iii) 8 building
blocks comprising a loop and a dipole element8. The evaluation focused on achievable implant SAR reduction and
homogenization of B1+ in a cylindrical (r=20mm and
L=110mm) region of interest (ROI) around the implant. EMF shimming was
optimized using a multi-objective genetic algorithm that calculates the
tradeoff-curve of B1 homogeneity in the ROI vs. the implant SAR. The
outcome of the GA was benchmarked against CP and OP shimming patterns.
In simulations,
a cylindrical phantom (r=85mm and L=300mm) mimicking muscle electrical
properties at 297.2 MHz was used. The same PVP-based phantom9 was used for the experiments A 70mm long copper wire with 1mm
radius was used to mimic an implant and aligned in parallel with the phantom at
a 30mm distance from its surface. The RF modules were positioned with 20mm
distance from the phantom and shielded with 30mm distance from the elements. Phantom
experiments were performed on a 7T MRI scanner (SIEMENS Magnetom) where the RF
transceiver array was connected to the scanner through an interface (MRI.TOOLS
GmbH, Germany) to utilize the pTx mode. B1+ maps were
acquired using a fast gradient-echo imaging with pre-saturation10. Characterization of implant heating was conducted with SAR
mapping (simulations) and temperature mapping (experiments) using a high-power
heating regime (Pavearage=175W, t=5 min). Result
Our EMF simulations
revealed that an array of 8 RF building blocks comprising a loop and dipole
element as shown in figure 1 is the superior design. We obtained excellent agreement between
the measured and simulated B1+ maps as outlined in figure 2
and 3, respectively. The GA approach outperforms the CP and OP in the ROI in
terms of both B1+ magnitude and homogeneity as well as SAR efficiency. Statistical
parameters of the simulated B1+ in the ROI are presented in the table of figure2. The SAR maps of CP, OP and GA maps are presented in figure 4 and the CP
and GA temperature maps in figure 5.
The simulated
value of maximum average SAR10g (normalized to 1W excitation) in the whole
phantom using CP, OP and GA is 0.2 $$$ {1}/{kg}$$$ , 0.31 $$$ {1}/{kg}$$$ and 0.24 $$$ {1}/{kg}$$$, respectively.
For the CP mode, the maximum is found at the implant, whereas for both OP and
GA excitation, peak SAR occurs at the surface. Even though overall peak SAR for unit power is
slightly increased comparing GA to CP excitation, B1+ in the ROI is also
significantly increased, leading to $$${B^+_1}/{\sqrt {SAR_{Max}}}$$$ efficiencies of 0.54
$$${\mu T}/{\sqrt {{W}/{kg}}}$$$ and 0.67 $$${\mu T}/{\sqrt {{W}/{kg}}}$$$ for CP and GA, respectively, i.e. showing a 24%
efficiency increase. While OP can significantly reduce implant
SAR, it nevertheless suffers from a peak SAR increase of more than 50%. On the
other hand, the GA method successfully limits both implant as well as overall
peak SAR while still delivering improved B1 in the ROI. The maximum temperature increment in the experiment using GA is 61% lower compared to CP.Discussion
CP excitation
leads to B1 artifacts close to the
implant as well as excessive local RF power deposition. The orthogonal
projection method reduces implant heating but suffers from B1 degradation in
the implant vicinity as well as excessive peak SAR increase outside the implant
region. Our results demonstrate that the multi-objective GA approach affords SAR
reduction in conjunction with a strong and uniform B1+ field in the implant
region.Conclusion
Parallel
transmission using the proposed 8-channel RF array in conjunction with the GA approach helps to meet the requirements of MR safety and transmission field uniformity
to support MRI based monitoring of tissue healing and for the benefit of MRI characterization of
the degradation state of bio-degradable implants. Acknowledgements
This project is
supported by the H. 2020 European Training Network of the European Union
(MgSafe, grant agr. No. 811226)References
[1] B. J. C. Luthringer, F. Feyerabend, and R.
W. Römer, “Magnesium-Based Implants: A Mini-Review,” Magnesium Research.
2014, doi: 10.1684/mrh.2015.0375.
[2] A. Myrissa et al., “In vitro and in
vivo comparison of binary Mg alloys and pure Mg,” Materials Science and
Engineering C, 2016, doi: 10.1016/j.msec.2015.12.064.
[3] D. Paramitha, M. F. Ulum, A. Purnama, D.
H. B. Wicaksono, D. Noviana, and H. Hermawan, “Monitoring degradation products
and metal ions in vivo,” in Monitoring and Evaluation of Biomaterials and
their Performance In Vivo, 2017.
[4] L. Winter, F. Seifert, L. Zilberti, and M.
Murbach, “MRI-Related Heating of Implants and Devices : A Review Physics of
Implant Heating,” pp. 1–20, 2020, doi: 10.1002/jmri.27194.
[5] A. J. E. Raaijmakers et al., “The
fractionated dipole antenna: A new antenna for body imaging at 7 Tesla,” Magnetic
Resonance in Medicine, 2016, doi: 10.1002/mrm.25596.
[6] V. Alagappan et al., “Degenerate
mode band-pass birdcage coil for accelerated parallel excitation,” Magnetic
Resonance in Medicine, 2007, doi: 10.1002/mrm.21247.
[7] L. Winter et al., “Parallel
transmission medical implant safety testbed: Real-time mitigation of RF induced
tip heating using time-domain E-field sensors,” Magnetic Resonance in
Medicine, 2020, doi: 10.1002/mrm.28379.
[8] M. A. Ertürk, A. J. E. Raaijmakers, G.
Adriany, K. Uğurbil, and G. J. Metzger, “A 16-channel combined loop-dipole
transceiver array for 7 Tesla body MRI,” Magnetic Resonance in Medicine,
2017, doi: 10.1002/mrm.26153.
[9] S. Chung, D. Kim, E. Breton, and L. Axel, “Rapid
B1+ mapping using a preconditioning RF pulse with turboFLASH readout,” Magnetic
Resonance in Medicine, 2010, doi: 10.1002/mrm.22423.